DQR ID | Subject | Data Streams Affected |
---|---|---|
D050301.1 | TWP/SMET/C2 - Instrument failure | twpsmet60sC2.b1 |
D050310.1 | TWP/VCEIL/C2 - Ceilometer failure | twpvceil25kC2.00, twpvceil25kC2.b1 |
D050601.5 | TWP/SMET/C2 - Reprocess: Incorrect Units Listed for Vapor Pressure | twpsmet60sC2.b1 |
D060621.4 | TWP/MPL/C2 - Missing data | twpmplC2.00, twpmplC2.a1 |
D060818.2 | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC2.b1 |
D061011.3 | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/2005 | 1955 | 03/03/2005 | 0430 |
Subject: | TWP/SMET/C2 - Instrument failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Upper wind sensor failed. Wind direction was reported at a near constant value of 69 degrees and was significantly different than the lower wind sensor's direction values. Sensor replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2005 | 0600 | 04/01/2005 | 0007 |
Subject: | TWP/VCEIL/C2 - Ceilometer failure |
DataStreams: | twpvceil25kC2.00, twpvceil25kC2.b1 |
Description: | Large data gaps began occurring frequently at the TWP Nauru site 26 February at 06Z. Prior to 25 February, optical cross talk errors had been observed. Beginning on 25 February, data errors were observed. By 26 February, these errors were exhibiting themselves as extended data gaps. These gaps became increasingly frequent between 2/26 and 3/4. All data are missing after 20050304.1550. The instrument was replaced on 29 March and after some subsequent configuration work, the instrument was revived just after 00Z on 1 April. |
Measurements: | twpvceil25kC2.00:
twpvceil25kC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2004 | 0000 | 05/19/2005 | 2359 |
Subject: | TWP/SMET/C2 - Reprocess: Incorrect Units Listed for Vapor Pressure |
DataStreams: | twpsmet60sC2.b1 |
Description: | The netcdf header file incorrectly listed the units of Vapor Pressure as hPa when in fact they were kPa. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/2004 | 2330 | 02/01/2005 | 2130 |
Subject: | TWP/MPL/C2 - Missing data |
DataStreams: | twpmplC2.00, twpmplC2.a1 |
Description: | The MPL was turned off due to a dead detector. |
Measurements: | twpmplC2.00:
twpmplC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 07/13/2006 | 2232 |
Subject: | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC2.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 09/26/2006 | 1400 |
Subject: | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC2.b1 |
Description: | A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 datastream. When reprocessing of this datastream occurs these variables will be added with -9999 (missing) as the value for dates prior to gauge installation. |
Measurements: | twpsmet60sC2.b1:
|