DQR ID | Subject | Data Streams Affected |
---|---|---|
D051209.2 | SGP/AERI/E14 - QC Data corrupted by static | sgpaerisummaryE14.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2005 | 0000 | 11/30/2005 | 2359 |
Subject: | SGP/AERI/E14 - QC Data corrupted by static |
DataStreams: | sgpaerisummaryE14.b1 |
Description: | The E14 AERI had elevated channel 2 short band responsivity and noise equivalent radiance values. These values are normally used to evaluate data quality. The cause of these high values is thought to be due to a static charge. The actual radiance plots are not affected. |
Measurements: | sgpaerisummaryE14.b1:
|