Subject: | SGP/NFOV/C1 - Head temp heater circuit failed |
DataStreams: | sgpnfovC1.a0, sgpnfovC1.b1
|
Description: | The head temp values were erratic. The instrument was removed from service and found the
head temp heater circuit failed in the full-on position frying the detectors. Data are
not useful. |
Measurements: | sgpnfovC1.a0: - Fields flagged by callang: 0 = no flag\n,(callang_flags)
- NIMFR Tube Temperature(tube_temp)
- Detector Temperature(head_temp)
- Ceilometer ambient temperature(ambient_temp)
- Uncalibrated Zenith Radiance(radiance)
- Raw counts from instrument before calibration(raw_counts)
sgpnfovC1.b1: - Uncalibrated Zenith Radiance(radiance)
- Calibrated zenith radiance data corrected for changes in head temperature.(radiance_temp_cor)
- Cosine of Apparent Solar Zenith Angle(cosine_solar_zenith_angle)
- NIMFR Tube Temperature(tube_temp)
- Raw counts from instrument before calibration(raw_counts)
- Ceilometer ambient temperature(ambient_temp)
- Fields flagged by callang: 0 = no flag\n,(callang_flags)
- Detector Temperature(head_temp)
|