DQR ID | Subject | Data Streams Affected |
---|---|---|
D060530.1 | NIM/ECOR/M1 - Data Suspect for NW and East Wind Directions | nim30ecorM1.b1 |
D061117.1 | NIM/MWRP/M1 - Reprocess: K-band calibration drift | nimmwrpM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/26/2005 | 1530 | 12/31/2006 | 2359 |
Subject: | NIM/ECOR/M1 - Data Suspect for NW and East Wind Directions |
DataStreams: | nim30ecorM1.b1 |
Description: | problem will remain for the entire extend of the NIM deployment. |
Measurements: | nim30ecorM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/2006 | 0000 | 11/17/2006 | 1713 |
Subject: | NIM/MWRP/M1 - Reprocess: K-band calibration drift |
DataStreams: | nimmwrpM1.b1 |
Description: | During October and November there was a drift in the K-Band calibration of the MWRP. This resulted in brightness temperatures that are too high. New calibration coefficients were derived from the median value of the noise diode temperature for the months of October and November. |
Measurements: | nimmwrpM1.b1:
|