Data Quality Reports for Session: 103615 User: perez Completed: 01/09/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040421.3NSA/SKYRAD/C1 - itermittant IRT biasnsaskyrad60sC1.b1
D050219.2TWP/SKYRAD/C2 - IRT biastwpskyrad60sC2.b1
D051202.1NSA/SKYRAD/C1 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
nsaskyrad20sC1.a0, nsaskyrad60sC1.b1
D051202.2NSA/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
nsaskyrad20sC2.a0, nsaskyrad60sC2.b1
D051202.3TWP/SKYRAD/C3 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
twpskyrad60sC3.b1
D051202.5TWP/SKYRAD/C1 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
twpskyrad20sC1.a0, twpskyrad60sC1.b1
D051202.6TWP/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2
downwelling longwave
twpskyrad20sC2.a0, twpskyrad60sC2.b1
D060630.31TWP/SKYRAD/C1 - Reprocess: Longwave Calibration errortwpskyrad60sC1.b1
D060630.32TWP/SKYRAD/C2 - Reprocess: Longwave Calibration errortwpskyrad60sC2.b1
D060630.33TWP/SKYRAD/C3 - Reprocess: Longwave Calibration errortwpskyrad60sC3.b1
D060630.34TWP/GNDRAD/C1 - Reprocess: Longwave Calibration errortwpgndrad60sC1.b1
D060630.35TWP/GNDRAD/C2 - Reprocess: Longwave Calibration errortwpgndrad60sC2.b1
D060630.36TWP/GNDRAD/C3 - Reprocess: Longwave Calibration errortwpgndrad60sC3.b1
D060630.37NSA/SKYRAD/C1 - Reprocess: Longwave Calibration errornsaskyrad60sC1.b1
D060630.38NSA/SKYRAD/C2 - Reprocess: Longwave Calibration errornsaskyrad60sC2.b1
D060630.39NSA/GNDRAD/C1 - Reprocess: Longwave Calibration errornsagndrad60sC1.b1
D060630.40NSA/GNDRAD/C2 - Reprocess: Longwave Calibration errornsagndrad60sC2.b1


DQRID : D040421.3
Start DateStart TimeEnd DateEnd Time
03/28/2003020006/21/20060215
Subject:
NSA/SKYRAD/C1 - itermittant IRT bias
DataStreams:nsaskyrad60sC1.b1
Description:
A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky 
temperature measurement then a sudden correction, as compared to AERI, appears to have been 
caused by a combination of a failing IR detector, a poor signal cable, a contaminated 
(with condensation or frost) mirror or lens, and low ambient temperatures. The positive 
bias was apparently not consistant in either frequency or duration but on days when the 
unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. 

The instrument was repaired and recalibrated and the signal cable replaced.
Measurements:nsaskyrad60sC1.b1:
  • Sky/Cloud Infra-Red Temperature Minima(sky_ir_temp_min)
  • Sky/Cloud Infra-Red Temperature Maxima(sky_ir_temp_max)
  • Sky/Cloud Infra-Red Temperature Standard Deviation(sky_ir_temp_std)
  • Sky/Cloud Infra-Red Temperature(sky_ir_temp)


Back To Table of Contents

DQRID : D050219.2
Start DateStart TimeEnd DateEnd Time
02/07/2005045404/17/20052023
Subject:
TWP/SKYRAD/C2 - IRT bias
DataStreams:twpskyrad60sC2.b1
Description:
The IRT was measuring higher sky temperatures than AERI. The problem was corrected by 
changing the mirror.
Measurements:twpskyrad60sC2.b1:
  • Sky/Cloud Infra-Red Temperature Minima(sky_ir_temp_min)
  • Sky/Cloud Infra-Red Temperature Maxima(sky_ir_temp_max)
  • Sky/Cloud Infra-Red Temperature(sky_ir_temp)
  • Sky/Cloud Infra-Red Temperature Standard Deviation(sky_ir_temp_std)


Back To Table of Contents

DQRID : D051202.1
Start DateStart TimeEnd DateEnd Time
11/04/2004190009/09/20052142
Subject:
NSA/SKYRAD/C1 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:nsaskyrad20sC1.a0, nsaskyrad60sC1.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2 and thus the calculated PIR1 and PIR2 
irradiances.  The errors are inversely proportial to temperature.

Users can apply the temperature dependent irradiance corrections shown in the table below. 
 Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 
W/m2.  The error corrections become the same order of magnitude or larger than the 
instrument uncertainties at or below 260K or 9F. 

Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance				
	DIR1	DIR1	DIR2	DIR2
T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]	Correction[W/m2]
109.4	316.13	0.00	316.13	0.00
100.5	311.20	-0.15	311.20	0.12
93.2	307.14	-0.12	307.14	0.10
87.0	303.70	-0.10	303.70	0.08
81.6	300.70	-0.17	300.70	0.14
77.0	298.15	-0.22	298.15	0.18
59.4	288.35	-0.34	288.35	0.27
47.4	281.72	-0.23	281.72	0.18
38.5	276.74	-0.42	276.74	0.34
31.3	272.77	-0.48	272.77	0.38
25.4	269.48	-0.61	269.48	0.49
20.3	266.67	-0.67	266.67	0.54
16.0	264.25	-0.82	264.25	0.66
12.1	262.11	-0.98	262.11	0.79
8.7	260.19	-0.97	260.19	0.78
5.6	258.46	-1.07	258.46	0.86
2.7	256.89	-1.17	256.89	0.94
0.1	255.44	-1.33	255.44	1.07
-2.3	254.11	-1.41	254.11	1.13
-4.5	252.87	-1.43	252.87	1.15
-6.6	251.71	-1.52	251.71	1.22
-8.5	250.63	-1.63	250.63	1.31
-10.4	249.62	-1.73	249.62	1.39
-12.2	248.58	-1.79	248.58	1.44
-15.4	246.82	-1.93	246.82	1.55
-18.3	245.23	-2.05	245.23	1.65
-20.9	243.78	-2.22	243.78	1.78
-23.3	242.44	-2.36	242.44	1.89
-25.5	241.21	-2.58	241.21	2.07
-27.5	240.07	-2.65	240.07	2.13
-29.5	239.01	-2.75	239.01	2.21
-31.3	238.00	-3.02	238.00	2.42
Measurements:nsaskyrad60sC1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)

nsaskyrad20sC1.a0:
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)


Back To Table of Contents

DQRID : D051202.2
Start DateStart TimeEnd DateEnd Time
11/16/2004000009/09/20052133
Subject:
NSA/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:nsaskyrad20sC2.a0, nsaskyrad60sC2.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2 and thus the calculated PIR1 and PIR2 
irradiances.  The errors are inversely proportial to temperature.

Users can apply temperature dependent irradiance corrections shown in the table below.  
Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 W/m2.  
The error corrections become the same order of magnitude or larger than the instrument 
uncertainties at or below 260K or 9F.

Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance				
	DIR1	DIR1	DIR2	DIR2
T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]	Correction[W/m2]
109.4	316.13	0.00	316.13	0.00
100.5	311.20	-0.15	311.20	0.12
93.2	307.14	-0.12	307.14	0.10
87.0	303.70	-0.10	303.70	0.08
81.6	300.70	-0.17	300.70	0.14
77.0	298.15	-0.22	298.15	0.18
59.4	288.35	-0.34	288.35	0.27
47.4	281.72	-0.23	281.72	0.18
38.5	276.74	-0.42	276.74	0.34
31.3	272.77	-0.48	272.77	0.38
25.4	269.48	-0.61	269.48	0.49
20.3	266.67	-0.67	266.67	0.54
16.0	264.25	-0.82	264.25	0.66
12.1	262.11	-0.98	262.11	0.79
8.7	260.19	-0.97	260.19	0.78
5.6	258.46	-1.07	258.46	0.86
2.7	256.89	-1.17	256.89	0.94
0.1	255.44	-1.33	255.44	1.07
-2.3	254.11	-1.41	254.11	1.13
-4.5	252.87	-1.43	252.87	1.15
-6.6	251.71	-1.52	251.71	1.22
-8.5	250.63	-1.63	250.63	1.31
-10.4	249.62	-1.73	249.62	1.39
-12.2	248.58	-1.79	248.58	1.44
-15.4	246.82	-1.93	246.82	1.55
-18.3	245.23	-2.05	245.23	1.65
-20.9	243.78	-2.22	243.78	1.78
-23.3	242.44	-2.36	242.44	1.89
-25.5	241.21	-2.58	241.21	2.07
-27.5	240.07	-2.65	240.07	2.13
-29.5	239.01	-2.75	239.01	2.21
-31.3	238.00	-3.02	238.00	2.42
Measurements:nsaskyrad60sC2.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)

nsaskyrad20sC2.a0:
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)


Back To Table of Contents

DQRID : D051202.3
Start DateStart TimeEnd DateEnd Time
05/21/2004000009/09/20052012
Subject:
TWP/SKYRAD/C3 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:twpskyrad60sC3.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 
irradiances.  The errors are inversely proportial to temperature.

Users can apply temperature dependent irradiance corrections as shown in the table below.  
Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 
W/m2. The error corrections become the same order of magnitude or larger than the instrument 
uncertainties at or below 260K or 9F. Irradiance Correction Table for Skyrad (TWP and 
NSA) DIR1 and DIR2 Irradiance

		DIR1	DIR1	DIR2	DIR2
	T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]       
		Correction[W/m2]
	109.4	316.13	0.00	316.13	0.00
	100.5	311.20	-0.15	311.20	0.12
	93.2	307.14	-0.12	307.14	0.10
	87.0	303.70	-0.10	303.70	0.08
	81.6	300.70	-0.17	300.70	0.14
	77.0	298.15	-0.22	298.15	0.18
	59.4	288.35	-0.34	288.35	0.27
	47.4	281.72	-0.23	281.72	0.18
	38.5	276.74	-0.42	276.74	0.34
	31.3	272.77	-0.48	272.77	0.38
	25.4	269.48	-0.61	269.48	0.49
	20.3	266.67	-0.67	266.67	0.54
	16.0	264.25	-0.82	264.25	0.66
	12.1	262.11	-0.98	262.11	0.79
	8.7	260.19	-0.97	260.19	0.78
	5.6	258.46	-1.07	258.46	0.86
	2.7	256.89	-1.17	256.89	0.94
	0.1	255.44	-1.33	255.44	1.07
	-2.3	254.11	-1.41	254.11	1.13
	-4.5	252.87	-1.43	252.87	1.15
	-6.6	251.71	-1.52	251.71	1.22
	-8.5	250.63	-1.63	250.63	1.31
	-10.4	249.62	-1.73	249.62	1.39
	-12.2	248.58	-1.79	248.58	1.44
	-15.4	246.82	-1.93	246.82	1.55
	-18.3	245.23	-2.05	245.23	1.65
	-20.9	243.78	-2.22	243.78	1.78
	-23.3	242.44	-2.36	242.44	1.89
	-25.5	241.21	-2.58	241.21	2.07
	-27.5	240.07	-2.65	240.07	2.13
	-29.5	239.01	-2.75	239.01	2.21
	-31.3	238.00	-3.02	238.00	2.42
Measurements:twpskyrad60sC3.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)


Back To Table of Contents

DQRID : D051202.5
Start DateStart TimeEnd DateEnd Time
08/02/2004020009/09/20052022
Subject:
TWP/SKYRAD/C1 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:twpskyrad20sC1.a0, twpskyrad60sC1.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 
irradiances.  The absolute value of the errors are inversely proportial to temperature.

Users can apply temperature dependent irradiance corrections shown in the table below.  
Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 W/m2. 
The error corrections become the same order of magnitude or larger than the instrument 
uncertainties at or below 260K or 9F. 

Irradiance Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance

		DIR1	DIR1	DIR2	DIR2
	T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]       
		Correction[W/m2]
	109.4	316.13	0.00	316.13	0.00
	100.5	311.20	-0.15	311.20	0.12
	93.2	307.14	-0.12	307.14	0.10
	87.0	303.70	-0.10	303.70	0.08
	81.6	300.70	-0.17	300.70	0.14
	77.0	298.15	-0.22	298.15	0.18
	59.4	288.35	-0.34	288.35	0.27
	47.4	281.72	-0.23	281.72	0.18
	38.5	276.74	-0.42	276.74	0.34
	31.3	272.77	-0.48	272.77	0.38
	25.4	269.48	-0.61	269.48	0.49
	20.3	266.67	-0.67	266.67	0.54
	16.0	264.25	-0.82	264.25	0.66
	12.1	262.11	-0.98	262.11	0.79
	8.7	260.19	-0.97	260.19	0.78
	5.6	258.46	-1.07	258.46	0.86
	2.7	256.89	-1.17	256.89	0.94
	0.1	255.44	-1.33	255.44	1.07
	-2.3	254.11	-1.41	254.11	1.13
	-4.5	252.87	-1.43	252.87	1.15
	-6.6	251.71	-1.52	251.71	1.22
	-8.5	250.63	-1.63	250.63	1.31
	-10.4	249.62	-1.73	249.62	1.39
	-12.2	248.58	-1.79	248.58	1.44
	-15.4	246.82	-1.93	246.82	1.55
	-18.3	245.23	-2.05	245.23	1.65
	-20.9	243.78	-2.22	243.78	1.78
	-23.3	242.44	-2.36	242.44	1.89
	-25.5	241.21	-2.58	241.21	2.07
	-27.5	240.07	-2.65	240.07	2.13
	-29.5	239.01	-2.75	239.01	2.21
	-31.3	238.00	-3.02	238.00	2.42
Measurements:twpskyrad20sC1.a0:
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)

twpskyrad60sC1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)


Back To Table of Contents

DQRID : D051202.6
Start DateStart TimeEnd DateEnd Time
07/05/2004220009/09/20052058
Subject:
TWP/SKYRAD/C2 - Reprocess: Incorrect datalogger program affecting PIR1 and PIR2 
downwelling longwave
DataStreams:twpskyrad20sC2.a0, twpskyrad60sC2.b1
Description:
The excitation delays using in measuring the PIR1 case thermistor commonly named "Case1" 
and the PIR2 dome thermistor called "Dome2" were discovered to have been incorrectly set 
to 0 in the Skyrad datalogger program when the new CR23X based Skyrad loggers were 
installed.  The result of no delay in these measurements results in an error in the measured 
resistance of the thermistors, Case1 and Dome2, and thus the calculated PIR1 and PIR2 
irradiances.  The absolute value of the errors are inversely proportial to temperature.

Users can apply the temperature dependent irradiance corrections shown in the table below. 
 Recall that the reported uncertainty of PIRs is typically +-.25% to +-.5% or 1 to 2 
W/m2. The error corrections become the same order of magnitude or larger than the instrument 
uncertainties at or below 260K or 9F.
 
Irradiance Correction Table for Skyrad (TWP and NSA) DIR1 and DIR2 Irradiance
		DIR1	DIR1	DIR2	DIR2
	T[F]	Tcase[K]	Correction[W/m2]	Tdome[K]       
		Correction[W/m2]
	109.4	316.13	0.00	316.13	0.00
	100.5	311.20	-0.15	311.20	0.12
	93.2	307.14	-0.12	307.14	0.10
	87.0	303.70	-0.10	303.70	0.08
	81.6	300.70	-0.17	300.70	0.14
	77.0	298.15	-0.22	298.15	0.18
	59.4	288.35	-0.34	288.35	0.27
	47.4	281.72	-0.23	281.72	0.18
	38.5	276.74	-0.42	276.74	0.34
	31.3	272.77	-0.48	272.77	0.38
	25.4	269.48	-0.61	269.48	0.49
	20.3	266.67	-0.67	266.67	0.54
	16.0	264.25	-0.82	264.25	0.66
	12.1	262.11	-0.98	262.11	0.79
	8.7	260.19	-0.97	260.19	0.78
	5.6	258.46	-1.07	258.46	0.86
	2.7	256.89	-1.17	256.89	0.94
	0.1	255.44	-1.33	255.44	1.07
	-2.3	254.11	-1.41	254.11	1.13
	-4.5	252.87	-1.43	252.87	1.15
	-6.6	251.71	-1.52	251.71	1.22
	-8.5	250.63	-1.63	250.63	1.31
	-10.4	249.62	-1.73	249.62	1.39
	-12.2	248.58	-1.79	248.58	1.44
	-15.4	246.82	-1.93	246.82	1.55
	-18.3	245.23	-2.05	245.23	1.65
	-20.9	243.78	-2.22	243.78	1.78
	-23.3	242.44	-2.36	242.44	1.89
	-25.5	241.21	-2.58	241.21	2.07
	-27.5	240.07	-2.65	240.07	2.13
	-29.5	239.01	-2.75	239.01	2.21
	-31.3	238.00	-3.02	238.00	2.42
Measurements:twpskyrad60sC2.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)

twpskyrad20sC2.a0:
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)


Back To Table of Contents

DQRID : D060630.31
Start DateStart TimeEnd DateEnd Time
08/02/2004020012/21/20052205
Subject:
TWP/SKYRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:twpskyrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpskyrad60sC1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)


Back To Table of Contents

DQRID : D060630.32
Start DateStart TimeEnd DateEnd Time
07/05/2004050012/21/20052150
Subject:
TWP/SKYRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:twpskyrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpskyrad60sC2.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)


Back To Table of Contents

DQRID : D060630.33
Start DateStart TimeEnd DateEnd Time
05/20/2004220012/21/20052136
Subject:
TWP/SKYRAD/C3 - Reprocess: Longwave Calibration error
DataStreams:twpskyrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpskyrad60sC3.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)


Back To Table of Contents

DQRID : D060630.34
Start DateStart TimeEnd DateEnd Time
08/02/2004070012/21/20052204
Subject:
TWP/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.35
Start DateStart TimeEnd DateEnd Time
07/05/2004220212/21/20052149
Subject:
TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC2.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.36
Start DateStart TimeEnd DateEnd Time
05/20/2004010012/21/20052137
Subject:
TWP/GNDRAD/C3 - Reprocess: Longwave Calibration error
DataStreams:twpgndrad60sC3.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:twpgndrad60sC3.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.37
Start DateStart TimeEnd DateEnd Time
11/04/2004030012/21/20051853
Subject:
NSA/SKYRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:nsaskyrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsaskyrad60sC1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)


Back To Table of Contents

DQRID : D060630.38
Start DateStart TimeEnd DateEnd Time
11/16/2004163012/21/20052244
Subject:
NSA/SKYRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:nsaskyrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsaskyrad60sC2.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)


Back To Table of Contents

DQRID : D060630.39
Start DateStart TimeEnd DateEnd Time
11/04/2004160012/21/20051855
Subject:
NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D060630.40
Start DateStart TimeEnd DateEnd Time
11/17/2004003712/22/20052152
Subject:
NSA/GNDRAD/C2 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC2.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC2.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)


Back To Table of Contents



END OF DATA