Data Quality Reports for Session: 103679 User: perez Completed: 01/12/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050104.1TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrecttwpskyradC3.00, twpskyrad60sC3.b1


DQRID : D050104.1
Start DateStart TimeEnd DateEnd Time
05/20/2004000009/02/20040400
Subject:
TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect
DataStreams:twpskyradC3.00, twpskyrad60sC3.b1
Description:
On 05/20 during installtion of the new Campbell datalogger system and associated program 
the NIP (S/N 33551E6) was assigned an incorrect calibration coefficient.  A coefficient of 
116.01 W/m2/mV was applied when it should have been 124.378 W/m2/mV.
TWP reports SV0405D and SV0409D detail the activities which lead to this DQR.
Measurements:twpskyradC3.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)

twpskyrad60sC3.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Standard Deviation(short_direct_normal_std)


Back To Table of Contents



END OF DATA