Data Quality Reports for Session: 104013 User: nab_essc Completed: 01/31/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060713.1NIM/MWRP/M1 - 51.25 GHz channel calibration driftednimmwrpM1.b1
D061117.1NIM/MWRP/M1 - Reprocessed: K-band calibration drift correctednimmwrpM1.b1


DQRID : D060713.1
Start DateStart TimeEnd DateEnd Time
Subject:
NIM/MWRP/M1 - 51.25 GHz channel calibration drifted
DataStreams:nimmwrpM1.b1
Description:
After a power outage on May 5 the 51.25 GHz had a slight change in the calibration. The 
resulting LWP computed by using all 6 channels increased of about 0.025 mm (25 g/m2).
Measurements:nimmwrpM1.b1:
  • Microwave brightness temperature(brightnessTemperature)


Back To Table of Contents

DQRID : D061117.1
Start DateStart TimeEnd DateEnd Time
10/01/2006000011/17/20061713
Subject:
NIM/MWRP/M1 - Reprocessed: K-band calibration drift corrected
DataStreams:nimmwrpM1.b1
Description:
During October and November there was a drift in the K-Band calibration of the MWRP. This 
resulted in brightness temperatures that were too high.

New calibration coefficients were derived from the median value of the noise diode 
temperature for the months of October and November and the data were reprocessed to apply the 
corrected calibrations.  The reprocessed data were archived in January 2007.
Measurements:nimmwrpM1.b1:
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved total precipitable water vapor(totalPrecipitableWater)


Back To Table of Contents



END OF DATA