DQR ID | Subject | Data Streams Affected |
---|---|---|
D000219.1 | SGP/SONDE/C1 - Incorrect surface temperature | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D000308.1 | SGP/SONDE - Questionable absolute pressures | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D000308.2 | SGP/SONDE - Questionable absolute pressures | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D000625.1 | SGP/SONDE/C1 - Operator used wrong calibration tape | sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewndcalcC1.c1 |
D000801.2 | SGP/SONDE/C1 - Incorrect absolute pressure | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewndcalcC1.c1 |
D001107.1 | SGP/SONDE/C1 - Incorrect surface temperature | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D001122.1 | SGP/MWR/C1 - calibration checks | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D001205.1 | SGP/MWR/C1 - Time drift | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D010215.5 | SGP/MWR/C1 - IRT failure | sgpmwrlosC1.a1, sgpmwrlosC1.b1 |
D030312.10 | SGP/MWR/C1 - Intermittent Negative Sky Brightness Temperatures | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgp1mwravgC1.c1, sgp5mwravgC1.c1 |
D040318.8 | SGP/SMOS/E13 - 6-month Calibration Checks | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
D040824.1 | SGP/MWR/C1 - Wet window flag "on" more frequently than expected | sgpmwrlosC1.a1, sgpmwrlosC1.b1 |
D980126.1 | Incorrect radiosonde serial numbers | sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D990124.2 | SGP/SONDE - Dry bias in sonde RH | sgpsondeB1.00, sgpsondeB1.a0, sgpsondeB1.a1, sgpsondeB4.00, sgpsondeB4.a0, sgpsondeB4.a1, sgpsondeB5.00, sgpsondeB5.a0, sgpsondeB5.a1, sgpsondeB6.00, sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, DsgpsondeB1.00, DsgpsondeB1.a0, DsgpsondeB1.a1, DsgpsondeB4.00, DsgpsondeB4.a0, DsgpsondeB4.a1, DsgpsondeB5.00, DsgpsondeB5.a0, DsgpsondeB5.a1, DsgpsondeC1.00, DsgpsondeC1.a0, DsgpsondeC1.a1, sgpsondeptuC1.00, DsgpsondeptuB1.00, DsgpsondeptuB4.00, DsgpsondeptuB5.00, DsgpsondeptuB6.00, DsgpsondeptuC1.00, sgpsondewnpnB1.a0, sgpsondewnpnB1.a1, sgpsondewnpnB4.a0, sgpsondewnpnB4.a1, sgpsondewnpnB5.a0, sgpsondewnpnB5.a1, sgpsondewnpnB6.a0, sgpsondewnpnB6.a1, sgpsondewnpnC1.a0, sgpsondewnpnC1.a1, sgpsondewnprC1.a0, sgpsondewnprC1.a1, sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.00, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.00, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewrprB1.00, sgpsondewrprB1.a0, sgpsondewrprB1.a1, sgpsondewrprB4.00, sgpsondewrprB4.a0, sgpsondewrprB4.a1, sgpsondewrprB5.00, sgpsondewrprB5.a0, sgpsondewrprB5.a1, sgpsondewrprC1.00, sgpsondewrprC1.a0, sgpsondewrprC1.a1, DsgpsondeptucalcB1.c1, DsgpsondeptucalcB4.c1, DsgpsondeptucalcB5.c1, DsgpsondeptucalcC1.c1, DsgpsondenogcwrpnB1.c1, DsgpsondenogcwrpnB4.c1, DsgpsondenogcwrpnB5.c1, DsgpsondenogcwrpnC1.c1, DsgpsondenogcptucalcB1.c1, DsgpsondenogcptucalcB4.c1, DsgpsondenogcptucalcB5.c1, DsgpsondenogcptucalcC1.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2000 | 2330 | 02/17/2000 | 2330 |
Subject: | SGP/SONDE/C1 - Incorrect surface temperature |
DataStreams: | sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | It appears as if the operator mis-typed the temperature when entering the surface values into the system. The value appears as 89.0 degC instead of 8.9 degC. Also note that there are no RH data in this sounding other than the initial value of 100%. The reason for this loss of RH is unknown. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2000 | 2328 | 03/04/2000 | 0051 |
Subject: | SGP/SONDE - Questionable absolute pressures |
DataStreams: | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | The sonde surface pressure reading was considerably higher than the corresponding THWAPS pressure (988.0 hPa vs 982.3 hPa). The sonde readings are consistent during the sounding, however, suggesting that this is not a case of a pressure jump at launch but one of sensor offset. This problem will only affect those who use the absolute pressure values. The file altitudes, which are based on the relative pressure changes from the surface, should be fine. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.00:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/04/2000 | 0229 | 03/04/2000 | 0359 |
Subject: | SGP/SONDE - Questionable absolute pressures |
DataStreams: | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | The sonde surface pressure reading was considerably higher than the corresponding THWAPS pressure (992.9 hPa vs 983.0 hPa). The sonde readings are consistent during the sounding, however, suggesting that this is not a case of a pressure jump at launch but one of sensor offset. This problem will only affect those who use the absolute pressure values. The file altitudes, which are based on the relative pressure changes from the surface, should be fine. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.00:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/23/2000 | 2330 | 06/24/2000 | 0100 |
Subject: | SGP/SONDE/C1 - Operator used wrong calibration tape |
DataStreams: | sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewndcalcC1.c1 |
Description: | It appears as if the operators had a problem with the original sounding and launched a second radiosonde. Apparently they used the calibration tape that came with the first radiosonde (serial number 943156405) in place of the one that was correct for the second radiosonde (sn 943156406) and all the data are bad after the first sample (surface values). |
Measurements: | sgpsondewrpnC1.a0:
sgpsondeC1.a0:
sgpsondewrpnC1.00:
sgpsondewndcalcC1.c1:
sgpsondeC1.a1:
sgpsondeC1.00:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/25/2000 | 0533 | 07/25/2000 | 0724 |
Subject: | SGP/SONDE/C1 - Incorrect absolute pressure |
DataStreams: | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewndcalcC1.c1 |
Description: | The surface pressure (983.5 hPa) reported by this radiosonde (sn 94315713) is substantially higher than that reported by the THWAPS (~977) or by the sondes used in the previous and following soundings (976.6, 977.3). This may be a calibration problem or a mechanic problem on launch. The pressure values reported in the remainder of the sounding are consistent with the reported surface pressure, so the altitudes should be correct. The absolute pressure values may be high by about 5 hPa. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.00:
sgpsondewndcalcC1.c1:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/03/2000 | 2030 | 11/03/2000 | 2030 |
Subject: | SGP/SONDE/C1 - Incorrect surface temperature |
DataStreams: | sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | The value for air temperature at the surface in this file is 23.3 degC. This value is too high and resulted from incorrect data entry (typographic mistake) by the operator. The true value as determined from the operator pre-launch logs is 12.3 degC. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.00:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/16/2000 | 2210 | 02/16/2000 | 2250 |
09/21/2000 | 1423 | 09/21/2000 | 1434 |
09/21/2000 | 1506 | 09/21/2000 | 1517 |
09/26/2000 | 1426 | 09/26/2000 | 1434 |
10/02/2000 | 1732 | 10/02/2000 | 1752 |
10/02/2000 | 2052 | 10/02/2000 | 2107 |
10/03/2000 | 1406 | 10/03/2000 | 1421 |
09/06/2001 | 1900 | 09/06/2001 | 2000 |
Subject: | SGP/MWR/C1 - calibration checks |
DataStreams: | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Description: | The MWR calibration was checked by a cryogenic (liquid nitrogen) blackbody target or (in the case of the Sept 2001 time period) by pointing the mirror at an internal blackbody target. The brightness temperatures and vap and liq retrievals are not representative of the sky/atmosphere. |
Measurements: | sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
sgpmwrlosC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/05/2000 | 1955 | 09/18/2000 | 2145 |
Subject: | SGP/MWR/C1 - Time drift |
DataStreams: | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Description: | On 9/18/00, I found that the time on the MWR laptop was 1m 21s slow. The Dimension4 utility had last synchronized the time on 5/5/00. Found that the network DNS settings had been disabled. Added the DNS entry for ntp host CF10. Dimension4 synchronized the clock by adding 80.91s. The time had drifted at a rate of -0.59s/day. |
Measurements: | sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
sgpmwrlosC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/17/2000 | 0000 | 01/24/2001 | 1600 |
Subject: | SGP/MWR/C1 - IRT failure |
DataStreams: | sgpmwrlosC1.a1, sgpmwrlosC1.b1 |
Description: | The IRT was damaged by internal moisture causing sky temperature measurements that are negatively biased compared to those from the AERI. IRT#0517 was returned to the manufacturer for repair and was replaced with spare IRT#1254. |
Measurements: | sgpmwrlosC1.b1:
sgpmwrlosC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/17/1999 | 1800 | 07/31/2002 | 2034 |
Subject: | SGP/MWR/C1 - Intermittent Negative Sky Brightness Temperatures |
DataStreams: | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgp1mwravgC1.c1, sgp5mwravgC1.c1 |
Description: | Several related and recurring problems with the SGP MWRs have been reported dating back to 1999. These problems were due to the occurrence of blackbody signals (in counts) that were half of those expected. The symptoms included noisy data (especially at Purcell), spikes in the data (especially at Vici), negative brightness temperatures, and apparent loss of serial communication between the computer and the radiometer, which results in a self-termination of the MWR program (especially at the CF). Because these all initially appeared to be hardware-related problems, the instrument mentor and SGP site operations personnel (1) repeatedly cleaned and replaced the fiber optic comm. components, (2) swapped radiometers, (3) sent radiometers back to Radiometrics for evaluation (which has not revealed any instrument problems), and (4) reconfigured the computer's operating system. Despite several attempts to isolate and correct it, the problem persisted. It became apparent that some component of the Windows98 configuration conflicted with the DOS-based MWR program or affected the serial port or the contents of the serial port buffer. This problem was finally corrected by upgrading the MWR software with a new Windows-compatible program. |
Measurements: | sgp5mwravgC1.c1:
sgpmwrlosC1.b1:
sgp1mwravgC1.c1:
sgpmwrlosC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/17/1994 | 1415 | 06/17/1994 | 1615 |
11/18/1994 | 2000 | 11/18/1994 | 2100 |
05/05/1995 | 1942 | 05/05/1995 | 2030 |
10/06/1995 | 1800 | 10/06/1995 | 1858 |
05/03/1996 | 1850 | 05/03/1996 | 2100 |
11/18/1996 | 2130 | 11/19/1996 | 0030 |
04/20/1997 | 0000 | 04/20/1997 | 2359 |
10/10/1997 | 1515 | 10/10/1997 | 1845 |
05/07/1998 | 2030 | 05/07/1998 | 2115 |
12/04/1998 | 1600 | 12/04/1998 | 1621 |
04/09/1999 | 1730 | 04/09/1999 | 2100 |
10/07/1999 | 1900 | 10/07/1999 | 1945 |
04/20/2000 | 1727 | 04/20/2000 | 1927 |
05/04/2000 | 1844 | 05/04/2000 | 1910 |
10/19/2000 | 1915 | 10/19/2000 | 2020 |
Subject: | SGP/SMOS/E13 - 6-month Calibration Checks |
DataStreams: | sgp1smosE13.a0, sgp30smosE13.a1, sgp1440smosE13.a0 |
Description: | During the time periods specified, the SMOS.E13 tower was lowered for 6-month Calibration Checks. During these times wind data are incorrect and temperature and humidity data may be incorrect. Note, exact times when tower was down or maintenance was being performed is sometimes unknown. Users should use the remaining data from these days with caution. |
Measurements: | sgp1440smosE13.a0:
sgp30smosE13.a1:
sgp1smosE13.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1999 | 0000 | 12/21/1999 | 2359 |
11/10/2000 | 0000 | 12/08/2000 | 2359 |
01/26/2001 | 0000 | 03/05/2001 | 2359 |
12/18/2001 | 0000 | 12/31/2001 | 2359 |
02/11/2002 | 0000 | 02/28/2002 | 2359 |
02/19/2003 | 0000 | 03/26/2003 | 2359 |
10/15/2003 | 0000 | 11/03/2003 | 2359 |
Subject: | SGP/MWR/C1 - Wet window flag "on" more frequently than expected |
DataStreams: | sgpmwrlosC1.a1, sgpmwrlosC1.b1 |
Description: | The wet window flag was set "on" more frequently than expected during the time periods specified. This indicates the heater has been running more than necessary. In most instances the moisture sensitivity was adjusted at the end of these periods. |
Measurements: | sgpmwrlosC1.b1:
sgpmwrlosC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/1998 | 0000 | 12/31/1999 | 2359 |
Subject: | Incorrect radiosonde serial numbers |
DataStreams: | sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | Radiosondes manufactured at Vaisala's Woburn, MA factory during the first of January 1998 have incorrect serial numbers. Instead of being numbered 8012XXXXX these sondes are numbered 7012XXXXX. Vaisala has been informed of this, has confirmed the problem, and has corrected their numbering for all subsequent sondes. 163 incorrectly numbered sondes were shipped to ARM on 1/9/98. Some have been used at the CF since 1/19/98:1430. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnB5.a1:
sgpsondewrpnB4.a0:
sgpsondewrpnB1.a0:
sgpsondewrpnB6.a0:
sgpsondewrpnB5.a0:
sgpsondewrpnB4.a1:
sgpsondewrpnB6.a1:
sgpsondewrpnC1.a1:
sgpsondewrpnB1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/1992 | 0000 | 05/31/2002 | 2359 |
Subject: | SGP/SONDE - Dry bias in sonde RH |
DataStreams: | sgpsondeB1.00, sgpsondeB1.a0, sgpsondeB1.a1, sgpsondeB4.00, sgpsondeB4.a0, sgpsondeB4.a1, sgpsondeB5.00, sgpsondeB5.a0, sgpsondeB5.a1, sgpsondeB6.00, sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, DsgpsondeB1.00, DsgpsondeB1.a0, DsgpsondeB1.a1, DsgpsondeB4.00, DsgpsondeB4.a0, DsgpsondeB4.a1, DsgpsondeB5.00, DsgpsondeB5.a0, DsgpsondeB5.a1, DsgpsondeC1.00, DsgpsondeC1.a0, DsgpsondeC1.a1, sgpsondeptuC1.00, DsgpsondeptuB1.00, DsgpsondeptuB4.00, DsgpsondeptuB5.00, DsgpsondeptuB6.00, DsgpsondeptuC1.00, sgpsondewnpnB1.a0, sgpsondewnpnB1.a1, sgpsondewnpnB4.a0, sgpsondewnpnB4.a1, sgpsondewnpnB5.a0, sgpsondewnpnB5.a1, sgpsondewnpnB6.a0, sgpsondewnpnB6.a1, sgpsondewnpnC1.a0, sgpsondewnpnC1.a1, sgpsondewnprC1.a0, sgpsondewnprC1.a1, sgpsondewrpnB1.00, sgpsondewrpnB1.a0, sgpsondewrpnB1.a1, sgpsondewrpnB4.00, sgpsondewrpnB4.a0, sgpsondewrpnB4.a1, sgpsondewrpnB5.00, sgpsondewrpnB5.a0, sgpsondewrpnB5.a1, sgpsondewrpnB6.00, sgpsondewrpnB6.a0, sgpsondewrpnB6.a1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1, sgpsondewrprB1.00, sgpsondewrprB1.a0, sgpsondewrprB1.a1, sgpsondewrprB4.00, sgpsondewrprB4.a0, sgpsondewrprB4.a1, sgpsondewrprB5.00, sgpsondewrprB5.a0, sgpsondewrprB5.a1, sgpsondewrprC1.00, sgpsondewrprC1.a0, sgpsondewrprC1.a1, DsgpsondeptucalcB1.c1, DsgpsondeptucalcB4.c1, DsgpsondeptucalcB5.c1, DsgpsondeptucalcC1.c1, DsgpsondenogcwrpnB1.c1, DsgpsondenogcwrpnB4.c1, DsgpsondenogcwrpnB5.c1, DsgpsondenogcwrpnC1.c1, DsgpsondenogcptucalcB1.c1, DsgpsondenogcptucalcB4.c1, DsgpsondenogcptucalcB5.c1, DsgpsondenogcptucalcC1.c1 |
Description: | Vaisala has confirmed ARM findings of an apparent dry bias in the relative humidity measured by RS-80H radiosondes. The cause of the dry bias is thought to be contamination of the humidity sensor by volatile organic substances originating from some plastic parts of the radiosonde. The amount of contamination is a function of the time between the date of sonde manufacture and its use. All RS-80H sondes manufactured before week 34 of 1998 will show this bias. After week 34 of 1998 Vaisala changed its packaging to reduce, but not eliminate the contamination problem. Starting with RS-80 radiosonde manufactured in late June 2000 Vaisala enclosed the sensor boom in an inert plastic shield, thereby eliminating the contamination that caused the dry bias. Starting in May 2001 at the SGP, May 2002 at the TWP, and later 2002 at the NSA, ARM has moved to using RS-90 radiosondes. These sondes are not subject to the contaminatino that caused the dry bias. Vaisala is in the process of developing an algorithm that can be used to estimate the correct RH from knowledge of the sonde age. All of the ARM sounding data have sufficient metadata available to apply the correction. Additionally, ARM has funded a Science Team effort (Milosevich) to develop a 'best' correction algorithm for the RS-80 radiosonde humidity data. When completed this algorithm will allow us to reprocess the accumulated RS-80 data and produce a new data platform with what we hope will be more accurate data. |
Measurements: | sgpsondeB4.00:
DsgpsondeB5.00:
sgpsondewrpnC1.00:
sgpsondewnpnC1.a1:
DsgpsondeptuB6.00:
DsgpsondenogcptucalcB1.c1:
sgpsondewrpnB6.a0:
DsgpsondeptuC1.00:
sgpsondewnprC1.a0:
DsgpsondenogcptucalcB5.c1:
DsgpsondeptucalcC1.c1:
sgpsondewnpnB4.a1:
sgpsondeB6.00:
sgpsondewrprB4.00:
DsgpsondeptuB1.00:
DsgpsondenogcptucalcC1.c1:
sgpsondewrpnC1.a1:
sgpsondewrpnB1.a1:
sgpsondewrprC1.00:
sgpsondewrpnB1.a0:
sgpsondewrprB5.a1:
sgpsondeB5.a1:
sgpsondewrprB5.00:
DsgpsondeC1.00:
sgpsondewrprB1.a1:
sgpsondeB1.00:
sgpsondeptuC1.00:
DsgpsondenogcptucalcB4.c1:
DsgpsondeB1.a0:
DsgpsondeC1.a1:
DsgpsondeptucalcB5.c1:
sgpsondewrprB1.00:
sgpsondewrpnC1.a0:
sgpsondewrpnB1.00:
sgpsondeB1.a0:
DsgpsondeC1.a0:
DsgpsondeB4.a1:
DsgpsondeptuB4.00:
sgpsondeB1.a1:
sgpsondewrprB5.a0:
sgpsondeC1.00:
sgpsondewnpnB1.a1:
sgpsondewrpnB6.a1:
sgpsondewrprC1.a0:
sgpsondeC1.a0:
sgpsondewrprC1.a1:
DsgpsondeptuB5.00:
sgpsondewrpnB5.a0:
sgpsondewrprB4.a1:
DsgpsondeB1.a1:
sgpsondeB5.a0:
sgpsondewnpnB6.a0:
DsgpsondeB5.a1:
DsgpsondenogcwrpnB1.c1:
sgpsondewrprB1.a0:
sgpsondewrpnB5.00:
DsgpsondeB4.a0:
DsgpsondenogcwrpnB4.c1:
DsgpsondeB5.a0:
sgpsondewnpnB4.a0:
sgpsondewnprC1.a1:
sgpsondewrpnB5.a1:
sgpsondewnpnB5.a0:
DsgpsondeB4.00:
sgpsondewrpnB6.00:
sgpsondewrprB4.a0:
sgpsondewnpnB1.a0:
sgpsondeB4.a1:
sgpsondeC1.a1:
sgpsondewrpnB4.00:
DsgpsondenogcwrpnB5.c1:
sgpsondewnpnC1.a0:
sgpsondewrpnB4.a0:
DsgpsondeB1.00:
sgpsondeB4.a0:
sgpsondewnpnB6.a1:
sgpsondewrpnB4.a1:
sgpsondewnpnB5.a1:
DsgpsondeptucalcB4.c1:
DsgpsondenogcwrpnC1.c1:
sgpsondeB5.00:
DsgpsondeptucalcB1.c1:
|