Subject: | TWP/VCEIL/C3 - low receiver sensitivity |
DataStreams: | twpvceil25kC3.b1 |
Description: | The ceilometer reported frequent receiver alarms. There are periods when the receiver sensitivity dropped to about 50%. The problem was corrected by replacing the instrument with a spare. |
Measurements: | twpvceil25kC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2005 | 0000 | 03/31/2006 | 2359 |
Subject: | TWP/ARSCL/C3 - Reprocess: ARSCL CloudBasePrecipitation field incorrect |
DataStreams: | twparscl1clothC3.c1, twparsclcbh1clothC3.c1, twpmmcrmode1bl200511041clothC3.c1, twpmmcrmode2ci200511041clothC3.c1, twpmmcrmode3ge200511041clothC3.c1, twpmmcrmode4pr200511041clothC3.c1 |
Description: | The CloudBasePrecipitation field is incorrect and has no physical meaning. An ARSCL processing error occurred when the smet60s data stream format changed, causing the wrong data field to be used to supply precipitation measurements to the ARSCL VAP. The error caused no significant degradation in other ARSCL fields, since manual quality control corrected any related errors. |
Measurements: | twparsclcbh1clothC3.c1:
twpmmcrmode4pr200511041clothC3.c1:
twpmmcrmode2ci200511041clothC3.c1:
twpmmcrmode3ge200511041clothC3.c1:
twparscl1clothC3.c1:
twpmmcrmode1bl200511041clothC3.c1:
|