DQR ID | Subject | Data Streams Affected |
---|---|---|
D040424.3 | TWP/MPL/C1 - SunShade timing problem | twpmplC1.00, twpmplC1.a1, twpmplcmask1clothC1.c1, twpmplsmask1clothC1.c1 |
D040424.4 | TWP/MPL/C1 - SP laser diode supply failed | twpmplC1.00, twpmplC1.a1, twpmplcmask1clothC1.c1, twpmplsmask1clothC1.c1 |
D050718.1 | TWP/MPL/C1 - energy monitor spikes | twpmplC1.00, twpmplC1.a1 |
D060225.1 | TWP/MPL/C1 - Energy monitor values are erroneous | twpmplC1.00, twpmplC1.a1 |
D060324.1 | TWP/MPL/C1 - Time stamps are incorrect. | twpmplC1.00, twpmplC1.a1 |
D060901.1 | TWP/MPL/C1 - Data dropouts | twpmplC1.a1, twpmplcmask1clothC1.c1, twpmplsmask1clothC1.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/2003 | 0000 | 02/17/2004 | 2359 |
Subject: | TWP/MPL/C1 - SunShade timing problem |
DataStreams: | twpmplC1.00, twpmplC1.a1, twpmplcmask1clothC1.c1, twpmplsmask1clothC1.c1 |
Description: | For an unknown reason, the sun shade was dropping for less then 10 minutes during the night at about 19:00 UTC between Dec 14, 2003 and Feb 17, 2004. Data taken during that ten minute interval is not looking at the sky and should be considered as though it were missing. This is a fairly small effect. Data users can simply average across the interval or block the records as "missing". |
Measurements: | twpmplC1.a1:
twpmplC1.00:
twpmplsmask1clothC1.c1:
twpmplcmask1clothC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2004 | 0000 | 04/18/2004 | 2359 |
Subject: | TWP/MPL/C1 - SP laser diode supply failed |
DataStreams: | twpmplC1.00, twpmplC1.a1, twpmplcmask1clothC1.c1, twpmplsmask1clothC1.c1 |
Description: | The Spectra Physics laser diode supply failed on March 2, 2004. A replacement was installed April 18. Any data between these dates is worthless. |
Measurements: | twpmplC1.a1:
twpmplC1.00:
twpmplsmask1clothC1.c1:
twpmplcmask1clothC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2005 | 0000 | 07/16/2005 | 2300 |
Subject: | TWP/MPL/C1 - energy monitor spikes |
DataStreams: | twpmplC1.00, twpmplC1.a1 |
Description: | The energy monitor spiked to abnormal values, from 2 up to 25 microjoules per pulse. At the beginning, it was only a few intermittent spikes, but the spikes became more frequent. Most of June does not show a spiky energy monitor. On June 28-30, July 1-10 a diurnal pattern shows up. On July 12-13 it abruptly goes away and returns July 14-16 for 24hrs. Then abruptly stops again 17-19. Except for the 12th, the last changes occurred at 00:00 UTC suggesting a piece of equipment is turning on and off. I do not know why a noisy energy monitor would have a day/night or 00:00UTC dependence. |
Measurements: | twpmplC1.a1:
twpmplC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/11/2005 | 1200 | 02/25/2006 | 2300 |
Subject: | TWP/MPL/C1 - Energy monitor values are erroneous |
DataStreams: | twpmplC1.00, twpmplC1.a1 |
Description: | Energy monitor values should not be used to normalization the backscatter profiles. A problem with the energy monitor electronics forced the value to maximum. The problem was corrected. |
Measurements: | twpmplC1.a1:
twpmplC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2006 | 0500 | 03/22/2006 | 2100 |
Subject: | TWP/MPL/C1 - Time stamps are incorrect. |
DataStreams: | twpmplC1.00, twpmplC1.a1 |
Description: | The computer was experiencing time synchronization problems. The time stamp for the data is incorrect. The synchronization continued to get worse throughout the period. Secondary results of the sync problem are data gaps. The problem exhibited itself with data gaps at the end of every hour. As the computer continued to get further out of sync, the duration of the data gaps increased. The error in time synchronization can be corrected by comparing the MPL data with another instrument such as the ceilometers or MMCR. The error in time can be determined by looking for identical cloud structure. Then shift the MPL data in time to match the other instruments. Remember the error is not constant; it continues to get worst throughout the period. |
Measurements: | twpmplC1.a1:
twpmplC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/19/2006 | 1800 | 08/28/2006 | 0041 |
Subject: | TWP/MPL/C1 - Data dropouts |
DataStreams: | twpmplC1.a1, twpmplcmask1clothC1.c1, twpmplsmask1clothC1.c1 |
Description: | Data dropouts occurred during this period. Problem ended when MPL computer was rebooted. |
Measurements: | twpmplC1.a1:
twpmplsmask1clothC1.c1:
twpmplcmask1clothC1.c1:
|