DQR ID | Subject | Data Streams Affected |
---|---|---|
D060928.3 | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry | sgpsondewnpnC1.b1 |
D070710.2 | SGP/SONDE/C1 - Reprocess: Incorrect surface temperature and RH | sgpsondewnpnC1.b1 |
D070719.1 | SGP/SONDE/C1 - Reprocess: Bad surface RH | sgpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2006 | 1728 | 12/31/9999 | 2359 |
Subject: | SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry |
DataStreams: | sgpsondewnpnC1.b1 |
Description: | The output precision of the following variables was increased 20060928 alt changed from 1m to 0.1m dp changed from 0.1degC to 0.01degC pres changed from 0.1 hPa to 0.01 hPa RH changed from 1%RH to 0.01%RH tdry changed from .1degC to 0.01degC Note that the increase in precision does not necessarily imply an increase in accuracy. Users are cautioned to continue to use their best scientific judgement when using these or any other data. |
Measurements: | sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/03/2007 | 1729 | 07/03/2007 | 1730 |
07/03/2007 | 2331 | 07/03/2007 | 2332 |
Subject: | SGP/SONDE/C1 - Reprocess: Incorrect surface temperature and RH |
DataStreams: | sgpsondewnpnC1.b1 |
Description: | Because of a system problem these two soundings (20070703:1729 and 20070703:2331) have incorrect surface values for temperature and RH. Sounding BadTemp BadRH GoodTemp GoodRH 17:29 22.60 22.61 27.61 66.97 23:31 22.59 0.00 29.17 64.31 The system has been fixed. |
Measurements: | sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2007 | 1127 | 07/11/2007 | 1337 |
Subject: | SGP/SONDE/C1 - Reprocess: Bad surface RH |
DataStreams: | sgpsondewnpnC1.b1 |
Description: | It appears that there was some sort of problem with the RH sensor heating cycle regulator on this radiosonde. The surface RH value (66.6% RH) is incorrect and should be 78.7% RH. When I looked at the raw data to try to understand the problem, I noted that the RH value in the non-heated sensor did not fully recover to ambient before the next sensor heating began. The problem seemed to go away after launch, but it is hard to be 100% sure. |
Measurements: | sgpsondewnpnC1.b1:
|