DQR ID | Subject | Data Streams Affected |
---|---|---|
D060808.1 | SGP/MWR/B6 - Reprocessed: Corrected calibration | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
D060821.1 | SGP/MWR/B6 - Data missing | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
D060927.2 | SGP/MWR/B6 - Computer problem | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
D061009.2 | SGP/MWR/B6 - Computer failure; Data missing | sgpmwrB6.00, sgpmwrlosB6.b1, sgpmwrtipB6.a1, sgp5mwravgB6.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/15/2006 | 1900 | 08/18/2006 | 0800 |
Subject: | SGP/MWR/B6 - Reprocessed: Corrected calibration |
DataStreams: | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Description: | The instrument was overheating so the radiometer temperature set point was increased from 56.3 C to 57.8 C. The data have been reprocessed with corrected calibration coefficients. Reprocessed MWRLOS data were archived January 2007. MWRTIP data will not be reprocessed. |
Measurements: | sgpmwrtipB6.a1:
sgpmwrlosB6.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/02/2006 | 0300 | 07/05/2006 | 2100 |
07/26/2006 | 1200 | 08/04/2006 | 1500 |
Subject: | SGP/MWR/B6 - Data missing |
DataStreams: | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Description: | Data are missing because of computer problem and power outage at the site. |
Measurements: | sgpmwrtipB6.a1:
sgpmwrlosB6.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/2006 | 0500 | 08/11/2006 | 1600 |
08/26/2006 | 0900 | 08/29/2006 | 1900 |
Subject: | SGP/MWR/B6 - Computer problem |
DataStreams: | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Description: | The computer was found powered off several times and the operators were unable to reboot it remotely |
Measurements: | sgpmwrtipB6.a1:
sgpmwrlosB6.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/08/2006 | 1600 | 09/12/2006 | 2100 |
Subject: | SGP/MWR/B6 - Computer failure; Data missing |
DataStreams: | sgpmwrB6.00, sgpmwrlosB6.b1, sgpmwrtipB6.a1, sgp5mwravgB6.c1 |
Description: | The computer powered off and it was not possible to restart it remotely. Data are missing between 9/8 and 9/12. |
Measurements: | sgpmwrtipB6.a1:
sgpmwrB6.00:
sgpmwrlosB6.b1:
sgp5mwravgB6.c1:
|