DQR ID | Subject | Data Streams Affected |
---|---|---|
D050920.2 | NSA/MWR/C1 - Instrument problem, Heater problem | nsamwrC1.00, nsamwrlosC1.b1 |
D051020.1 | NSA/MWR/C1 - Incorrect time stamp | nsamwrC1.00, nsamwrlosC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/13/2005 | 0000 | 10/18/2005 | 2000 |
Subject: | NSA/MWR/C1 - Instrument problem, Heater problem |
DataStreams: | nsamwrC1.00, nsamwrlosC1.b1 |
Description: | The heater of the MWR was ON all the time even when the moisture sensor was dry. Jim Liljegren suspected a problem with the relay that activates the heater. The suggested action was to send the faulty assembly to SGP for repair and get a spare unit from SGP. The spare unit was sent on 9/15 and was installed on 10/18. |
Measurements: | nsamwrC1.00:
nsamwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2005 | 2200 | 09/30/2005 | 2000 |
Subject: | NSA/MWR/C1 - Incorrect time stamp |
DataStreams: | nsamwrC1.00, nsamwrlosC1.b1 |
Description: | On Sept 27 the motherboard of the MWR computer was replaced. Following the replacement, the date on the computer was not set to the correct time of day, therefore the date and time in the files were wrong. The time on the computer was reset on Sept. 30. Time stamps between Sept. 27 and Sept. 30 were later corrected based on gathered information, however they may have as much as 5 minutes of uncertainty. |
Measurements: | nsamwrC1.00:
nsamwrlosC1.b1:
|