Data Quality Reports for Session: 106911 User: kmraisan Completed: 08/08/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D051031.3SGP/ECOR/E14 - CO2/H2O Incorrectsgp30ecorE14.b1
D051216.2SGP/ECOR/E14 - Flux Measurements Incorrectsgp30ecorE14.b1


DQRID : D051031.3
Start DateStart TimeEnd DateEnd Time
08/25/2005000008/29/20050230
Subject:
SGP/ECOR/E14 - CO2/H2O Incorrect
DataStreams:sgp30ecorE14.b1
Description:
The LI-7500 CO2/H2O sensor stopped working after a rainstorm.  CO2 and H2O data are 
incorrect during the stated period.
Measurements:sgp30ecorE14.b1:
  • skewness of variable q(skew_q)
  • rotated covariance uq(cvar_rot_uq)
  • mean value of out of range values and spikes of c -9999 if no spikes(mean_spk_c)
  • covariance of qc(cvar_qc)
  • number of c samples removed due to spikes(n_spk_c)
  • rotated covariance wq(cvar_rot_wq)
  • covariance uq(cvar_uq)
  • rotated covariance uc(cvar_rot_uc)
  • number of valid q samples(n_good_q)
  • number of samples with \IRGA hardware problem\ flag(n_bad_irga)
  • 0=real or 1=dummy value of lv(real_lv)
  • average temperature (IGRA internal sensor)(temp_irga)
  • rotated covariance vc(cvar_rot_vc)
  • number of bad or out of range c samples(n_bad_c)
  • mean value of out of range values and spikes of q -9999 if no spikes(mean_spk_q)
  • variance of variable q(var_q)
  • number of samples with IRGA optical path blocked flag(n_bad_irga_light)
  • variance of variable c(var_c)
  • rotated covariance wc(cvar_rot_wc)
  • covariance uc(cvar_uc)
  • covariance of tq(cvar_tq)
  • number of q samples removed due to spikes(n_spk_q)
  • mean water vapor concentration (q)(mean_q)
  • kurtosis of variable c(kurt_c)
  • covariance of tc(cvar_tc)
  • covariance vc(cvar_vc)
  • covariance wq(cvar_wq)
  • number of valid c samples(n_good_c)
  • skewness of variable c(skew_c)
  • latent heat flux(lv_e)
  • mean co2 concentration (c)(mean_c)
  • CO2 flux(fc)
  • rotated covariance vq(cvar_rot_vq)
  • covariance wc(cvar_wc)
  • number of bad or out of range q samples(n_bad_q)
  • kurtosis of variable q(kurt_q)
  • covariance vq(cvar_vq)


Back To Table of Contents

DQRID : D051216.2
Start DateStart TimeEnd DateEnd Time
11/04/2005133011/17/20051630
12/01/2005193012/13/20051630
01/24/2006150002/13/20061930
05/05/2006210006/13/20061400
07/07/2006193007/14/20061500
Subject:
SGP/ECOR/E14 - Flux Measurements Incorrect
DataStreams:sgp30ecorE14.b1
Description:
The ECOR datastream was corrupted; this happens when the ECOR program produces a 
datastream of the wrong length (i.e., leaves one or more measurements out).

Fluxes and some other measurements are incorrect.

Reprocessing will not correct this problem; the raw data are corrupted.
Measurements:sgp30ecorE14.b1:
  • mean co2 concentration (c)(mean_c)
  • CO2 flux(fc)
  • mean water vapor concentration (q)(mean_q)
  • average temperature (IGRA internal sensor)(temp_irga)
  • average atmospheric pressure (IGRA internal sensor)(atm_pres)
  • corrected sensible heat flux(h)
  • latent heat flux(lv_e)


Back To Table of Contents



END OF DATA