Data Quality Reports for Session: 107463 User: pching Completed: 09/01/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050823.1PYE/SONDE/M1 - Bad temperature and RHpyesondewnpnM1.b1


DQRID : D050823.1
Start DateStart TimeEnd DateEnd Time
06/30/2005052806/30/20050711
Subject:
PYE/SONDE/M1 - Bad temperature and RH
DataStreams:pyesondewnpnM1.b1
Description:
The temperature appears to be approximately 10 degC too high during the entire sounding.  
This also causes the RH to be too low.  Most likely the problem was a bad temperature 
sensor.
Measurements:pyesondewnpnM1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents