Data Quality Reports for Session: 107543 User: bgkim Completed: 09/06/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D000801.4SGP/AOS/C1 - Nephelometer failuresgpaosC1.a0, sgpaosauxC1.a0
D000801.5SGP/AOS/C1 - Nephelometer failuresgpaosC1.a0, sgpaosauxC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0
D990611.1SGP/AOS/C1 - Ozone Instrument FailuresgpaosC1.a0, sgpaosauxC1.a0
D990614.1SGP/AOS/C1 - Invalid absorption datasgpaosC1.a0, sgpaosauxC1.a0
D990618.1SGP/AOS/C1 - Incorrect absorption and scattering datasgpaosC1.a0


DQRID : D000801.4
Start DateStart TimeEnd DateEnd Time
06/20/2000134007/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the reference nephelometer, which was out of service for 2 periods 
due to 2 separate problems.  This DQR is for the second period, from 000620 (1340 GMT) to 
000703 (1833 GMT).  Light absorption measurements are invalid whenever the reference 
nephelometer is not operational, because those data are required for correction of the 
absorption data.
Measurements:sgpaosauxC1.a0:
  • TSI Low RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_LRH)
  • TSI Low RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_LRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_LRH)
  • TSI Low RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_LRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_LRH)
  • TSI Low RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_LRH)
  • TSI Low RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_LRH)
  • TSI Low RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_LRH)
  • TSI Low RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_LRH)
  • TSI Low RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_LRH)
  • TSI Low RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_LRH)
  • TSI Low RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_LRH)
  • CLAP Signal(BapSignal)
  • CLAP reference signal(BapRef)
  • TSI Low RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_LRH)

sgpaosC1.a0:
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • 1 um Absorption coefficient(Bap_I_1um)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)


Back To Table of Contents

DQRID : D000801.5
Start DateStart TimeEnd DateEnd Time
06/02/2000184207/03/20001833
Subject:
SGP/AOS/C1 - Nephelometer failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
An inexperienced AOS technician inadvertently closed off the flow to the dual-nephelometer 
humidograph system during a Friday Preventative Maintenance.  This resulted in liquid 
water getting into the humidified nephelometer.  Flow was restored to the system on Monday, 
but the nephelometers were producing strange data.  A number of tests were performed 
over the next few days to remotely troubleshoot the situation, to no avail.  Eventually, the 
decision was made to pull both nephelometers and return them to Boulder for repair.  
This DQR refers only to the humidified nephelometer, which was out of service for the entire 
period from 000602 through 000703.
Measurements:sgpaosauxC1.a0:
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)
  • CLAP Signal(BapSignal)
  • CLAP reference signal(BapRef)

sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • 1 um Absorption coefficient(Bap_I_1um)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents

DQRID : D990611.1
Start DateStart TimeEnd DateEnd Time
05/07/1999155605/14/19991445
Subject:
SGP/AOS/C1 - Ozone Instrument Failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
A malfunction occurred with the AOS surface ozone instrument during the period 990507 1556 
GMT to 990514 1445 GMT.  The instrument reported an ozone concentration of 9.91e-01 ppm 
(991 ppb) continuously during this time.  Quality checking of the data the following week 
revealed that the problem began at about the time weekly maintenance check was made on 
Friday.  The problem continued until the next weekly maintenance check was done.  Site Ops 
was notified by the mentor of the problem on Friday, 5/14/99, and a technician was sent 
to inspect the system.  No switches on the instrument front panel were found in the wrong 
position (for example, if the instrument was inadvertantly left in a calibration mode).  
By the time the technician arrived at the trailer, the instrument appeared to be working 
correctly.  No explanation is given as to why these erroneous values were being 
generated by the ozone instrument, except that possibly it locked up with the throwing of the 
weekly preventative maintenance switch and this problem was not corrected until the same 
switch was thrown the following week.
Measurements:sgpaosauxC1.a0:
  • Ozone sample frequency(O3_Samp)
  • Ozone offset value(O3_Offset)
  • Ozone sample pressure(O3_Press)
  • Ozone span value(O3_Span)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Ozone sample temperature(O3_Temp)
  • Ozone line temperature set point(O3_SetPt)
  • Ozone span concentration(O3_Spn)
  • Ozone zero concentration(O3_Zro)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • unknown(O3_Cont)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • Ozone concentration(Ozone)


Back To Table of Contents

DQRID : D990614.1
Start DateStart TimeEnd DateEnd Time
05/21/1999141305/27/19991819
Subject:
SGP/AOS/C1 - Invalid absorption data
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
During the period noted, a valve position error with the 1-um/10-um impactor switching 
valve occurred.  The result of this error was to invalidate all switched data.  The software 
correctly inserted missing value codes in the nephelometer data, but the PSAP light 
absorption photometer for some reason reported a constant value over this period.
Measurements:sgpaosauxC1.a0:
  • Transmittance through CLAP filter(BapFiltr)
  • CLAP Signal(BapSignal)
  • CLAP reference signal(BapRef)
  • CLAP flow(BapFlow)

sgpaosC1.a0:
  • 10 um Absorption coefficient(Bap_I_10um)
  • 1 um Absorption coefficient(Bap_I_1um)


Back To Table of Contents

DQRID : D990618.1
Start DateStart TimeEnd DateEnd Time
05/21/1999134805/27/19991822
Subject:
SGP/AOS/C1 - Incorrect absorption and scattering data
DataStreams:sgpaosC1.a0
Description:
The electronically switched valve that alternately switches flow through 1-um and 10-um 
impactors showed an error during this time period.  This error invalidated all downstream 
data.  In the raw data files, aerosol light scattering coefficients for the TSI reference 
and humidified nephelometers were assigned missing value codes.  The aerosol light 
absorption coefficient was observed to stick at a constant value over this period.  The problem 
went away after the weekly maintenance was performed, although nothing specific was done 
to the valve.  The flipping of the maintenance switch appears to have solved the problem.
Measurements:sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_LRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • 1 um Absorption coefficient(Bap_I_1um)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • 10 um Absorption coefficient(Bap_I_10um)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_LRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI Low RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_LRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_LRH)
  • TSI Low RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_LRH)
  • TSI Low RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_LRH)
  • TSI Low RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_LRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents



END OF DATA