DQR ID | Subject | Data Streams Affected |
---|---|---|
D060308.1 | TWP/SONDE/C2 - Broken temperature sensor | twpsondewnpnC2.b1 |
D060403.2 | TWP/SONDE/C2 - Broken temperature sensor | twpsondewnpnC2.b1 |
D060418.1 | TWP/SONDE/C2 - Bad RH in lower 8km | twpsondewnpnC2.b1 |
D060522.1 | TWP/SMET/C1 - Upper wind speed sensor failure | twpsmet60sC1.b1 |
D060614.2 | TWP/SMET/C2 - Temp/RH probe failure | twpsmet60sC2.b1 |
D060817.3 | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC1.b1 |
D060818.2 | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC2.b1 |
D061002.1 | TWP/MWR/C2 - Missing data | twp5mwravgC2.c1, twpmwrC2.00, twpmwrlosC2.b1, twpmwrtipC2.a1 |
D061011.2 | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC1.b1 |
D061011.3 | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC2.b1 |
D061114.2 | TWP/MWR/C1 - Radiometer failure | twpmwrC1.00, twpmwrlosC1.b1, twpmwrtipC1.a1 |
D061213.3 | TWP/MWR/C2 - Reprocessed: New Calibration coefficients | twpmwrlosC2.b1, twpmwrtipC2.a1 |
D070105.1 | TWP/MWR/C1 - Missing data | twpmwrlosC1.b1 |
D070308.3 | TWP/SONDE/C2 - Surface Temperature Incorrect | twpsondewnpnC2.b1 |
D070308.4 | TWP/SONDE/C2 - Surface RH value incorrect | twpsondewnpnC2.b1 |
D070313.2 | TWP/SONDE/C2 - Surface RH value incorrect | twpsondewnpnC2.b1 |
D070313.3 | TWP/SONDE/C1 - Surface RH value incorrect | twpsondewnpnC1.b1 |
D070329.2 | TWP/SONDE/C2 - Surface RH Value incorrect | twpsondewnpnC2.b1 |
D070404.1 | TWP/SONDE/C2 - Surface RH too high | twpsondewnpnC2.b1 |
D070412.3 | TWP/MWR/C1/C2 - Sun in the field of view | twpmwrlosC1.b1, twpmwrlosC2.b1, twpmwrtipC1.a1, twpmwrtipC2.a1 |
D070428.1 | TWP/SONDE/C2 - Broken temperature sensor at launch | twpsondewnpnC2.b1 |
D070501.2 | TWP/SONDE/C2 - Broken temperature sensor at launch | twpsondewnpnC2.b1 |
D070613.1 | TWP/MWR/C1 - Intermittent data | twpmwrlosC1.b1, twpmwrtipC1.a1 |
D070824.1 | twpsondewnpnC1.b1 | |
D070824.3 | twpsondewnpnC2.b1 | |
D070824.4 | twpsondewnpnC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/15/2006 | 1144 | 01/15/2006 | 1328 |
01/19/2006 | 1318 | 01/19/2006 | 1320 |
01/22/2006 | 1149 | 01/22/2006 | 1315 |
02/14/2006 | 0055 | 02/14/2006 | 0207 |
02/21/2006 | 0232 | 02/21/2006 | 0359 |
Subject: | TWP/SONDE/C2 - Broken temperature sensor |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The temperature sensor on this radiosonde broke after launch. All the temperature-related data are incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2006 | 0043 | 03/27/2006 | 0208 |
Subject: | TWP/SONDE/C2 - Broken temperature sensor |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The temperature sensor on this radiosonde broke after launch. All the temperature-related data are incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2006 | 0247 | 04/10/2006 | 0314 |
Subject: | TWP/SONDE/C2 - Bad RH in lower 8km |
DataStreams: | twpsondewnpnC2.b1 |
Description: | It appears that the RH sensor heating circuit failed during this sounding. The RH oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems to stop when the temperature is below ~35 degC. According to Vaisala these are symptoms of failure in the ASIC (application specific integrated circuit) that controls the heating. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2006 | 0000 | 10/14/2006 | 0237 |
Subject: | TWP/SMET/C1 - Upper wind speed sensor failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind speed sensor was binding at low wind speeds. Comparison to the lower wind sensor showed that the upper sensor flat-lines for long periods of time. The upper anemometer was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/05/2006 | 2144 | 06/15/2006 | 0427 |
Subject: | TWP/SMET/C2 - Temp/RH probe failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Temperature/RH probe failed and was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 07/13/2006 | 2225 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 07/13/2006 | 2232 |
Subject: | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC2.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2006 | 1500 | 11/27/2006 | 1900 |
Subject: | TWP/MWR/C2 - Missing data |
DataStreams: | twp5mwravgC2.c1, twpmwrC2.00, twpmwrlosC2.b1, twpmwrtipC2.a1 |
Description: | The heater blower assembly was causing the fuse to blow. The instrument was taken offline until a replacement arrives. |
Measurements: | twpmwrlosC2.b1:
twpmwrC2.00:
twp5mwravgC2.c1:
twpmwrtipC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 10/17/2006 | 0000 |
Subject: | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC1.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C1 SMET suite of instruments on 20061017. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC1.b1 datastream. Between 9/26 and 10/17, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19961009-20060926 and data from 19961009-20061017 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 11/29/2006 | 2014 |
Subject: | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC2.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C2 SMET suite of instruments on 20061129. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC2.b1 datastream. Between 9/26 and 11/29, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19981021-20060926 and data from 19981021-20061129 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2006 | 1000 | 01/30/2007 | 0457 |
Subject: | TWP/MWR/C1 - Radiometer failure |
DataStreams: | twpmwrC1.00, twpmwrlosC1.b1, twpmwrtipC1.a1 |
Description: | The instrument stopped responding and had to be powered off. Suspect cause of failure is a faulty digital board. |
Measurements: | twpmwrC1.00:
twpmwrlosC1.b1:
twpmwrtipC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/27/2006 | 1900 | 11/30/2006 | 0700 |
Subject: | TWP/MWR/C2 - Reprocessed: New Calibration coefficients |
DataStreams: | twpmwrlosC2.b1, twpmwrtipC2.a1 |
Description: | On 11/27 a spare radiometer was installed to replace a faulty radiometer. Once the radiometer started to collect data it used the default calibration coefficients until it collected enough tip data to determine its own coefficients. The data were reprocessed to apply corrected calibrations and the reprocessed data were archived 20061222. |
Measurements: | twpmwrlosC2.b1:
twpmwrtipC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1996 | 2338 | 12/11/1996 | 0003 |
02/28/1997 | 0118 | 06/27/1997 | 2048 |
11/06/1997 | 1959 | 11/12/1997 | 1800 |
01/05/1998 | 0204 | 01/08/1998 | 2048 |
04/21/1998 | 0017 | 04/29/1998 | 1803 |
09/13/1998 | 0759 | 09/20/1998 | 0900 |
11/20/1998 | 2226 | 11/27/1998 | 1026 |
02/27/1999 | 0600 | 03/02/1999 | 0159 |
03/06/1999 | 0800 | 03/09/1999 | 1807 |
09/03/1999 | 0300 | 09/10/1999 | 0400 |
10/01/1999 | 0900 | 10/18/1999 | 0000 |
10/20/1999 | 0500 | 10/22/1999 | 0600 |
01/14/2000 | 0800 | 01/16/2000 | 0342 |
01/14/2006 | 1258 | 01/16/2006 | 0638 |
Subject: | TWP/MWR/C1 - Missing data |
DataStreams: | twpmwrlosC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | twpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/02/2007 | 1128 | 01/02/2007 | 1129 |
Subject: | TWP/SONDE/C2 - Surface Temperature Incorrect |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The first two temperature values in the datafile are incorrect. It appears that the observer mis-typed the surface temperature as -27 instead of +27. The next value was 8 which is also incorrect and was caused by the system interpolating from the incorrect surface value and what the sonde temperature sensor was reading. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/23/2007 | 1130 | 01/23/2007 | 1131 |
Subject: | TWP/SONDE/C2 - Surface RH value incorrect |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The surface RH value was incorrect. It appears the observer mis-typed 70 instead of 80. This error also caused the dew point temperature to be incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2007 | 2350 | 03/08/2007 | 2353 |
Subject: | TWP/SONDE/C2 - Surface RH value incorrect |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The surface RH was reproted as 78%. The value should have been in the low 50's. It appears the observer did not allow the sonde to equilibrate to ambient conditions before taking the reading. This error also caused an incorrect dew point. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/08/2007 | 0000 | 03/08/2007 | 0001 |
Subject: | TWP/SONDE/C1 - Surface RH value incorrect |
DataStreams: | twpsondewnpnC1.b1 |
Description: | The surface RH value of 77% when the value should have been in the low 60's. It appears the observers did not allow the sonde to equilibrate to ambient conditions before collecting the surface data. This error also caused the dew point to be incorrect. |
Measurements: | twpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2007 | 2319 | 03/28/2007 | 2320 |
Subject: | TWP/SONDE/C2 - Surface RH Value incorrect |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The surface value entered by the observers for RH (96%) is incorrect. At the ascent time listed in the .ptu file the readings by the Sonde were around 65%. This low RH value is supported by the nearby SMET tower RH reading of 71.8%. This also caused the surface dew point to be incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/03/2007 | 2330 | 04/03/2007 | 2331 |
Subject: | TWP/SONDE/C2 - Surface RH too high |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The surface Rh was entered as 94%. The ascent time has 72% and this value is supported by RH values from the nearby SMET system. This error also will affect the dew point. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2007 | 0000 | 04/02/2007 | 2359 |
Subject: | TWP/MWR/C1/C2 - Sun in the field of view |
DataStreams: | twpmwrlosC1.b1, twpmwrlosC2.b1, twpmwrtipC1.a1, twpmwrtipC2.a1 |
Description: | Every day between 3/7 and 4/2 around noon local time there is an increase in the brightness temperatures due to the sun in the field of vie of the radiometer. |
Measurements: | twpmwrlosC2.b1:
twpmwrlosC1.b1:
twpmwrtipC1.a1:
twpmwrtipC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/27/2007 | 1140 | 04/27/2007 | 1420 |
Subject: | TWP/SONDE/C2 - Broken temperature sensor at launch |
DataStreams: | twpsondewnpnC2.b1 |
Description: | Temperature sensor broken at launch. All thermodynamic values are incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/30/2007 | 2359 | 05/01/2007 | 0200 |
Subject: | TWP/SONDE/C2 - Broken temperature sensor at launch |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The temperature sensor on this radiosonde broke after launch. All the temperature-related data are incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2007 | 0700 | 06/13/2007 | 2000 |
06/16/2007 | 2100 | 06/17/2007 | 2100 |
Subject: | TWP/MWR/C1 - Intermittent data |
DataStreams: | twpmwrlosC1.b1, twpmwrtipC1.a1 |
Description: | After a computer change the MWR program did not restart properly. Data are intermittent between 6/12 and 6/17. |
Measurements: | twpmwrlosC1.b1:
twpmwrtipC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/23/2007 | 1208 | 08/23/2007 | 1209 |
Subject: | |
DataStreams: | twpsondewnpnC1.b1 |
Description: | |
Measurements: |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/23/2007 | 0128 | 08/23/2007 | 0129 |
Subject: | |
DataStreams: | twpsondewnpnC2.b1 |
Description: | |
Measurements: |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/18/2007 | 2258 | 08/24/2007 | 1900 |
Subject: | |
DataStreams: | twpsondewnpnC2.b1 |
Description: | |
Measurements: |