Data Quality Reports for Session: 107582 User: sivaraman Completed: 09/06/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D020909.1SGP/SONDE/C1 - Bad Temperature SensorsgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D020909.2SGP/SONDE/C1 - Bad Temperature SensorsgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D021006.1Bad temperature sensorsgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D030411.3SGP/Sonde/C1 - Failed Sonde Launches due to Antenna problemssgplssondeC1.c1, sgpsondeC1.00, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D050912.2SGP/SONDE/C1 - Temperature/RH incorrectsgplssondeC1.c1, sgpsondewnpnC1.b1


DQRID : D020909.1
Start DateStart TimeEnd DateEnd Time
09/06/2002172709/06/20021854
Subject:
SGP/SONDE/C1 - Bad Temperature Sensor
DataStreams:sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
It appears that the temperature sensor broke upon launch and therefore the reported 
temperature is approximately 40 degC too low.
Measurements:sgpsondewnpnC1.a1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)

sgpqmemwrcolC1.c1:
  • Integrated vapor column from sonde using MWR Instrument Performance Model (IPM)(integ_vap_sonde)

sgplssondeC1.c1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D020909.2
Start DateStart TimeEnd DateEnd Time
09/06/2002052709/06/20020718
Subject:
SGP/SONDE/C1 - Bad Temperature Sensor
DataStreams:sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
It appears that the temperature sensor broke shortly after launch. Temperatures are offset 
by approxmiately (but not constantly) 40 degC.
Measurements:sgpsondewnpnC1.a1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)

sgpqmemwrcolC1.c1:
  • Integrated vapor column from sonde using MWR Instrument Performance Model (IPM)(integ_vap_sonde)

sgplssondeC1.c1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D021006.1
Start DateStart TimeEnd DateEnd Time
10/03/2002112010/03/20021320
Subject:
Bad temperature sensor
DataStreams:sgplssondeC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
It appears that the temperture sensor broke a few seconds into flight.  This results in an 
unreasonably low temperature for the rest of the flight.  For some reason, the rh values 
also go missing shortly after launch.  Only the wind and pressure data are reliable.
Measurements:sgpsondewnpnC1.a1:
  • Dummy altitude for Zeb(alt)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Ascent Rate(asc)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Ascent Rate(asc)
  • Dry bulb temperature(tdry)
  • Dummy altitude for Zeb(alt)

sgplssondeC1.c1:
  • Dummy altitude for Zeb(alt)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D030411.3
Start DateStart TimeEnd DateEnd Time
04/01/2003173004/01/20031930
04/02/2003053004/02/20030730
04/02/2003113004/02/20031330
Subject:
SGP/Sonde/C1 - Failed Sonde Launches due to Antenna problems
DataStreams:sgplssondeC1.c1, sgpsondeC1.00, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
These sonde launches where unsuccessful due to problems
with the antenna configuration resulting in signal loss
before the launch.
Measurements:sgpsondewnpnC1.a1:
  • Mean Wind Speed(wspd)
  • V-component(v_wind)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Relative humidity inside the instrument enclosure(rh)
  • Ascent Rate(asc)
  • U-component(u_wind)
  • Wind Status(wstat)
  • Wind Direction(deg)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Time offset of tweaks from base_time(time_offset)
  • lat(lat)
  • lon(lon)
  • Retrieved pressure profile(pres)

sgpsondewnpnC1.b1:
  • U-component(u_wind)
  • Mean Wind Speed(wspd)
  • lat(lat)
  • Ascent Rate(asc)
  • lon(lon)
  • Time offset of tweaks from base_time(time_offset)
  • V-component(v_wind)
  • Dummy altitude for Zeb(alt)
  • Surface dew point temperature(dp)
  • base time(base_time)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)
  • Wind Direction(deg)
  • Wind Status(wstat)
  • Dry bulb temperature(tdry)

sgpsondeC1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)

sgplssondeC1.c1:
  • lon(lon)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Retrieved pressure profile(pres)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)
  • lat(lat)
  • Time offset of tweaks from base_time(time_offset)


Back To Table of Contents

DQRID : D050912.2
Start DateStart TimeEnd DateEnd Time
09/09/2005113209/09/20051332
Subject:
SGP/SONDE/C1 - Temperature/RH incorrect
DataStreams:sgplssondeC1.c1, sgpsondewnpnC1.b1
Description:
The temperature throughout this sounding is too high and as a result the RH values will be 
incorrect as well.  I'm not sure what is wrong here but the operators noted reasonable 
values (18.1 degC, 87%) in the log before launch.  Review of the raw data, however, shows 
the sonde reporting temperatures nearly 20 degC higher.  I suppose it is possible that 
the operator entered into the log the reference temperature from the GC25 ground check set 
rather than the temperature reported by the sonde - I can't think of any other 
explanation - there is no doubt that the temperature sensor on this sonde was bad.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)

sgplssondeC1.c1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents



END OF DATA