Data Quality Reports for Session: 107912 User: chubarova Completed: 09/24/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060630.19SGP/SIRS/E11 - Reprocessed: Longwave Calibration errorsgpsirsE11.b1


DQRID : D060630.19
Start DateStart TimeEnd DateEnd Time
02/17/2004175302/14/20062016
Subject:
SGP/SIRS/E11 - Reprocessed: Longwave Calibration error
DataStreams:sgpsirsE11.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in February 2004. 
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.									      
The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records. 
Still, these data are considered far superior to the originally processed data. The 
reprocessed data were archived in August 2006.
Measurements:sgpsirsE11.b1:
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)


Back To Table of Contents



END OF DATA