DQR ID | Subject | Data Streams Affected |
---|
D060717.2 | SGP/MWR/C1 - Spikes in ambient temperature readings | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D060821.3 | SGP/MWR/C1 - Missing data | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D061213.4 | SGP/MWR/C1 - Instrument testing | sgpmwrC1.00, sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D070314.2 | SGP/MWR/C1/E14 - Freezing rain-Incorrect rain flag | sgpmwrlosC1.b1, sgpmwrlosE14.b1, sgpmwrtipC1.a1, sgpmwrtipE14.a1 |
D070802.3 | SGP/MWR/C1 - Noisy data | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Subject: | SGP/MWR/C1/E14 - Freezing rain-Incorrect rain flag |
DataStreams: | sgpmwrlosC1.b1, sgpmwrlosE14.b1, sgpmwrtipC1.a1, sgpmwrtipE14.a1
|
Description: | Between 01/12 and 01/15 brightness temperatures show high values (Tb > 100 K) indicative
of rain, however the wet_window flag is 0 (indicative of no rain). During that time there
where freezing conditions with ice pellets that were not detected by the sensor. The high
brightness temperatures may be due to melting ice on the window. |
Measurements: | sgpmwrtipC1.a1: - Water on Teflon window (1=WET, 0=DRY)(wet_window)
sgpmwrlosC1.b1: - Water on Teflon window (1=WET, 0=DRY)(wet_window)
sgpmwrtipE14.a1: - Water on Teflon window (1=WET, 0=DRY)(wet_window)
sgpmwrlosE14.b1: - Water on Teflon window (1=WET, 0=DRY)(wet_window)
|