Data Quality Reports for Session: 108508 User: bart Completed: 10/19/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060112.4NIM/SKYRAD/M1 - IRTs do not agree with AERInimskyrad60sM1.b1
D060222.1NIM/SKYRAD/M1 - Instrument ventilator problemnimskyrad20sM1.a0, nimskyrad60sM1.b1, nimskyradM1.00
D060915.1NIM/SKYRAD/M1 - Unknown problem with PIR1nimskyrad20sM1.a0, nimskyrad60sM1.b1, nimskyradM1.00
D061006.1NIM/SKYRAD/M1 - Tracker inoperativenimskyrad20sM1.a0, nimskyrad60sM1.b1, nimskyradM1.00


DQRID : D060112.4
Start DateStart TimeEnd DateEnd Time
11/26/2005000006/07/20061900
Subject:
NIM/SKYRAD/M1 - IRTs do not agree with AERI
DataStreams:nimskyrad60sM1.b1
Description:
Since deployment at PYE, and then at NIM, the AMF SKYRAD IRT measured about 10K higher sky 
temperatures than the AERI and the MWRP IRT measureed about 20K higher than the AERI.  
Several actions were taken to diagnose the problem including confirming the correct 
configuration of the IRTs and data logger, cleaning the mirror and lens, and replacing the mirror.

After several days of rain beginning 6/2/2006, the three instruments came into agreement. 
It is unknown whether this was a problem with the AERI, MWRP-IRT or SKYRAD-IRT.
Measurements:nimskyrad60sM1.b1:
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Sky Infra Red Temperature Minima(sky_ir_temp_min)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)


Back To Table of Contents

DQRID : D060222.1
Start DateStart TimeEnd DateEnd Time
12/23/2005000002/23/20060800
Subject:
NIM/SKYRAD/M1 - Instrument ventilator problem
DataStreams:nimskyrad20sM1.a0, nimskyrad60sM1.b1, nimskyradM1.00
Description:
Data quality office first reported that PIR1 and PIR2 appear to be deviating more than 
typical during the daylight hours.  It was found that the Skyrad PIR ventilator fans were 
building up debris on the intake air filter.  Obstructed air flow results in solar heating 
of the PIR ventilator housing in the later part of the day and biasing the incomming IR 
irradiance values. The slow buildup was insidious and wasn't discovered and corrected 
(i.e. ventilator fans cleaned on a regular and frequent basis) until Feb 23rd. On or around 
December 23rd was when this problem may have first started.  We are unable to quantify the 
extent of the problem in W/m2.  Normal deviation between PIRs is typically 3-5 W/m2.
Measurements:nimskyrad20sM1.a0:
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)

nimskyrad60sM1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Temperature, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_temp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Temperature, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_temp)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Temperature, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_temp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Temperature, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_temp)

nimskyradM1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060915.1
Start DateStart TimeEnd DateEnd Time
09/02/2006000009/19/20060745
Subject:
NIM/SKYRAD/M1 - Unknown problem with PIR1
DataStreams:nimskyrad20sM1.a0, nimskyrad60sM1.b1, nimskyradM1.00
Description:
Spikes in Shaded 1 occur frequently during this time period, resulting in differences b/t 
shaded1 and shaded2 exceeding 20W/m2.  Instrument connections were checked and a spare 
PIR instrument was tried in the place of PIR1 (shaded1).  After returning the original PIR1 
to its original location data again looked good and very close to PIR2.  Problem appears 
resolved but exact cure is unknown.
Measurements:nimskyrad20sM1.a0:
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_resist)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_resist)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Resistance, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_resist)

nimskyrad60sM1.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp_shaded1)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Maxima(down_long_hemisp_shaded2_max)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Minima(down_long_hemisp_shaded1_min)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer2(inst_down_long_hemisp_shaded2_tp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Standard
    Deviation(down_long_hemisp_shaded2_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Standard
    Deviation(down_long_hemisp_shaded1_std)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1, Maxima(down_long_hemisp_shaded1_max)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Temperature, Shaded
    Pyrgeometer2(inst_down_long_shaded2_dome_temp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2, Minima(down_long_hemisp_shaded2_min)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer2(down_long_hemisp_shaded2)
  • Instantaneous Downwelling Pyrgeometer Dome Thermistor Temperature, Shaded
    Pyrgeometer1(inst_down_long_shaded1_dome_temp)
  • Instantaneous Downwelling Pyrgeometer Thermopile, Shaded Pyrgeometer1(inst_down_long_hemisp_shaded1_tp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Temperature, Shaded
    Pyrgeometer2(inst_down_long_shaded2_case_temp)
  • Instantaneous Downwelling Pyrgeometer Case Thermistor Temperature, Shaded
    Pyrgeometer1(inst_down_long_shaded1_case_temp)

nimskyradM1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D061006.1
Start DateStart TimeEnd DateEnd Time
09/12/2006120010/05/20061330
Subject:
NIM/SKYRAD/M1 - Tracker inoperative
DataStreams:nimskyrad20sM1.a0, nimskyrad60sM1.b1, nimskyradM1.00
Description:
Starting on 9/12 the NIM radiometer tracker drifted enough due to sinking supports and a 
drifting clock that the direct normal measurments (NIP) and the downwelling hemispheric 
diffuse shortwave measurments (8-48) were adversely affected in the afternoon hours.  The 
shortwave hemispheric global instrument (PSP) and downelling hemispheric longwave shaded1 
and shaded2 (PIR) are unaffected by this problem. 

The tracker problem greatly affected direct measurements and affected the diffuse 
measurement to a lesser extent. There are no redundant measurements for these instruments.  
However, during this time range the data collected before local noon (aprox 1140GMT) for all 
shortwave Skyrad instruments appear to be good.
Measurements:nimskyrad20sM1.a0:
  • Instantaneous Direct Normal Shortwave Irradiance, Pyheliometer Thermopile
    Voltage(inst_direct_normal)
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)

nimskyrad60sM1.b1:
  • Shortwave Direct Normal Irradiance, Pyrgeometer(short_direct_normal)
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Standard Deviation(down_short_diffuse_hemisp_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Minima(short_direct_normal_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)

nimskyradM1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents



END OF DATA