Data Quality Reports for Session: 108720 User: wlj Completed: 11/02/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D061213.1NSA/METTWR/C1 - CMH Sensor Calibrated Incorrectlynsamettwr4hC1.b1
D061213.2NSA/METTWR/C1 - 40m wind sensor heater failurensamettwr4hC1.b1
D070226.1NSA/METTWR/C1 - 40m Wind Direction Data Suspectnsamettwr4hC1.b1
D070529.1NSA/METTWR/C1 - 20m RH Data Questionablensamettwr4hC1.b1


DQRID : D061213.1
Start DateStart TimeEnd DateEnd Time
12/12/2006150003/14/20070234
Subject:
NSA/METTWR/C1 - CMH Sensor Calibrated Incorrectly
DataStreams:nsamettwr4hC1.b1
Description:
The CMH temp and DP were higher than the co-located T/RH probe at the 2m level.  An 
incorrect calibration along with a failing thermistor caused the readings to be incorrect for 
both measured values Temp and Dew Point.
Measurements:nsamettwr4hC1.b1:
  • Chilled Mirror Calculated Vapor Pressure(VPCMH)
  • Chilled Mirror Calculated Relative Humidity(CMHRH)
  • Chilled Mirror Dew Point(CMHDP)
  • Chilled Mirror Temperature(CMHTemp)


Back To Table of Contents

DQRID : D061213.2
Start DateStart TimeEnd DateEnd Time
11/01/2006000005/22/20072347
Subject:
NSA/METTWR/C1 - 40m wind sensor heater failure
DataStreams:nsamettwr4hC1.b1
Description:
The 40m wind direction became frozen at times.  The values from the sensor were flat-lined 
(stdev = 0) much of the time and did not match sensors from other levels.
Measurements:nsamettwr4hC1.b1:
  • 40m Vector Averaged Wind Direction(WD40M_DU_WVT)
  • Standard Deviation of 40m Vector Averaged Wind Direction(WD40M_SDU_WVT)


Back To Table of Contents

DQRID : D070226.1
Start DateStart TimeEnd DateEnd Time
11/20/2006003011/27/20060254
12/08/2006012512/14/20062229
12/16/2006154512/21/20061105
12/23/2006161201/18/20072357
01/20/2007135701/29/20072140
Subject:
NSA/METTWR/C1 - 40m Wind Direction Data Suspect
DataStreams:nsamettwr4hC1.b1
Description:
The 40m wind direction data was suspect due to icing on the vane or a failing shafter 
heater or a combination of both.  The affected data can be easily seen on comparison plots 
with other levels data or by checking the standard deviation of the 40m wind direction.  
Any standard deviation of less than 0.10 degrees for the 40m level suggest a sluggish/ice 
affected sensor.
Measurements:nsamettwr4hC1.b1:
  • 40m Vector Averaged Wind Direction(WD40M_DU_WVT)
  • Standard Deviation of 40m Vector Averaged Wind Direction(WD40M_SDU_WVT)


Back To Table of Contents

DQRID : D070529.1
Start DateStart TimeEnd DateEnd Time
10/18/2006000005/22/20072330
Subject:
NSA/METTWR/C1 - 20m RH Data Questionable
DataStreams:nsamettwr4hC1.b1
Description:
The 20m RH data was questionable at times.  When the data appeared to be poor was when the 
2m and 20m agreed better than the 20m and the 40m and/or 10m.
Measurements:nsamettwr4hC1.b1:
  • Standard Deviation of 20m Calculated Dew Point(DP20M_STD)
  • Standard Deviation of 20m Calculated Vapor Pressure(VP20M_STD)
  • Standard Deviation of 20m Relative Humidity(RH20M_STD)
  • 20m Average Relative Humidity(RH20M_AVG)
  • 20m Average Calculated Vapor Pressure(VP20M_AVG)
  • 20m Average Calculated Dew Point(DP20M_AVG)


Back To Table of Contents



END OF DATA