Data Quality Reports for Session: 109042 User: mather Completed: 11/26/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040228.1TWP/SONDE/C1 - Broken temperature sensortwpsondewnpnC1.b1
D041011.3TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D050105.1TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D050201.2TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1


DQRID : D040228.1
Start DateStart TimeEnd DateEnd Time
02/07/2004234202/08/20040115
02/09/2004113302/09/20041233
04/05/2004233204/06/20040105
04/22/2004115604/22/20041317
04/24/2004113004/24/20041312
04/29/2004113704/29/20041312
04/30/2004233605/01/20040112
05/03/2004233105/04/20040105
05/05/2004113505/05/20041319
05/07/2004113205/07/20041315
05/11/2004114205/11/20041311
05/12/2004114505/12/20041206
05/18/2004113005/18/20041321
07/10/2004113007/11/20041255
07/13/2004113907/13/20041320
07/16/2004234607/17/20040107
07/22/2004235507/23/20040140
07/26/2004234107/27/20040115
08/14/2004233008/15/20040104
08/22/2004233508/23/20040105
09/05/2004234009/06/20040119
09/14/2004232809/14/20042329
09/17/2004233209/18/20040102
10/28/2004233610/29/20040057
11/28/2004235211/29/20040044
Subject:
TWP/SONDE/C1 - Broken temperature sensor
DataStreams:twpsondewnpnC1.b1
Description:
The temperature sensor on these radiosondes broke after launch
All the temperature-related data are incorrect.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D041011.3
Start DateStart TimeEnd DateEnd Time
10/08/2004113110/08/20041200
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~1 %RH) and 
high (~80 %RH).  According to Vaisala these are symptoms of failure in the ASIC 
(application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D050105.1
Start DateStart TimeEnd DateEnd Time
12/30/2004233012/31/20040108
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~5 %RH) and 
not so low (~70 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D050201.2
Start DateStart TimeEnd DateEnd Time
01/07/2005233001/08/20050102
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 12km).  The RH oscillates rapidly between low (~2 %RH) and 
not so low (~80 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents



END OF DATA