DQR ID | Subject | Data Streams Affected |
---|---|---|
D030114.1 | SGP/SONDE/C1 - Bad RH Data | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D030122.1 | SGP/SONDE/C1 - Broken temperature sensor SGP/CF sounding | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
D030206.5 | SGP/SONDE/C1 - Bad temperature/RH | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2003 | 0536 | 01/11/2003 | 0551 |
Subject: | SGP/SONDE/C1 - Bad RH Data |
DataStreams: | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | The RH values during the first 15 minutes of this sounding oscillate between reasonable values and near-zero values. The oscillation is irregular and seems to be related to failure of the RH sensor heating process. I have informed Vaisala of the problem and provided raw data files to them for analysis. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/19/2003 | 1134 | 01/19/2003 | 1259 |
Subject: | SGP/SONDE/C1 - Broken temperature sensor SGP/CF sounding |
DataStreams: | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | The radiosonde temperature sensor apparently broke during launch. This results in bad temperature, dew point, and rh data. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/02/2003 | 1732 | 02/02/2003 | 1832 |
Subject: | SGP/SONDE/C1 - Bad temperature/RH |
DataStreams: | sgpsondewnpnC1.a1, sgpsondewnpnC1.b1 |
Description: | The temperature sensor broke shortly after launch. All temperature and RH-related data are incorrect. |
Measurements: | sgpsondewnpnC1.a1:
sgpsondewnpnC1.b1:
|