DQR ID | Subject | Data Streams Affected |
---|---|---|
D071001.1 | SGP/EBBR/E26 - Sensible and Latent Heat Fluxes Affected by bad temp sensor | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
D071115.1 | SGP/EBBR/E20 - Reprocess: Wind Direction 65 Degrees Low | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
D080102.2 | SGP/EBBR/E9 - Net Radiation Incorrect | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/28/2007 | 0140 | 10/11/2007 | 1730 |
Subject: | SGP/EBBR/E26 - Sensible and Latent Heat Fluxes Affected by bad temp sensor |
DataStreams: | sgp15ebbrE26.b1, sgp30ebbrE26.b1, sgp5ebbrE26.b1 |
Description: | The left Thum (temperature of the humidity sensor chamber) measurement was incorrect at times, affecting several measurements, including the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE26.b1:
sgp5ebbrE26.b1:
sgp30ebbrE26.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2007 | 0000 | 10/24/2007 | 1900 |
Subject: | SGP/EBBR/E20 - Reprocess: Wind Direction 65 Degrees Low |
DataStreams: | sgp15ebbrE20.b1, sgp30ebbrE20.b1, sgp5ebbrE20.b1 |
Description: | The wind direction sensor reported wind directions that were 65 degrees too low because of a mis-alignment. This bias should have no effect on the res_ws or sigma_wd measurements. The wind direction data could be re-processed by adding 65 degrees. |
Measurements: | sgp30ebbrE20.b1:
sgp15ebbrE20.b1:
sgp5ebbrE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/22/2007 | 1500 | 01/08/2008 | 1800 |
Subject: | SGP/EBBR/E9 - Net Radiation Incorrect |
DataStreams: | sgp15ebbrE9.b1, sgp30ebbrE9.b1, sgp5ebbrE9.b1 |
Description: | There was water in the net radiometer. The net radiation data was incorrect. |
Measurements: | sgp15ebbrE9.b1:
sgp5ebbrE9.b1:
sgp30ebbrE9.b1:
|