DQR ID | Subject | Data Streams Affected |
---|---|---|
D050420.3 | SGP/SIRS/C1 - Reprocess: Adjust Direct Normal Irradiance | sgp15swfanalsirs1longC1.c1, sgp1swfanalsirs1longC1.c1, sgpsirsC1.00, sgpsirsC1.01 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/2000 | 1300 | 08/15/2001 | 1350 |
Subject: | SGP/SIRS/C1 - Reprocess: Adjust Direct Normal Irradiance |
DataStreams: | sgp15swfanalsirs1longC1.c1, sgp1swfanalsirs1longC1.c1, sgpsirsC1.00, sgpsirsC1.01 |
Description: | Direct normal broadband shortwave irradiance measured by NIP s/n 30722E6 was found to have a 0.62%% negative bias when compared in-situ with absolute cavity radiometer measurements taken on August 21, 2000. The estimated measurement uncertainty of the calibration of NIP s/n 30722E6 on July 18, 2000 was -2.6% to +4.8%. Data from NIP s/n 30722E6 can be adjusted according to the in-situ comparisons to reduce the apparent measurement bias. |
Measurements: | sgpsirsC1.01:
sgp1swfanalsirs1longC1.c1:
sgpsirsC1.00:
sgp15swfanalsirs1longC1.c1:
|