DQR ID | Subject | Data Streams Affected |
---|---|---|
D040227.3 | SGP/SIRS/E2 - Reprocess: Incorrect shortwave radiometer calibration coefficients | sgpsirsE2.00, sgpsirsE2.b1 |
D040229.2 | SGP/SIRS/E21 - Shadowband Misalignment | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
D040320.1 | SGP/SIRS/E5 - anomalously low NIP readings | sgpsirs20sE5.a0, sgpsirsE5.00, sgpsirsE5.b1 |
D040320.2 | SGP/SIRS/E8 - Possible Contamination of DS PSP instrument | sgpsirs20sE8.a0, sgpsirsE8.00, sgpsirsE8.b1 |
D040329.1 | SGP/SIRS/E21 - Instrument logger time problem, Unknown | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
D040405.2 | SGP/SIRS/E7 - PSP dome contamination | sgpsirs20sE7.a0, sgpsirsE7.b1 |
D040831.2 | SGP/SIRS/E7 - Reprocess: Incorrect shortwave radiometer calibration coefficients | sgpsirsE7.00, sgpsirsE7.b1 |
D040831.3 | SGP/SIRS/E8 - Reprocess: Incorrect shortwave radiometer calibration coefficients | sgpsirsE8.00, sgpsirsE8.b1 |
D060630.10 | SGP/SIRS/E2 - Reprocessed: Longwave Calibration error | sgpsirsE2.b1 |
D060630.11 | SGP/SIRS/E3 - Reprocessed: Longwave Calibration error | sgpsirsE3.b1 |
D060630.12 | SGP/SIRS/E4 - Reprocessed: Longwave Calibration error | sgpsirsE4.b1 |
D060630.13 | SGP/SIRS/E5 - Reprocessed: Longwave Calibration error | sgpsirsE5.b1 |
D060630.14 | SGP/SIRS/E6 - Reprocessed: Longwave Calibration error | sgpsirsE6.b1 |
D060630.15 | SGP/SIRS/E7 - Reprocessed: Longwave Calibration error | sgpsirsE7.b1 |
D060630.16 | SGP/SIRS/E8 - Reprocessed: Longwave Calibration error | sgpsirsE8.b1 |
D060630.17 | SGP/SIRS/E9 - Reprocessed: Longwave Calibration error | sgpsirsE9.b1 |
D060630.26 | SGP/SIRS/E20 - Reprocessed: Longwave Calibration error | sgpsirsE20.b1 |
D060630.27 | SGP/SIRS/E21 - Reprocessed: Longwave Calibration error | sgpsirsE21.b1 |
D060630.28 | SGP/SIRS/E22 - Reprocessed: Longwave Calibration error | sgpsirsE22.b1 |
D060630.29 | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error | sgpsirsE24.b1 |
D060630.30 | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error | sgpsirsE27.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/2004 | 1612 | 03/25/2004 | 1600 |
Subject: | SGP/SIRS/E2 - Reprocess: Incorrect shortwave radiometer calibration coefficients |
DataStreams: | sgpsirsE2.00, sgpsirsE2.b1 |
Description: | During the pyrgeometer instrument (PIR) changout a datalogger program was loaded with incorrect calibration coefficients for the shortwave instruments. The affected measurements were : downwelling diffuse (DD), upwelling shortwave(US), direct normal (NIP) and downwelling shortwave(DS) Recalculate irradiances by determining original mV signal and reapplying the correct calibration coefficients. For each one minute irradiance value perform the following calculations by instrument: DD(corrected) = (DD/107.24)*118.11 US(corrected) = (US/113.19)*118.21 NIP(corrected)= (NIP/120.17)*119.08 DS(corrected) = (DS/118.46)*117.33 |
Measurements: | sgpsirsE2.b1:
sgpsirsE2.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 1300 | 03/03/2004 | 1300 |
Subject: | SGP/SIRS/E21 - Shadowband Misalignment |
DataStreams: | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
Description: | The NIP reported values close to zero and downwelling shortwave global and diffuse were equivalent due to tracker failure. Maintenance replaced tracker 2/27, but couldn't replace instruments until 03/02 2000 GMT due to bird poop contamination. |
Measurements: | sgpsirsE21.00:
sgpsirs20sE21.a0:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/17/2004 | 1843 | 03/28/2004 | 0300 |
Subject: | SGP/SIRS/E5 - anomalously low NIP readings |
DataStreams: | sgpsirs20sE5.a0, sgpsirsE5.00, sgpsirsE5.b1 |
Description: | Direct normal (NIP instrument) irradiance readings fall anomalously on 3/17 18:43GMT to less than half the expected values during clear sky conditions. NIP readings returned to normal on 3/28/04 12:00. Maintenance on 3/29/04 19:48 revealed no unusual contaminents. Precipitation on 3/27 (measured by the SMOS system) probably cleared the NIP window of contaminents before scheduled maintenance. |
Measurements: | sgpsirsE5.b1:
sgpsirs20sE5.a0:
sgpsirsE5.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/15/2004 | 1500 | 03/31/2004 | 1200 |
Subject: | SGP/SIRS/E8 - Possible Contamination of DS PSP instrument |
DataStreams: | sgpsirs20sE8.a0, sgpsirsE8.00, sgpsirsE8.b1 |
Description: | Downwelling hemispheric shortwave values drop at about 1500 GMT under clear sky conditions, typically from about 14:00GMT to 16:00GMT. This may be due to a dirty PSP but maintenance on 03/23/04 18:11 GMT revealed no more than typical light dust. The problem was essentially undetectable by 03/31/04 12:00. The time of day that the problem exists appears to be an exagerated PSP cosine response. No obvious reason presents why the problem appeared and then disappeared within this two week period. |
Measurements: | sgpsirsE8.00:
sgpsirs20sE8.a0:
sgpsirsE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/2004 | 1200 | 03/24/2004 | 0200 |
Subject: | SGP/SIRS/E21 - Instrument logger time problem, Unknown |
DataStreams: | sgpsirs20sE21.a0, sgpsirsE21.00, sgpsirsE21.b1 |
Description: | Values of diffuse hemispheric shortwave radiation differ from derived values by at most 70 Wm^(-2). The incident appears to be isolated to 3/21, 3/22, and 3/23. An estimation of solar noon using downwelling shortwave hemispheric irradiance from the PSP shows solar noon to oocur at around 1845GMT on 3/21. Calculated solar noon occurs at E21 at 1831GMT for 3/21/2004. There appears to be a time discrepancy in the datalogger for these dates. 04/01 is the next date with enough sun to determine solar noon and compare to calculated. The data from this date appears correct. Ronald Reed [rreed@ops.sgp.arm.gov] reports the logger did not receive any reportable time updates during the late March time period. We don't know what happened during these three dates but from 3/24 on data looks acceptable. |
Measurements: | sgpsirsE21.00:
sgpsirs20sE21.a0:
sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/30/2004 | 1430 | 04/06/2004 | 2100 |
Subject: | SGP/SIRS/E7 - PSP dome contamination |
DataStreams: | sgpsirs20sE7.a0, sgpsirsE7.b1 |
Description: | Values of hemispheric downwelling shortwave radiation differed from derived values by 45 W/m^2 on 03/30 around 1430 GMT under clear sky conditions due to dirt on the dome. Maintenance reported cleaning heavy dust from dome on 04/06 2030 GMT. |
Measurements: | sgpsirs20sE7.a0:
sgpsirsE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 1845 | 04/20/2004 | 1825 |
Subject: | SGP/SIRS/E7 - Reprocess: Incorrect shortwave radiometer calibration coefficients |
DataStreams: | sgpsirsE7.00, sgpsirsE7.b1 |
Description: | During the pyrgeometer instrument (PIR) changout a datalogger program was loaded with incorrect calibration coefficients for the shortwave instruments. The affected measurements (instruments) were : downwelling diffuse (8-48), upwelling shortwave(PSP), direct normal (NIP) and downwelling shortwave (PSP) Recalculate irradiances by determining original mV signal and reapplying the correct calibration coefficients. For each one minute irradiance value perform the following calculations by instrument: DD(corrected) = (DD/107.24)*108.50 US(corrected) = (US/113.19)*110.10 NIP(corrected)= (NIP/120.17)*122.26 DS(corrected) = (DS/118.46)*122.56 |
Measurements: | sgpsirsE7.b1:
sgpsirsE7.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/2004 | 1825 | 04/20/2004 | 1825 |
Subject: | SGP/SIRS/E8 - Reprocess: Incorrect shortwave radiometer calibration coefficients |
DataStreams: | sgpsirsE8.00, sgpsirsE8.b1 |
Description: | During the pyrgeometer instrument (PIR) changeout a datalogger program was loaded with incorrect calibration coefficients for the shortwave instruments. The affected measurements (instruments) were: downwelling diffuse (8-48), upwelling shortwave (PSP), direct normal (NIP) and downwelling shortwave (PSP) Recalculate irradiances by determining original mV signal and reapplying the correct calibration coefficients. For each one minute irradiance value perform the following calculations by instrument: DD(corrected) = (DD/107.24)*116.37 US(corrected) = (US/113.19)*121.78 NIP(corrected)= (NIP/120.17)*122.47 DS(corrected) = (DS/118.46)*122.34 |
Measurements: | sgpsirsE8.00:
sgpsirsE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/2004 | 1630 | 02/23/2005 | 1615 |
Subject: | SGP/SIRS/E2 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in September 2006. |
Measurements: | sgpsirsE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2004 | 1715 | 02/08/2006 | 1730 |
Subject: | SGP/SIRS/E3 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE3.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2004 | 1720 | 02/22/2006 | 1615 |
Subject: | SGP/SIRS/E4 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE4.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE4.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2004 | 2030 | 02/22/2006 | 1817 |
Subject: | SGP/SIRS/E5 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE5.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE5.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/11/2004 | 1610 | 02/09/2006 | 1553 |
Subject: | SGP/SIRS/E6 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE6.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE6.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 2000 | 02/07/2006 | 2005 |
Subject: | SGP/SIRS/E7 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE7.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/2004 | 1845 | 02/21/2006 | 1840 |
Subject: | SGP/SIRS/E8 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE8.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in September 2006. |
Measurements: | sgpsirsE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2003 | 1640 | 02/07/2006 | 1730 |
Subject: | SGP/SIRS/E9 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE9.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 2100 | 02/15/2006 | 2030 |
Subject: | SGP/SIRS/E20 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE20.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in November 2006. |
Measurements: | sgpsirsE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1520 | 02/15/2006 | 1515 |
Subject: | SGP/SIRS/E21 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE21.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in November 2006. |
Measurements: | sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1830 | 02/15/2006 | 1741 |
Subject: | SGP/SIRS/E22 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE22.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1640 | 02/16/2006 | 1705 |
Subject: | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE24.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE24.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2003 | 0002 | 02/15/2006 | 1750 |
Subject: | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE27.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in May 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE27.b1:
|