Subject: | PYE/SKYRAD/M1 - Reprocessed: IRT configuration and calibration |
DataStreams: | pyeskyrad20sM1.a0, pyeskyrad60sM1.b1, pyeskyradM1.00
|
Description: | The configuration of the IRT was inadvertently set to default and the program of the data
loggers were changed to that used at TWP.C3. This caused the calibration of the data to
be wrong. It was corrected by reconfiguring the IRT and changing the conversion factors
in the data logger program.
These data have been reprocessed to apply correct calibrations and the reprocessed data
were archived in February, 2007. |
Measurements: | pyeskyrad60sM1.b1: - Sky Infra Red Temperature Minima(sky_ir_temp_min)
- Sky Infra Red Temperature Maxima(sky_ir_temp_max)
- Standard Deviation of Sky Infra Red Temperature(sky_ir_temp_std)
- Sky Infra-Red Temperature(sky_ir_temp)
pyeskyradM1.00: - null(Raw data stream - documentation not supported)
pyeskyrad20sM1.a0: - Instantaneous Sky/Cloud Infrared(inst_sky_ir_signal)
|