DQR ID | Subject | Data Streams Affected |
---|---|---|
D000302.1 | TWP/SMET/C1 - Reprocess: Skewed wind direction 1 data | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1 |
D000504.5 | TWP/SMET/C1 - Wind Direction alignment method | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1 |
D000608.7 | TWP/C2/SMET Wind Speeds | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
D000608.8 | TWP/C2/SMET Wind Speeds | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
D000608.9 | TWP/C2/SMET Winds Speeds | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
D000613.2 | TWP/C2/SMET Temperature spike | twpsmet60sC2.a1, twpsmet60sC2.b1 |
D000613.3 | TWP/C2/SMET - Pressure sensitivity to rain | twpsmet60sC2.a1, twpsmet60sC2.b1 |
D000621.1 | TWP/SMET/C1 - Incorrect calibration for ARCS1 Wind speed | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmetC1.00 |
D000718.11 | TWP/SMET/C1 - Periodic spikes in temperature and water vapor pressure | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D000718.12 | TWP/SMET/C1 - Elevated relative humidity | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D010709.15 | TWP/SMET/C1 - C1 TWP Wind speed calibration factor problem/ Wind1 | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D010709.16 | TWP/SMET/C1 - C1 SMET Incorrect wind speed calibration factors | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D011211.3 | TWP/SMET/C1 - precip, wind, temperature | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D020109.2 | TWP/SMET/C2 - RH Sensor bad at Nauru | twpsmet60sC2.a1, twpsmet60sC2.b1 |
D020214.9 | TWP/SMET/C1 - Sensor replacement | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D020613.9 | TWP/C2 Installation | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
D020620.61 | TWP/C1 - Incorrect elevation reported in Manus Data | twpgndrad20sC1.a1, twpgndrad60sC1.a1, twpgndrad60sC1.b1, twpsmet60sC1.a1, twpsmet60sC1.b1 |
D021211.1 | TWP/MET/C3 - WIND DIRECTION SENSOR #2 MISALIGNED | twpsmet60sC3.a1, twpsmet60sC3.b1 |
D030131.1 | TWP/SMET/C2 - Temperature Sensor Failed, Data Missing | twpsmet60sC2.a1, twpsmet60sC2.b1 |
D030212.1 | TWP/SMET/C3 - Intermittent Pressure Sensor Failure | twpsmet60sC3.a1, twpsmet60sC3.b1 |
D030227.3 | TWP/SMET/C1 - Reprocess: Miscalibrated windspeed2 data | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D030317.1 | TWP/SMET/C1 - Reprocess: Wind Speed #2 incorrect | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D030501.5 | TWP/SMET/C2 - WIND SENSOR #1 (HIGHER) INOPERATIVE | twpsmet60sC2.b1 |
D031103.2 | TWP/SMET/C3 - Data reported at 15sec rather than 60s intervals | twpsmet60sC3.a1, twpsmet60sC3.b1 |
D031215.3 | TWP/SMET/C1 - Tower Lowered for Cal Activities | twpsmet60sC1.b1, twpsmetC1.00 |
D040421.1 | TWP/SMET/C2 - Wind Sensor failure | twpsmet60sC2.b1 |
D040428.1 | TWP/SMET/C1 - Poor Performing Sensor | twpsmet60sC1.b1 |
D040722.4 | TWP/SMET/C3 - Rain blocked pressure port | twpsmet60sC3.b1 |
D040722.5 | TWP/SMET/C3 - Data missing due to Logger Change | twpsmet60sC3.b1 |
D040722.6 | TWP/SMET/C2 - All data missing due to logger change | twpsmet60sC2.b1 |
D040722.7 | TWP/SMET/C2 - connection and timing issues with new logger | twpsmet60sC2.b1 |
D040726.1 | TWP/SMET/C2 - Negative rainfall rates | twpsmet60sC2.b1 |
D040726.3 | TWP/SMET/C3 - Negative rainfall rates | twpsmet60sC3.b1 |
D040813.11 | TWP/SMET/C1 - RESET14 activities | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D040824.2 | TWP/SMET/C2 - Incorrect rainrate equation used. | twpsmet60sC2.b1 |
D040824.3 | TWP/SMET/C1 - Incorrect rainrate equation used | twpsmet60sC1.b1 |
D050210.1 | TWP/SMET/C3 - Data non-representative | twpsmet60sC3.b1 |
D050301.1 | TWP/SMET/C2 - Instrument failure | twpsmet60sC2.b1 |
D050322.2 | TWP/SMET/C2 - Upper wind sensor failure | twpsmet60sC2.b1 |
D050418.2 | TWP/SMET/C3 - Instrument logger problem, Sensor failure | twpsmet60sC3.b1 |
D050421.6 | TWP/SMET/C1 - Negative rainfall rates | twpsmet60sC1.b1 |
D050714.1 | TWP/SMET/C2 - RH failure | twpsmet60sC2.b1 |
D050808.3 | TWP/SMET/C3 - Data not representative | twpsmet60sC3.b1 |
D050831.4 | TWP/SMET/C3 - Towrr lowered for sensor comparisons, data not representative | twpsmet60sC3.b1 |
D050920.4 | TWP/SMET/C2 - Instrument maintenance | twpsmet60sC2.b1 |
D050921.2 | TWP/SMET/C2 - ORG data incorrect | twpsmet60sC2.b1 |
D051005.6 | TWP/SMET/C2 - Tower down for calibration activities | twpsmet60sC2.b1 |
D051115.1 | TWP/SMET/C3 - Upper wind sensor failure | twpsmet60sC3.b1 |
D051117.2 | TWP/SMET/C3 - RH Values Artificially High | twpsmet60sC3.b1 |
D051117.3 | TWP/SMET/C3 - Instrument failure | twpsmet60sC3.b1 |
D060426.8 | TWP/SMET/C3 - ORG data noisy due to degraded ground | twpsmet60sC3.b1 |
D060502.3 | TWP/SMET/C3 - Reprocess: Windspeed2 calibration error | twpsmet60sC3.b1 |
D060522.1 | TWP/SMET/C1 - Upper wind speed sensor failure | twpsmet60sC1.b1 |
D060614.2 | TWP/SMET/C2 - Temp/RH probe failure | twpsmet60sC2.b1 |
D060627.1 | TWP/SMET/C3 - Anemometer Readings Incorrect | twpsmet60sC3.b1 |
D060817.3 | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC1.b1 |
D060818.2 | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC2.b1 |
D060818.3 | TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC3.b1 |
D061011.2 | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC1.b1 |
D061011.3 | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC2.b1 |
D061011.4 | TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC3.b1 |
D070306.2 | TWP/SMET/C3 - Barometer Data Incorrect | twpsmet60sC3.b1 |
D070306.3 | TWP/SMET/C3 - Data Incorrect Due to Water Intrusion | twpsmet60sC3.b1 |
D070409.1 | TWP/SMET/C1 - Upper winds failure | twpsmet60sC1.b1 |
D070510.5 | TWP/SMET/C2 - Lower windspeed sensor questionable | twpsmet60sC2.b1 |
D070520.1 | TWP/SMET/C2 - Lower Wind Sensor Direction Incorrect | twpsmet60sC2.b1 |
D070803.1 | TWP/SMET/C3 - Upper wind speed artificially slow | twpsmet60sC3.b1 |
D071105.1 | TWP/SMET/C3 - ORG overestimated rainfall | twpsmet60sC3.b1 |
D071219.1 | TWP/SMET/C2 - Lower Wind Speed Data Suspect | twpsmet60sC2.b1 |
D080225.1 | TWP/SMET/C3 - Upper Wind Sensor Data Questionable | twpsmet60sC3.b1 |
D980416.33 | TWP/SMET/C1 - Surface pressure spike | twpsmet60sC1.b1 |
D990910.2 | Calibration interruption at TWP Nauru | twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmetC2.00 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/1999 | 0000 | 11/25/1999 | 0450 |
Subject: | TWP/SMET/C1 - Reprocess: Skewed wind direction 1 data |
DataStreams: | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | Wind direction 1 data is incorrect. Manus SMET wind direction for Wind-1 (upper unit) was apparently skewed from its original position as a result of checking direction calibration. As near as we can tell the position of the sensor was rotated clockwise Somewhere in the range of 45-90 degrees, most likely in steps 22.5 degrees. Data comparison performed on 3-2-00, using a small amount of data from this instrument. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.00:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/1999 | 0450 | 04/02/2000 | 0000 |
Subject: | TWP/SMET/C1 - Wind Direction alignment method |
DataStreams: | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | During RESET10 we discovered a problem with wind direction from WIND1 (upper anemometer). During RESET8, WIND1 was corrected by rotating the alignment ring instead of the anemometer. This resulted in a better agreement between the 2 anemometers, but is not the correct way to align an anemometer. The alignment ring was readjusted for true north and the anemometer reset in the proper slot (the original problem). |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.00:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/24/1998 | 0000 | 01/08/1999 | 2100 |
Subject: | TWP/C2/SMET Wind Speeds |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
Description: | Beginning on December 24, Anemometer #1 (there are two anemometers on the SMET stand) occasionally reported wind speed values that were low relative to the other sensor. It is clear from the data that the low reading here is the anomalous one. |
Measurements: | twpsmetC2.00:
twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/09/1999 | 2300 | 01/10/1999 | 0600 |
Subject: | TWP/C2/SMET Wind Speeds |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
Description: | Beginning on December 24, Anemometer #1 (there are two anemometers on the SMET stand) occasionally reported wind speed values that were low relative to the other sensor. It is clear from the data that the low reading here is the anomalous one. |
Measurements: | twpsmetC2.00:
twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/1999 | 1430 | 01/14/1999 | 2100 |
Subject: | TWP/C2/SMET Winds Speeds |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
Description: | Beginning on December 24, Anemometer #1 (there are two anemometers on the SMET stand) occasionally reported wind speed values that were low relative to the other sensor. It is clear from the data that the low reading here is the anomalous one. |
Measurements: | twpsmetC2.00:
twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/1998 | 2105 | 12/03/1999 | 2105 |
Subject: | TWP/C2/SMET Temperature spike |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1 |
Description: | There was a brief -4 C spike in the temperature at approximately 21.5 Z on December 3. This spike also showed up in the vapor pressure which is calculated from the temperature and relative humidity. |
Measurements: | twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/20/1998 | 0000 | 12/31/1998 | 2359 |
Subject: | TWP/C2/SMET - Pressure sensitivity to rain |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1 |
Description: | The pressure behaves anomalously during rain events - even very mild ones. Normally, the pressure undergoes a smooth semi-diurnal oscillation with little higher frequency variability. However, during and shortly after rain events, the pressure signal exhibits abrupt changes - this sort of behavior has occasionally been observed at Manus but not with the regularity seen here. Examples of this behaviour occur on (date in the form YYYYMMDD): 19981212 19981215 19981216 19981217 19981223 19981224 19981225 19981230 19981231 |
Measurements: | twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/1999 | 0000 | 04/01/2000 | 2359 |
Subject: | TWP/SMET/C1 - Incorrect calibration for ARCS1 Wind speed |
DataStreams: | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmetC1.00 |
Description: | On 23 OCT 99, during Reset 8, an incorrect slope of was entered for the upper, windspeed1, wind speed sensor resulting in a 1% overestimate of the wind speed. The correct calibration was entered on 1 APR 2000 during RESET 10. |
Measurements: | twpsmet60sC1.b1:
twpsmetC1.00:
twpsmet60sC1.00:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/1998 | 0000 | 05/01/1998 | 0000 |
Subject: | TWP/SMET/C1 - Periodic spikes in temperature and water vapor pressure |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | Beginning on 980212 (day #43) spikes began to occur in the surface temperature and water vapor pressure. These spikes occur approximately once per hour (the spacing is very regular but occasionally an hour is skipped). In the one-minute averaged data, the temperature spike lasts for only one minute and dips approximately 4-5 C below the actual temperature. The one-minute standard deviation is more dramatic, rising from about 0.1 to about 35. The one minute temperature minimum is the most easily identifiable, each hour when the spike occurs, the minimum is -245.84. The vapor pressure mean jumps from about 30 to over 50 and the standard deviation jumps from less than 1 to more than 200. This problem was fixed by the RESET team on 980501. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/16/1998 | 0000 | 08/12/1999 | 0000 |
Subject: | TWP/SMET/C1 - Elevated relative humidity |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | The relative humidity has been exceeding 100% with unusual regularity. The most dramatic case occured on December 31 when the RH exceeded 100% for 18 hours and exceeded 103.5% for 6 hours with a maximum value of 103.8%. A PIF was submitted to propose changing this sensor. This change occurred on August 12, 1999. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/25/1996 | 0000 | 03/02/1999 | 2359 |
Subject: | TWP/SMET/C1 - C1 TWP Wind speed calibration factor problem/ Wind1 |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | Wind speed calibration factors were reversed between wind speed sensor 1 and 2. Results in less than 1% error, but correction factors may be used. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/25/1996 | 0000 | 03/02/1999 | 0000 |
Subject: | TWP/SMET/C1 - C1 SMET Incorrect wind speed calibration factors |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | Wind speed calibration factors were reversed between wind speed sensors 1 and 2. Error results in less than a 1% error in wind speed. A correction factor may be used. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2001 | 0445 | 10/23/2001 | 0645 |
Subject: | TWP/SMET/C1 - precip, wind, temperature |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | On 10/23/01 from 4:45 to 6:45 precipitation, wind, and temperature measurements were directly affected by RESET14 visit activities. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/2001 | 0300 | 01/03/2002 | 0841 |
Subject: | TWP/SMET/C2 - RH Sensor bad at Nauru |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1 |
Description: | RH sensor bad from 16 Dec 2001 03 GMT to 3 Jan 2002 |
Measurements: | twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2001 | 0040 | 10/24/2001 | 0100 |
Subject: | TWP/SMET/C1 - Sensor replacement |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | On 10/24/01 from 00:40 to 01:00 the T/RH probe was being replaced. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0312 | 11/19/1998 | 2359 |
Subject: | TWP/C2 Installation |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1, twpsmetC2.00 |
Description: | The Nauru ARCS was installed during the period September-November 1998. Data were being collected by the end of October, but for the first several weeks of November, the site was undergoing considerable change, so data from that period should be used with caution. The official site dedication occured on November 20. Most on-site activity ended following the dedication so we will consider November 20 to be the beginning of the Nauru operations. The last members of the installation team left Nauru on November 24. |
Measurements: | twpsmetC2.00:
twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1996 | 0001 | 06/22/2000 | 2359 |
Subject: | TWP/C1 - Incorrect elevation reported in Manus Data |
DataStreams: | twpgndrad20sC1.a1, twpgndrad60sC1.a1, twpgndrad60sC1.b1, twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | (From Bill Clements) We have been using 6m MSL as the station elevation of Momote (Manus). This was based on information posted near the station barometer in the Momote Weather Service Office. It turns out that the official WMO station elevation for Momote is 4m MSL and the upper air elevation is 5m MSL. All Manus instruments except the sonde station are at 4m MSL. The sonde launching station is at 5m MSL. |
Measurements: | twpgndrad20sC1.a1:
twpgndrad60sC1.b1:
twpsmet60sC1.b1:
twpgndrad60sC1.a1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/21/2002 | 0000 | 12/09/2002 | 0000 |
Subject: | TWP/MET/C3 - WIND DIRECTION SENSOR #2 MISALIGNED |
DataStreams: | twpsmet60sC3.a1, twpsmet60sC3.b1 |
Description: | Wind direction sensor #2 was misaligned and off by 15 degrees. |
Measurements: | twpsmet60sC3.a1:
twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/24/2003 | 0100 | 01/30/2003 | 0500 |
Subject: | TWP/SMET/C2 - Temperature Sensor Failed, Data Missing |
DataStreams: | twpsmet60sC2.a1, twpsmet60sC2.b1 |
Description: | The T/RH sensor on Nauru failed and was replaced (only the temperature sensor failed). |
Measurements: | twpsmet60sC2.b1:
twpsmet60sC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2002 | 0000 | 02/06/2003 | 2203 |
Subject: | TWP/SMET/C3 - Intermittent Pressure Sensor Failure |
DataStreams: | twpsmet60sC3.a1, twpsmet60sC3.b1 |
Description: | The Barometric Pressure sensor was intermittently failing at TWP site C3 Darwin. At times there were large pressure fluctuations occurring on the order of 24 mb in a few minutes. Times of large pressure fluctuations are readily apparent in the data. At these times the data should be carefully analyzed before determining it's usefulness. In between the periods of large fluctuations the readings appear to be stable. It is still suggested that independent pressure data be obtained for comparison before use. |
Measurements: | twpsmet60sC3.a1:
twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/15/2002 | 0000 | 02/27/2003 | 1100 |
Subject: | TWP/SMET/C1 - Reprocess: Miscalibrated windspeed2 data |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | A replacement anemometer (WSP2) was installed but the new anemometer calibration coefficients were not put into the logger. The multiplication factor was 0.09708 when it should have been 0.09763 (very small difference)the offset was 0.36 when it should have been 0.3 (a difference 0.06 m/s). |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/27/2003 | 1750 | 03/03/2003 | 1900 |
Subject: | TWP/SMET/C1 - Reprocess: Wind Speed #2 incorrect |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | Wind Speed #2 sensor was reporting values 10 times above the actual value. During an update to the logger configuration file the calibration multiplier was incorrectly entered. The multiplier inserted was .97630 when it should have been .09763. This caused the readings from this sensor to be incorrect. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/12/2003 | 1000 | 04/24/2003 | 0020 |
Subject: | TWP/SMET/C2 - WIND SENSOR #1 (HIGHER) INOPERATIVE |
DataStreams: | twpsmet60sC2.b1 |
Description: | Wind speed sensor #1 (higher sensor) failed. Wind speeds were reported as 0 m/s and wind direction data was not in agreement with wind sensor #2. Anemometer was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0628 | 06/03/2002 | 2359 |
Subject: | TWP/SMET/C3 - Data reported at 15sec rather than 60s intervals |
DataStreams: | twpsmet60sC3.a1, twpsmet60sC3.b1 |
Description: | Data in these data streams should be reports at 60sec intervals, but the logger program running during this time had the interval set to 15s. Users will have to average the 15sec data to get 60sec data. |
Measurements: | twpsmet60sC3.a1:
twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/2003 | 2300 | 12/02/2003 | 0300 |
Subject: | TWP/SMET/C1 - Tower Lowered for Cal Activities |
DataStreams: | twpsmet60sC1.b1, twpsmetC1.00 |
Description: | Tower lowered for instrument calibration |
Measurements: | twpsmet60sC1.b1:
twpsmetC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/10/2004 | 0000 | 04/24/2004 | 2303 |
Subject: | TWP/SMET/C2 - Wind Sensor failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Wind sensor #2 (lower one) failed. Wind direction showed little or no variation, unlike the colocated sensor #1. This caused lower wind speed values as the sensor was not rotating properly to orient itself parallel to the wind stream. The sensor was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2004 | 0000 | 01/19/2005 | 1530 |
Subject: | TWP/SMET/C1 - Poor Performing Sensor |
DataStreams: | twpsmet60sC1.b1 |
Description: | On 4/23/2004 condensation was found inside the target area of the ORG lens. A spare was sent and installed but did not solve the problem of high background voltage values causing rainrate values to be recorded during no rainfall. Values range anywhere from 0.06 to 0.12 mm/hr. A new Optical Rain Gage was installed on 01/19/2005 @ 13:30 GMT alleviating the high background voltage values that lead to consistent rainfall rates of around 0.11 to 0.15 mm/hr on a consistent basis. The formula used to calculate rainfall rates from the new Optical Rain Gage is: RR (mm/hr) = 25(V^1.87) - 0.15. Older model formula = 20(V^2) - 0.05 The two formulas result in comparable data. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2004 | 0000 | 05/17/2004 | 2259 |
Subject: | TWP/SMET/C3 - Rain blocked pressure port |
DataStreams: | twpsmet60sC3.b1 |
Description: | Atmospheric pressure values were suspect due to anomolous readings during heavy rain events. Suspect blockage of port due to rainfall. New logger and enclosure corrected problem. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/17/2004 | 2300 | 05/19/2004 | 0600 |
Subject: | TWP/SMET/C3 - Data missing due to Logger Change |
DataStreams: | twpsmet60sC3.b1 |
Description: | All Data from this site are missing during this time period due to a change of the collection system, logger and enclosure. Data resumed after change. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2004 | 2003 | 07/07/2004 | 0137 |
Subject: | TWP/SMET/C2 - All data missing due to logger change |
DataStreams: | twpsmet60sC2.b1 |
Description: | All data are missing at this site during this time period due to collection system, logger and enclosure change. Data resumed after change. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2004 | 0138 | 07/07/2004 | 2018 |
Subject: | TWP/SMET/C2 - connection and timing issues with new logger |
DataStreams: | twpsmet60sC2.b1 |
Description: | Atmospheric pressure data are intermittent during this time period due to connection problems and program timing issues with new logger installation. Data resumed normal operations after this time period. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0311 | 07/06/2004 | 0138 |
Subject: | TWP/SMET/C2 - Negative rainfall rates |
DataStreams: | twpsmet60sC2.b1 |
Description: | Due to limitations of the Zeno dataloggers all voltage measurements from the optical rain gauge (ORG) were converted to rainfall rates. This resulted in negative values for rain rates. Any value below 0.10 mm/hr should be considered to be 0 mm/hr. New dataloggers alleviated this problem. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0628 | 05/17/2004 | 2300 |
Subject: | TWP/SMET/C3 - Negative rainfall rates |
DataStreams: | twpsmet60sC3.b1 |
Description: | Due to limitations of the Zeno dataloggers all voltage measurements from the optical rain gauge (ORG) were converted to rainfall rates. This resulted in negative values for rain rates. Any value below 0.10 mm/hr should be considered to be 0 mm/hr. New dataloggers alleviated this problem. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2001 | 0000 | 10/26/2001 | 2359 |
Subject: | TWP/SMET/C1 - RESET14 activities |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | RESET14 activities may have affected data in subtle ways during this period. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/07/2004 | 0137 | 08/17/2004 | 2157 |
Subject: | TWP/SMET/C2 - Incorrect rainrate equation used. |
DataStreams: | twpsmet60sC2.b1 |
Description: | There are two different Optical Raingages in use at the TWP sites. The incorrect equation was used at this site. The incorrect equation is (25*(V^1.87)) - 0.15. The correct equation is (20*(V^2)) - 0.05. The correct equation was installed. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/03/2004 | 0307 | 08/17/2004 | 2138 |
Subject: | TWP/SMET/C1 - Incorrect rainrate equation used |
DataStreams: | twpsmet60sC1.b1 |
Description: | There are two different Optical Raingages in use at the TWP sites. The incorrect equation was used at this site. The incorrect equation is (25*(V^1.87)) - 0.15. The correct equation is (20*(V^2)) - 0.05. The correct equation was installed. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2005 | 2310 | 02/09/2005 | 2320 |
Subject: | TWP/SMET/C3 - Data non-representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower mast was lowered for inspection. Data during this time is non-representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/2005 | 1955 | 03/03/2005 | 0430 |
Subject: | TWP/SMET/C2 - Instrument failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Upper wind sensor failed. Wind direction was reported at a near constant value of 69 degrees and was significantly different than the lower wind sensor's direction values. Sensor replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/09/2005 | 2300 | 04/17/2005 | 0332 |
Subject: | TWP/SMET/C2 - Upper wind sensor failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | After replacement on 03/03/2005 the upper wind sensor began reporting erroneous values again on 03/09/2005 becoming very apparent after 03/17/2005. Wind speed data appears reasonable but the wind direction data do not agree well with the lower wind sensor. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2005 | 0230 | 04/19/2005 | 0050 |
Subject: | TWP/SMET/C3 - Instrument logger problem, Sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | The wind_lower direction sensor failed on 4/14 and disagrees significantly with the wind_upper direction. The wind_lower sensor reports a constant value around 0-15 degrees since it failed. Testing showed that excitation channel two on the logger was the problem. Logger was replaced by spare. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 03/09/2004 | 2243 |
Subject: | TWP/SMET/C1 - Negative rainfall rates |
DataStreams: | twpsmet60sC1.b1 |
Description: | Due to limitations of the Zeno dataloggers all voltage measurements from the optical rain gauge (ORG) were converted to rainfall rates. This resulted in negative values for rain rates. Any value below 0.10 mm/hr should be considered to be 0 mm/hr. New dataloggers alleviated this problem. New loggers were installed 03/09/2004 @ 2243 GMT. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/10/2005 | 1900 | 07/14/2005 | 2226 |
Subject: | TWP/SMET/C2 - RH failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | The RH values were frequently over MAX value of 104% starting on 7/10/2005 through end of period. Sensor was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower was lowered for radiometer change and characterization. Wind, temp and RH data not representative of local conditions. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Towrr lowered for sensor comparisons, data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | The tower was lowered for radiometer change and comparisons. During the time period the data listed was not representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/2005 | 1625 | 09/21/2005 | 1815 |
Subject: | TWP/SMET/C2 - Instrument maintenance |
DataStreams: | twpsmet60sC2.b1 |
Description: | Instrument calibration and testing were taking place for brief periods. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2005 | 0809 | 09/19/2005 | 1902 |
Subject: | TWP/SMET/C2 - ORG data incorrect |
DataStreams: | twpsmet60sC2.b1 |
Description: | Data from the ORG was incorrect due to heavy condensation on the lens. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2005 | 0200 | 09/19/2005 | 0330 |
Subject: | TWP/SMET/C2 - Tower down for calibration activities |
DataStreams: | twpsmet60sC2.b1 |
Description: | Tower lowered, wind vanes spun at different frequencies, and ORG tested with electronic rain. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/15/2005 | 2215 |
Subject: | TWP/SMET/C3 - Upper wind sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | A lightning strike destroyed the upper (#1) wind sensor. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/15/2005 | 0050 |
Subject: | TWP/SMET/C3 - RH Values Artificially High |
DataStreams: | twpsmet60sC3.b1 |
Description: | Lightning strike caused RH readings to be higher than reasonable (130%). |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/23/2005 | 0000 |
Subject: | TWP/SMET/C3 - Instrument failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | A lightning strike caused a failure in the ORG. Rainfall was being continually recorded. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/2005 | 0000 | 08/24/2006 | 2359 |
Subject: | TWP/SMET/C3 - ORG data noisy due to degraded ground |
DataStreams: | twpsmet60sC3.b1 |
Description: | SMET Rain Data began to be noisy during December 2005 and the problem slowly became worse over time so that in February 2006 the ORG was reporting with a constant background value of around 0.10 mm/hr rain rate. ORG voltage showed a near constant value above the 85mV threshold. The ground for the ORG had become degraded and was lost due to corrosion. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/27/2006 | 2330 | 05/02/2006 | 0201 |
Subject: | TWP/SMET/C3 - Reprocess: Windspeed2 calibration error |
DataStreams: | twpsmet60sC3.b1 |
Description: | The lower anemometer was replaced afer being removed in anticipation of cyclone Monica. The calibration coefficient was not changed during this period so the wind speeds are a little low (0.2%). |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2006 | 0000 | 10/14/2006 | 0237 |
Subject: | TWP/SMET/C1 - Upper wind speed sensor failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind speed sensor was binding at low wind speeds. Comparison to the lower wind sensor showed that the upper sensor flat-lines for long periods of time. The upper anemometer was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/05/2006 | 2144 | 06/15/2006 | 0427 |
Subject: | TWP/SMET/C2 - Temp/RH probe failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Temperature/RH probe failed and was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2006 | 0115 | 06/27/2006 | 0344 |
Subject: | TWP/SMET/C3 - Anemometer Readings Incorrect |
DataStreams: | twpsmet60sC3.b1 |
Description: | The Anemometer readings were incorrect due to the tower being lowered for maintenance. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 07/13/2006 | 2225 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 07/13/2006 | 2232 |
Subject: | TWP/SMET/C2 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC2.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/2003 | 0000 | 07/13/2006 | 2219 |
Subject: | TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC3.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 10/17/2006 | 0000 |
Subject: | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC1.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C1 SMET suite of instruments on 20061017. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC1.b1 datastream. Between 9/26 and 10/17, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19961009-20060926 and data from 19961009-20061017 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 11/29/2006 | 2014 |
Subject: | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC2.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C2 SMET suite of instruments on 20061129. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC2.b1 datastream. Between 9/26 and 11/29, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19981021-20060926 and data from 19981021-20061129 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0000 | 09/26/2006 | 1435 |
Subject: | TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC3.b1 |
Description: | A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 datastream. When reprocessing of this datastream occurs these variables will be added with -9999 (missing) as the value for dates prior to gauge installation. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2007 | 1723 | 03/03/2007 | 1228 |
Subject: | TWP/SMET/C3 - Barometer Data Incorrect |
DataStreams: | twpsmet60sC3.b1 |
Description: | The barometer data was incorrect or missing due to water intrusion. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2007 | 0056 | 03/03/2007 | 0123 |
Subject: | TWP/SMET/C3 - Data Incorrect Due to Water Intrusion |
DataStreams: | twpsmet60sC3.b1 |
Description: | Data became erratic and unreasonable after water intrusion into the enclosure. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/19/2007 | 0000 | 06/12/2007 | 0406 |
Subject: | TWP/SMET/C1 - Upper winds failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind instrument was inoperative resulting in constant wind direction and speed. A failed cable was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2007 | 0000 | 08/19/2007 | 2135 |
Subject: | TWP/SMET/C2 - Lower windspeed sensor questionable |
DataStreams: | twpsmet60sC2.b1 |
Description: | Lower wind speed sensor had degraded bearings. This sensor was reporting as much as 0.4 m/s lower values than the upper sensors when the winds were 2 m/s or below. The sensor was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/11/2007 | 0000 | 05/22/2007 | 0047 |
Subject: | TWP/SMET/C2 - Lower Wind Sensor Direction Incorrect |
DataStreams: | twpsmet60sC2.b1 |
Description: | Wind directions between lower and upper anemometers were displaced by approximately 45 degrees from 05/10/2007 through 05/21/2007 when a new sensor was installed to replace a failing sensor. The new sensor direction data incorrect. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2007 | 0000 | 09/18/2007 | 2315 |
Subject: | TWP/SMET/C3 - Upper wind speed artificially slow |
DataStreams: | twpsmet60sC3.b1 |
Description: | The upper wind measurement was consistently slower by around 0.5 - 1 m/s than the lower wind measurement. The bearings on the upper wind measurement were binding. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/02/2007 | 1855 | 11/02/2007 | 2010 |
Subject: | TWP/SMET/C3 - ORG overestimated rainfall |
DataStreams: | twpsmet60sC3.b1 |
Description: | Optical rain gauge (ORG) measured approximately 10mm higher than two co-located tipping bucket rain gauges and a co-located disdrometer. The ORG is typically higher but usually not by an order of magnitude. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/22/2007 | 0048 | 08/19/2007 | 2150 |
Subject: | TWP/SMET/C2 - Lower Wind Speed Data Suspect |
DataStreams: | twpsmet60sC2.b1 |
Description: | The lower wind speed bearings appeared to be binding at speeds below 2 m/s as it did not compare well with the upper wind monitor. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/05/2008 | 0000 | 02/27/2008 | 0315 |
Subject: | TWP/SMET/C3 - Upper Wind Sensor Data Questionable |
DataStreams: | twpsmet60sC3.b1 |
Description: | The wind_up vs. wind_low wind directions diverged, with the differences between the two eventually reached 45 degrees. The Upper sensor was found to be mis-aligned and loose. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/04/1997 | 0249 | 09/04/1997 | 0304 |
Subject: | TWP/SMET/C1 - Surface pressure spike |
DataStreams: | twpsmet60sC1.b1 |
Description: | A spike occured in the surface pressure. The reported pressure rose as high as 1300 hPa but returned to normal levels after 15 minutes. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/09/1999 | 0030 | 08/11/1999 | 0300 |
Subject: | Calibration interruption at TWP Nauru |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmetC2.00 |
Description: | Occasional calibration interuption occured while calibration SMET instruments |
Measurements: | twpsmet60sC1.b1:
twpsmetC2.00:
twpsmet60sC1.a1:
|