DQR ID | Subject | Data Streams Affected |
---|---|---|
D040127.3 | TWP/SONDE/C2 - Broken temperature sensors | twpsondewnpnC2.b1 |
D040425.1 | TWP/SONDE/C2 - Bad RH values aloft | twpsondewnpnC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0040 | 10/13/2003 | 0203 |
11/08/2003 | 2335 | 11/09/2003 | 0215 |
11/18/2003 | 1205 | 11/18/2003 | 1340 |
11/19/2003 | 1156 | 11/19/2003 | 1346 |
11/21/2003 | 0044 | 11/21/2003 | 0113 |
12/20/2003 | 1225 | 12/20/2003 | 1408 |
12/21/2003 | 0056 | 12/21/2003 | 0126 |
12/21/2003 | 1155 | 12/21/2003 | 1344 |
12/23/2003 | 0042 | 12/23/2003 | 0212 |
12/23/2003 | 1137 | 12/23/2003 | 1259 |
12/23/2003 | 2342 | 12/24/2003 | 0132 |
01/08/2004 | 1135 | 01/08/2004 | 1309 |
01/09/2004 | 1157 | 01/09/2004 | 1342 |
01/09/2004 | 1359 | 01/09/2004 | 1537 |
01/12/2004 | 2342 | 01/13/2004 | 0214 |
01/16/2004 | 1409 | 01/16/2004 | 1548 |
01/16/2004 | 2337 | 01/17/2004 | 0109 |
01/18/2004 | 1135 | 01/18/2004 | 1256 |
01/19/2004 | 0223 | 01/19/2004 | 0346 |
01/19/2004 | 2329 | 01/20/2004 | 0157 |
01/21/2004 | 1252 | 01/21/2004 | 1400 |
01/22/2004 | 0052 | 01/22/2004 | 0227 |
01/23/2004 | 1137 | 01/23/2004 | 1319 |
01/26/2004 | 1202 | 01/26/2004 | 0146 |
02/17/2004 | 1406 | 02/17/2004 | 1536 |
02/26/2004 | 1221 | 02/26/2004 | 1350 |
03/09/2004 | 1145 | 03/09/2004 | 1315 |
03/14/2004 | 1131 | 03/14/2004 | 1300 |
03/22/2004 | 1134 | 03/22/2004 | 1300 |
03/29/2004 | 2332 | 03/30/2004 | 0047 |
03/30/2004 | 2334 | 03/31/2004 | 0059 |
04/11/2004 | 2348 | 04/12/2004 | 0120 |
05/01/2004 | 2335 | 05/02/2004 | 0110 |
05/02/2004 | 2336 | 05/03/2004 | 0109 |
05/13/2004 | 2334 | 05/14/2004 | 0112 |
05/14/2004 | 1406 | 05/14/2004 | 1411 |
06/07/2004 | 0224 | 06/07/2004 | 0351 |
06/25/2004 | 1146 | 06/25/2004 | 1328 |
07/02/2004 | 0059 | 07/02/2004 | 0217 |
07/20/2004 | 0220 | 07/20/2004 | 0358 |
07/24/2004 | 1235 | 07/24/2004 | 1355 |
08/22/2004 | 1144 | 08/22/2004 | 1207 |
09/08/2004 | 1425 | 09/08/2004 | 1605 |
09/08/2004 | 2338 | 09/08/2004 | 2345 |
10/18/2004 | 2346 | 10/19/2004 | 0128 |
11/29/2004 | 2321 | 11/30/2004 | 0105 |
12/08/2004 | 1143 | 12/08/2004 | 1306 |
12/30/2004 | 1130 | 12/30/2004 | 1239 |
Subject: | TWP/SONDE/C2 - Broken temperature sensors |
DataStreams: | twpsondewnpnC2.b1 |
Description: | The temperature sensors on these radiosondes broke after launch. All temperature-related data are incorrect. |
Measurements: | twpsondewnpnC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/21/2004 | 1232 | 04/21/2004 | 1248 |
Subject: | TWP/SONDE/C2 - Bad RH values aloft |
DataStreams: | twpsondewnpnC2.b1 |
Description: | It appears that the RH sensor heating circuit on this sonde failed about 24 minutes into the flight. The RH values look reasonable, though somewhat noisy, from the surface to approximately 6.9 km (439 hPa) at which point they begin to oscillate between 0 and 100 %. This behavior stops when the sonde reaches approximately 11.5km (232 hPa) when the air temperature drops below -40 degC. The oscillation stops at that point, though the RH remains higher than might be expected even under supersaturated conditions. |
Measurements: | twpsondewnpnC2.b1:
|