DQR ID | Subject | Data Streams Affected |
---|---|---|
D000625.1 | SGP/SONDE/C1 - Operator used wrong calibration tape | sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, sgpsondewndcalcC1.c1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D000801.2 | SGP/SONDE/C1 - Incorrect absolute pressure | sgpsondewndcalcC1.c1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
D990210.1 | SGP/SONDE - Wrong cal tape | sgpsondewndcalcC1.c1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/23/2000 | 2330 | 06/24/2000 | 0100 |
Subject: | SGP/SONDE/C1 - Operator used wrong calibration tape |
DataStreams: | sgpsondeC1.00, sgpsondeC1.a0, sgpsondeC1.a1, sgpsondewndcalcC1.c1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | It appears as if the operators had a problem with the original sounding and launched a second radiosonde. Apparently they used the calibration tape that came with the first radiosonde (serial number 943156405) in place of the one that was correct for the second radiosonde (sn 943156406) and all the data are bad after the first sample (surface values). |
Measurements: | sgpsondewrpnC1.a0:
sgpsondeC1.a0:
sgpsondewrpnC1.00:
sgpsondewndcalcC1.c1:
sgpsondeC1.a1:
sgpsondeC1.00:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/25/2000 | 0533 | 07/25/2000 | 0724 |
Subject: | SGP/SONDE/C1 - Incorrect absolute pressure |
DataStreams: | sgpsondewndcalcC1.c1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | The surface pressure (983.5 hPa) reported by this radiosonde (sn 94315713) is substantially higher than that reported by the THWAPS (~977) or by the sondes used in the previous and following soundings (976.6, 977.3). This may be a calibration problem or a mechanic problem on launch. The pressure values reported in the remainder of the sounding are consistent with the reported surface pressure, so the altitudes should be correct. The absolute pressure values may be high by about 5 hPa. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.00:
sgpsondewndcalcC1.c1:
sgpsondewrpnC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/1999 | 0834 | 02/09/1999 | 1016 |
Subject: | SGP/SONDE - Wrong cal tape |
DataStreams: | sgpsondewndcalcC1.c1, sgpsondewrpnC1.00, sgpsondewrpnC1.a0, sgpsondewrpnC1.a1 |
Description: | There is something wrong with this sounding, its not clear exactly what happened, however. The first values reported by the radiosonde are much different from those entered by the operator. My impression is that this is a calibration tape mixup. The raw data file shows that the sounding software was initialized twice for this sounding, the first time showing sonde 823360910, the second time showing sonde 819351814. There is no enties in the operations log that describe a problem with a sonde before launch. |
Measurements: | sgpsondewrpnC1.a0:
sgpsondewrpnC1.00:
sgpsondewndcalcC1.c1:
sgpsondewrpnC1.a1:
|