Data Quality Reports for Session: 113927 User: goulven1mo Completed: 09/01/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D080625.2SGP/SIRS/E8 - Ventilator Fan Failuresgpsirs20sE8.a0, sgpsirsE8.00, sgpsirsE8.b1
D080701.1SGP/SIRS/E5 - Incorrect Calibration Coefficientsgpsirs20sE5.a0, sgpsirsE5.00, sgpsirsE5.b1


DQRID : D080625.2
Start DateStart TimeEnd DateEnd Time
05/06/2008033007/08/20081750
Subject:
SGP/SIRS/E8 - Ventilator Fan Failure
DataStreams:sgpsirs20sE8.a0, sgpsirsE8.00, sgpsirsE8.b1
Description:
At the start time specified the PSP (down_short_hemisp) ventilator fan failed. PM replaced 
the fan 7/8/2008 1750.  Ventilator failure does adversely affect radiometer measurements 
but the extent has not been quantified.  Mentor suggests using component sum method 
detailed in SIRS instrument handbook where down_short_hemisp = down_short_diffuse + 
short_direct_normal * cos (zenith angle) for each time record.
Measurements:sgpsirsE8.00:
  • null(Raw data stream - documentation not supported)

sgpsirs20sE8.a0:
  • Instantaneous Downwelling Hemispheric Shortwave, Unshaded Pyranometer Thermopile
    Voltage(inst_global)

sgpsirsE8.b1:
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Maxima(down_short_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Minima(down_short_hemisp_min)
  • Down-welling unshaded pyranometer voltage(down_short_hemisp)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyranometer, Standard
    Deviation(down_short_hemisp_std)


Back To Table of Contents

DQRID : D080701.1
Start DateStart TimeEnd DateEnd Time
06/25/2008173607/10/20081530
Subject:
SGP/SIRS/E5 - Incorrect Calibration Coefficient
DataStreams:sgpsirs20sE5.a0, sgpsirsE5.00, sgpsirsE5.b1
Description:
When instruments were changed out on 6/25 the UIR (up_long_hemisp measurement) S/N and cal 
coefficient reverted to a previous UIR instrument that was damaged and removed on 
12/28/2007. The UIR S/N should be 31637F3 with corresponding cal K=.2526 instead of applied 
value of K=.2571. 
However, for the PIR longwave instrument in this inverted position (UIR) the K1 term 
contributes approximately 0-7% of the total upwelling longwave signal throughout the day, or 
about 0-30 W/m2.  The cal coefficient is too high by approximately 2%. Thus, the UIR data, 
as is, ranges from 0 to .6 W/m2 too low since this term is always negative.  The 
uncertainty of this instrument is approximately +-10 W/m2 with respect to the WSIG. Reprocessing 
is not recommended.
Measurements:sgpsirsE5.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Standard
    Deviation(up_long_hemisp_std)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)

sgpsirs20sE5.a0:
  • Instantaneous Upwelling Pyrgeometer Thermopile(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)

sgpsirsE5.00:
  • null(Raw data stream - documentation not supported)


Back To Table of Contents



END OF DATA