DQR ID | Subject | Data Streams Affected |
---|---|---|
D050210.1 | TWP/SMET/C3 - Data non-representative | twpsmet60sC3.b1 |
D050418.2 | TWP/SMET/C3 - Lower wind direction sensor failure | twpsmet60sC3.b1 |
D050808.3 | TWP/SMET/C3 - Data not representative | twpsmet60sC3.b1 |
D050920.4 | TWP/SMET/C2 - Instrument maintenance | twpsmet60sC2.b1 |
D050921.2 | TWP/SMET/C2 - ORG data incorrect | twpsmet60sC2.b1 |
D060614.2 | TWP/SMET/C2 - Temp/RH probe failure | twpsmet60sC2.b1 |
D060818.3 | TWP/SMET/C3 - Reprocess: Barometric pressure units change | twpsmet60sC3.b1 |
D061011.3 | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC2.b1 |
D061011.4 | TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC3.b1 |
D061214.2 | TWP/SMET/C2 - Barometer failure | twpsmet60sC2.b1 |
D080922.5 | TWP/SMET/C3 - Missing data | twpsmet60sC3.b1 |
D081022.8 | TWP/SMET/C3 - Wind Speed data contain offset due to induced voltage | twpsmet60sC3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2005 | 2310 | 02/09/2005 | 2320 |
Subject: | TWP/SMET/C3 - Data non-representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower mast was lowered for inspection. Data during this time is non-representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2005 | 0230 | 04/19/2005 | 0050 |
Subject: | TWP/SMET/C3 - Lower wind direction sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | The lower wind direction sensor failed on 4/14 and disagrees significantly with the upper wind direction sensor. The lower wind sensor reports a constant value around 0-15 degrees since it failed. Testing showed that excitation channel two on the logger was the problem. Logger was replaced by spare. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower was lowered for radiometer change and characterization. Data are not representative of local conditions. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/21/2005 | 1625 | 09/21/2005 | 1815 |
Subject: | TWP/SMET/C2 - Instrument maintenance |
DataStreams: | twpsmet60sC2.b1 |
Description: | Instrument calibration and testing were taking place for brief periods. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/19/2005 | 0809 | 09/19/2005 | 1902 |
Subject: | TWP/SMET/C2 - ORG data incorrect |
DataStreams: | twpsmet60sC2.b1 |
Description: | Data from the ORG was incorrect due to heavy condensation on the lens. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/05/2006 | 2144 | 06/15/2006 | 0427 |
Subject: | TWP/SMET/C2 - Temp/RH probe failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | Temperature/RH probe failed and was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/2003 | 0000 | 05/19/2004 | 2359 |
Subject: | TWP/SMET/C3 - Reprocess: Barometric pressure units change |
DataStreams: | twpsmet60sC3.b1 |
Description: | Barometric pressure data during this time period is reported in hPa. After this time, the data have been converted to kPa to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/21/1998 | 0000 | 11/29/2006 | 2014 |
Subject: | TWP/SMET/C2 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC2.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C2 SMET suite of instruments on 20061129. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC2.b1 datastream. Between 9/26 and 11/29, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19981021-20060926 and data from 19981021-20061129 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0000 | 09/26/2006 | 1435 |
Subject: | TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC3.b1 |
Description: | A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 datastream. When reprocessing of this datastream occurs these variables will be added with -9999 (missing) as the value for dates prior to gauge installation. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2006 | 1941 | 01/23/2007 | 0023 |
Subject: | TWP/SMET/C2 - Barometer failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | The barometer began experiencing problems with noise resulting in incorrect data. The problem becames worse over time; after 12/19/2006 @0151 GMT the barometer data is completely missing. The barometer was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/17/2004 | 2202 | 05/19/2004 | 0559 |
04/23/2006 | 2357 | 04/27/2006 | 0348 |
Subject: | TWP/SMET/C3 - Missing data |
DataStreams: | twpsmet60sC3.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2004 | 0000 | 10/17/2008 | 0627 |
Subject: | TWP/SMET/C3 - Wind Speed data contain offset due to induced voltage |
DataStreams: | twpsmet60sC3.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | twpsmet60sC3.b1:
|