Description: | The PIR exhibited a positive bias of approximately 6 W/m2 compared to the upwelling IRT
due to ventilator fan failure. Thus, the up_long_hemisp, data stream is high by about 6
W/m2.
Bias is not consistent and depends on atmospheric conditions. Either disregard the
up_long_hemisp data or use the IRT for an estimate of the up_long_hemisp values during this time
period. |
Measurements: | nsagndrad20sC1.a0: - Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_hemisp_case_resist)
- Instantaneous Upwelling Pyrgeometer Thermopile(inst_up_long_hemisp_tp)
- Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_hemisp_dome_resist)
nsagndrad60sC1.b1: - Instantaneous Upwelling Pyrgeometer Dome Thermistor Temperature(inst_up_long_hemisp_dome_temp)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Standard
Deviation(up_long_hemisp_std) - Instantaneous Upwelling Pyrgeometer Thermopile(inst_up_long_hemisp_tp)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Maxima(up_long_hemisp_max)
- Instantaneous Upwelling Pyrgeometer Case Thermistor Temperature(inst_up_long_hemisp_case_temp)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
|