DQR ID | Subject | Data Streams Affected |
---|---|---|
D050321.6 | PYE/TSI/M1 - Misalignment | pyetsicldmaskM1.a1, pyetsimovieM1.a1, pyetsiskycoverM1.b1, pyetsiskyimageM1.a1 |
D050503.2 | PYE/TSI/M1 - water in lens | pyetsiM1.00, pyetsicldmaskM1.a1, pyetsimovieM1.a1, pyetsiskycoverM1.b1, pyetsiskyimageM1.a1 |
D050811.6 | PYE/SONDE/M1 - Broken RH sensor | pyesondewnpnM1.b1 |
D050823.1 | PYE/SONDE/M1 - Bad temperature and RH | pyesondewnpnM1.b1 |
D050912.3 | PYE/SONDE/M1 - Bad RH sensor | pyesondewnpnM1.b1 |
D050912.4 | PYE/SONDE/M1 - Bad RH sensor | pyesondewnpnM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2005 | 1525 | 04/07/2005 | 1715 |
Subject: | PYE/TSI/M1 - Misalignment |
DataStreams: | pyetsicldmaskM1.a1, pyetsimovieM1.a1, pyetsiskycoverM1.b1, pyetsiskyimageM1.a1 |
Description: | The TSI was not aligned with true north. |
Measurements: | pyetsiskyimageM1.a1:
pyetsiskycoverM1.b1:
pyetsicldmaskM1.a1:
pyetsimovieM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2005 | 0032 | 05/07/2005 | 1822 |
Subject: | PYE/TSI/M1 - water in lens |
DataStreams: | pyetsiM1.00, pyetsicldmaskM1.a1, pyetsimovieM1.a1, pyetsiskycoverM1.b1, pyetsiskyimageM1.a1 |
Description: | The lower half of the TSI images were blurry due to water in the camera lens and/or filter. The problem was corrected when both were cleaned. |
Measurements: | pyetsiM1.00:
pyetsiskyimageM1.a1:
pyetsiskycoverM1.b1:
pyetsicldmaskM1.a1:
pyetsimovieM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2005 | 1129 | 08/02/2005 | 1225 |
Subject: | PYE/SONDE/M1 - Broken RH sensor |
DataStreams: | pyesondewnpnM1.b1 |
Description: | One of the RH sensors on this radiosonde was bad. The RH oscillates rapidly between near 0 and ambient up to approximately 11 km. The near-zero values are incorrect. |
Measurements: | pyesondewnpnM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/30/2005 | 0528 | 06/30/2005 | 0711 |
Subject: | PYE/SONDE/M1 - Bad temperature and RH |
DataStreams: | pyesondewnpnM1.b1 |
Description: | The temperature appears to be approximately 10 degC too high during the entire sounding. This also causes the RH to be too low. Most likely the problem was a bad temperature sensor. |
Measurements: | pyesondewnpnM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/31/2005 | 1130 | 08/31/2005 | 1140 |
Subject: | PYE/SONDE/M1 - Bad RH sensor |
DataStreams: | pyesondewnpnM1.b1 |
Description: | The RH sensor heating circuit failed on this sounding. RH values during the first part of the sounding oscillate between a reasonable value and a very low value. It appears that (a) the heating was out of phase, and (b) incorrect values were being reported during the first 10 minutes or so of flight. |
Measurements: | pyesondewnpnM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/29/2005 | 1128 | 08/29/2005 | 1340 |
Subject: | PYE/SONDE/M1 - Bad RH sensor |
DataStreams: | pyesondewnpnM1.b1 |
Description: | Two problems here - failure of the RH sensor heating circuit and a bad RH sensor. |
Measurements: | pyesondewnpnM1.b1:
|