Data Quality Reports for Session: 119199 User: dmfoper Completed: 04/29/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D011102.3SGP EBBR E19 - E19 EBBR Soil Temperature 3 Intermittenmtly Incorrectsgp15ebbrE19.b1, sgp30ebbrE19.b1
D011127.30SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020716.15SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020717.26SGP/EBBR/E19 - Reprocess: Soil Heat Flow #3 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D020717.27SGP/EBBR/E19 - Reprocess: Soil Temperature #3 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D020717.29SGP/EBBR/E19 - Most Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.30SGP/EBBR/E19 - All Measurements Zerosgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.31SGP/EBBR/E19 - Pressure Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020717.32SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacementsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.33SGP/EBBR/E19 - Low Battery Voltagesgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D021113.1SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR
Header
sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1,
sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1,
sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1
D030110.24SGP/EBBR/E19 - Data Incorrect during 6 month checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.10SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #3 Incorrect Calibrationsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030116.11SGP/EBBR/E19 - Water in Net Radiometer Domesgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.6SGP/EBBR/E19 - Incorrect data after power interruptionsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.7SGP/EBBR/E19 - Most Measurements Zero following Restartsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.8SGP/EBBR/E19 - REPROCESS: Soil Temperature #3 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030116.9SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #1 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030117.1SGP/EBBR/E19 - Pressure Lowsgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030117.4SGP/EBBR/E19 - Soil Temperatures #2 and #3 Intermittently Badsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030117.6SGP/EBBR/E19 - Measurements zero following logger restartsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030117.7SGP/EBBR/E19 - Change of Energy Storage Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.11SGP/EBBR/E19 - Soil Temperature #2 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.12SGP/EBBR/E19 - All Data Zerossgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030711.13SGP/EBBR/E19 - Several Soil Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.14SGP/EBBR/E19 -Several Soil Measurements Were Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.17SGP/EBBR/E19 - Soil Heat Flow #3 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D040525.4SGP/EBBR/E19 - Metadata errors sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D050719.12SGP/EBBR/E19 - Improved EBBR CR10 Programsgp30ebbrE19.b1


DQRID : D011102.3
Start DateStart TimeEnd DateEnd Time
05/07/2001000005/07/20012356
05/21/2001000005/21/20012356
Subject:
SGP EBBR E19 - E19 EBBR Soil Temperature 3 Intermittenmtly Incorrect
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1
Description:
Soil Temperature # 3 was incorrect at times during these periods.
Compare G3 with G1, G2, G4, and G5 to determine when TS3 was incorrect.
Measurements:sgp15ebbrE19.b1:
  • soil_temp_3_rr

sgp30ebbrE19.b1:
  • energy_storage_change_3
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • latent_heat_flux
  • soil_temp_3
  • surface_soil_heat_flux_avg


Back To Table of Contents

DQRID : D011127.30
Start DateStart TimeEnd DateEnd Time
04/19/2001170004/19/20012001
10/03/2001141310/03/20011506
Subject:
SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The data was incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • temp_air_left
  • soil_heat_flow_4_mv
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D020716.15
Start DateStart TimeEnd DateEnd Time
05/16/2002143005/16/20021501
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The indicated measurements were incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • temp_air_left
  • temp_air_right

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • vapor_pressure_bottom
  • temp_trh_bottom
  • bowen_ratio
  • temp_trh_top
  • rh_top_fraction
  • rh_bottom_fraction
  • temp_air_top
  • vapor_pressure_top
  • latent_heat_flux
  • temp_air_bottom


Back To Table of Contents

DQRID : D020717.26
Start DateStart TimeEnd DateEnd Time
02/08/2002163002/21/20021601
Subject:
SGP/EBBR/E19 - Reprocess: Soil Heat Flow #3 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil heat flow 3 was incorrect, with values too large positive and too small negative.  
Since the calibration in the program is the one supplied by the vendor, it appears 
that the calibration may be wrong.

Average soil heat flow, sensible heat flux and latent heat flux can be recalculated
using the remaining good sets of soil sensors as follows:
   ave_shf = (g1 + g2 + g4 + g5)/4
   e = -(q + ave_shf)/(1 + bowen)
   h = -(q + e + ave_shf)
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 3(shf3)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • corr_soil_heat_flow_3
  • latent_heat_flux
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg


Back To Table of Contents

DQRID : D020717.27
Start DateStart TimeEnd DateEnd Time
11/01/2001000006/30/20022356
Subject:
SGP/EBBR/E19 - Reprocess: Soil Temperature #3 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature 3 was incorrect for the period indicated, causing average soil flow, 
latent heat flux, and sensible heat flux to be incorrect also.

Average soil heat flow, sensible heat flux, and latent heat flux can be recalculated using 
the remaining good sets of soil sensors as follows:

ave_shf = (g1 + g2 + g4 + g5)/4

e = -(q + ave_shf)/(1 + bowen)

h = -(q + e + ave_shf)
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • energy_storage_change_3
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • latent_heat_flux
  • soil_temp_3
  • surface_soil_heat_flux_avg


Back To Table of Contents

DQRID : D020717.29
Start DateStart TimeEnd DateEnd Time
04/08/2002043004/08/20020431
05/19/2002040005/19/20020401
Subject:
SGP/EBBR/E19 - Most Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Most measurements were incorrect for the indicated periods.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • battery_voltage
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D020717.30
Start DateStart TimeEnd DateEnd Time
04/18/2002173004/18/20021731
05/05/2002140005/05/20021401
06/27/2002150006/27/20021501
Subject:
SGP/EBBR/E19 - All Measurements Zero
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Measurements are often zero the first period after a CR10 restart.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • battery_voltage
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D020717.31
Start DateStart TimeEnd DateEnd Time
04/24/2002160004/24/20022201
05/25/2002210005/25/20022101
05/25/2002220005/25/20022231
05/29/2002173005/29/20021731
05/29/2002183005/29/20021831
05/29/2002200005/29/20022001
05/29/2002213005/29/20022131
05/30/2002150005/30/20021501
06/15/2002190006/15/20021901
Subject:
SGP/EBBR/E19 - Pressure Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The pressure measurement was incorrect.  This is used to determine the Bowen ratio, 
thereby affecting the sensible and latent heat fluxes also.
Measurements:sgp5ebbrE19.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • h(h)
  • Retrieved pressure profile(pres)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • latent_heat_flux
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D020717.32
Start DateStart TimeEnd DateEnd Time
05/16/2002200005/17/20020531
Subject:
SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacement
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Replacement of the CR10 datalogger caused most measurements to be incorrect, reason 
unknown.
A different logger was installed at the end of the specified period and this resolved the 
problem.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • battery_voltage
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D020717.33
Start DateStart TimeEnd DateEnd Time
05/27/2002000006/02/20021031
Subject:
SGP/EBBR/E19 - Low Battery Voltage
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
A battery voltage of less than 10.5 causes many of the sensors to produce incorrect 
measurements.
The 15 minute data contains the battery voltage and should be used to determine whether 
the measurements are usable, based on battery voltage.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D021113.1
Start DateStart TimeEnd DateEnd Time
04/01/2001000006/18/20032359
Subject:
SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR 
Header
DataStreams:sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1,
sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1,
sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1
Description:
The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the 
EBBR 30 minute netcdf header have been incorrect from the beginning 
of processing the data into netcdf files.  The correct units are 
MJ/m^3/C.

This does not affect the data quality.
Measurements:sgp30ebbrE8.b1:
  • soil_heat_capacity_2
  • soil_heat_capacity_1
  • soil_heat_capacity_3
  • soil_heat_capacity_5
  • soil_heat_capacity_4

sgp30ebbrE4.b1:
  • soil_heat_capacity_4
  • soil_heat_capacity_2
  • soil_heat_capacity_3
  • soil_heat_capacity_5
  • soil_heat_capacity_1

sgp30ebbrE9.b1:
  • soil_heat_capacity_3
  • soil_heat_capacity_5
  • soil_heat_capacity_1
  • soil_heat_capacity_4
  • soil_heat_capacity_2

sgp30ebbrE18.b1:
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_heat_capacity_3
  • soil_heat_capacity_4
  • soil_heat_capacity_5

sgp30ebbrE26.b1:
  • soil_heat_capacity_3
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • soil_heat_capacity_2
  • soil_heat_capacity_1

sgp30ebbrE13.b1:
  • soil_heat_capacity_3
  • soil_heat_capacity_5
  • soil_heat_capacity_4
  • soil_heat_capacity_2
  • soil_heat_capacity_1

sgp30ebbrE7.b1:
  • soil_heat_capacity_5
  • soil_heat_capacity_4
  • soil_heat_capacity_2
  • soil_heat_capacity_3
  • soil_heat_capacity_1

sgp30ebbrE12.b1:
  • soil_heat_capacity_2
  • soil_heat_capacity_4
  • soil_heat_capacity_3
  • soil_heat_capacity_1
  • soil_heat_capacity_5

sgp30ebbrE19.b1:
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • soil_heat_capacity_3
  • soil_heat_capacity_1
  • soil_heat_capacity_2

sgp30ebbrE20.b1:
  • soil_heat_capacity_2
  • soil_heat_capacity_1
  • soil_heat_capacity_5
  • soil_heat_capacity_3
  • soil_heat_capacity_4

sgp30ebbrE22.b1:
  • soil_heat_capacity_3
  • soil_heat_capacity_5
  • soil_heat_capacity_2
  • soil_heat_capacity_4
  • soil_heat_capacity_1

sgp30ebbrE2.b1:
  • soil_heat_capacity_5
  • soil_heat_capacity_4
  • soil_heat_capacity_1
  • soil_heat_capacity_3
  • soil_heat_capacity_2

sgp30ebbrE15.b1:
  • soil_heat_capacity_5
  • soil_heat_capacity_1
  • soil_heat_capacity_3
  • soil_heat_capacity_2
  • soil_heat_capacity_4


Back To Table of Contents

DQRID : D030110.24
Start DateStart TimeEnd DateEnd Time
11/14/2002160011/14/20021630
Subject:
SGP/EBBR/E19 - Data Incorrect during 6 month checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Some of the measurements are incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • temp_air_left
  • temp_air_right

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • vapor_pressure_bottom
  • temp_trh_bottom
  • bowen_ratio
  • temp_trh_top
  • rh_top_fraction
  • rh_bottom_fraction
  • temp_air_top
  • vapor_pressure_top
  • latent_heat_flux
  • temp_air_bottom


Back To Table of Contents

DQRID : D030116.10
Start DateStart TimeEnd DateEnd Time
02/08/2002160004/18/20021700
Subject:
SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #3 Incorrect Calibration
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #3 was incorrect during these periods.

To use the measurements, multiply Soil Heat Flow #3 by 0.726.
  
See below for a work-around.
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 3(shf3)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • corr_soil_heat_flow_3
  • latent_heat_flux
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg


Back To Table of Contents

DQRID : D030116.11
Start DateStart TimeEnd DateEnd Time
08/25/2002210009/05/20021430
Subject:
SGP/EBBR/E19 - Water in Net Radiometer Dome
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The net radiation measurement is incorrect. Measurements 
calculated using net radiation are also incorrect.  

No work-around is possible for this situation.
Measurements:sgp5ebbrE19.a1:
  • net radiation(q)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • h(h)
  • net radiation(q)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • latent_heat_flux
  • net_radiation


Back To Table of Contents

DQRID : D030116.6
Start DateStart TimeEnd DateEnd Time
07/11/2002160007/11/20021629
Subject:
SGP/EBBR/E19 - Incorrect data after power interruption
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most measurements are suspect after a power outage, caused by
replacement of the solar panel regulator.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D030116.7
Start DateStart TimeEnd DateEnd Time
07/11/2002150007/11/20021529
Subject:
SGP/EBBR/E19 - Most Measurements Zero following Restart
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • battery_voltage
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D030116.8
Start DateStart TimeEnd DateEnd Time
07/24/2002143007/24/20021500
Subject:
SGP/EBBR/E19 - REPROCESS: Soil Temperature #3 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature #3 was incorrect while electronic connections
were being checked. The indicated measurements 
that are calculated using Soil Temperature #3 are 
also incorrect.

See a work-around below.
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • energy_storage_change_3
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • latent_heat_flux
  • soil_temp_3
  • surface_soil_heat_flux_avg


Back To Table of Contents

DQRID : D030116.9
Start DateStart TimeEnd DateEnd Time
07/25/2002093007/25/20021300
07/26/2002090007/26/20020900
07/26/2002113007/26/20021130
07/27/2002063007/27/20020830
07/28/2002030007/28/20020630
07/28/2002093007/28/20021230
07/29/2002080007/29/20020830
08/08/2002160008/08/20021600
08/09/2002023008/09/20020300
08/09/2002073008/09/20020730
08/09/2002093008/09/20021100
08/09/2002140008/09/20021600
08/10/2002013008/10/20020230
08/10/2002120008/10/20021330
08/13/2002060008/13/20020700
08/13/2002203008/13/20022030
08/13/2002213008/14/20020330
08/14/2002130008/14/20021330
08/21/2002163008/22/20021430
Subject:
SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #1 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #1 was incorrect during these periods.  
See below for a work-around.
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 1(shf1)
  • h(h)
  • soil heat flow, site 1(g1)

sgp30ebbrE19.b1:
  • surface_soil_heat_flux_1
  • corr_soil_heat_flow_1
  • sensible_heat_flux
  • latent_heat_flux
  • soil_heat_flow_1
  • surface_soil_heat_flux_avg


Back To Table of Contents

DQRID : D030117.1
Start DateStart TimeEnd DateEnd Time
10/17/2002143010/31/20021600
Subject:
SGP/EBBR/E19 - Pressure Low
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The barometric pressure sensor read about 10 mb low during this
period because a lead from the sensor was loose or disconnected.

The error does not affect any other measurements.
Measurements:sgp5ebbrE19.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE19.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE19.b1:
  • Barometric Pressure(atmos_pressure)


Back To Table of Contents

DQRID : D030117.4
Start DateStart TimeEnd DateEnd Time
07/14/2002143001/09/20031600
Subject:
SGP/EBBR/E19 - Soil Temperatures #2 and #3 Intermittently Bad
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperatures #2 and #3 were intermittently incorrect.
#3 appeared to be okay after the J-panel was replaced. 
#2 continues to be flaky, suggesting that the sensor
itself has a problem.  A DQPR has been submitted to
replace #2.  The indicated measurements 
that are calculated using soil temperature are 
also suspect.  However, the intermittent problems normally
occurred during night times and were fairly small in magnitude,
thereby having little effect on the sensible and latent heat fluxes.
Measurements:sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)

sgp30ebbrE19.b1:
  • energy_storage_change_3
  • energy_storage_change_2
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • soil_temp_3
  • surface_soil_heat_flux_2
  • surface_soil_heat_flux_avg
  • latent_heat_flux
  • soil_temp_2


Back To Table of Contents

DQRID : D030117.6
Start DateStart TimeEnd DateEnd Time
01/09/2003170001/09/20031729
Subject:
SGP/EBBR/E19 - Measurements zero following logger restart
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.  The restart on 01/09/03 occurred after J-panel
replacements.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D030117.7
Start DateStart TimeEnd DateEnd Time
01/09/2003173001/09/20031759
Subject:
SGP/EBBR/E19 - Change of Energy Storage Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Because soil temperature measurements were not available 
from the beginning of the half hour, the change in soil 
energy storage could not be calculated and thus is 
incorrect.

There is no fix to this problem. Average Soil Heat Flow, 
Sensible Heat Flux (h), and Latent Heat Flux (e) can not 
be calculated using the available, remaining measurements.
Measurements:sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 5(g5)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • energy_storage_change_2
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • surface_soil_heat_flux_5
  • energy_storage_change_1
  • energy_storage_change_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • latent_heat_flux
  • energy_storage_change_5
  • energy_storage_change_4
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D030711.11
Start DateStart TimeEnd DateEnd Time
01/09/2003163001/09/20031700
Subject:
SGP/EBBR/E19 - Soil Temperature #2 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature #2 was incorrect during J-panel replacements.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • soil_temp_2_rr

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • energy_storage_change_2
  • surface_soil_heat_flux_2
  • surface_soil_heat_flux_avg
  • latent_heat_flux
  • soil_temp_2


Back To Table of Contents

DQRID : D030711.12
Start DateStart TimeEnd DateEnd Time
01/09/2003170001/09/20031700
02/06/2003173002/06/20031730
04/03/2003183004/03/20031830
Subject:
SGP/EBBR/E19 - All Data Zeros
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
All data (5, 15, and 30 minute) was zero after replacement of sensors and restart of the 
CR10.

This is a common occurrence.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • wdir_vec_mean_mv
  • net_radiation_mv
  • rh_left_mv
  • temp_trh_left_rr
  • rh_right_mv
  • soil_moisture_3_rr
  • soil_moisture_1_rr
  • temp_reference_rr
  • soil_moisture_5_rr
  • soil_temp_1_rr
  • atmos_pressure_mv
  • soil_heat_flow_5_mv
  • soil_heat_flow_2_mv
  • soil_heat_flow_1_mv
  • soil_temp_4_rr
  • wspd_arith_mean
  • soil_heat_flow_3_mv
  • soil_moisture_4_rr
  • soil_moisture_2_rr
  • soil_temp_3_rr
  • temp_trh_right_rr
  • soil_temp_5_rr
  • battery_voltage
  • temp_air_left
  • soil_heat_flow_4_mv
  • home_signal
  • soil_temp_2_rr
  • temp_air_right

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Retrieved pressure profile(pres)
  • Soil heat capacity 5(cs5)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • soil_moisture_4
  • wspd_arith_mean
  • vapor_pressure_bottom
  • rh_top_fraction
  • rh_bottom_fraction
  • vapor_pressure_top
  • wdir_vec_mean
  • latent_heat_flux
  • energy_storage_change_2
  • soil_heat_flow_5
  • soil_moisture_3
  • temp_trh_bottom
  • home_signal_15
  • energy_storage_change_1
  • temp_air_top
  • wspd_vec_mean
  • soil_moisture_2
  • soil_moisture_5
  • soil_heat_flow_2
  • corr_soil_heat_flow_5
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • corr_soil_heat_flow_2
  • corr_soil_heat_flow_3
  • soil_heat_flow_1
  • bowen_ratio
  • corr_soil_heat_flow_4
  • home_signal_30
  • soil_heat_capacity_4
  • soil_heat_capacity_5
  • energy_storage_change_3
  • net_radiation
  • soil_temp_3
  • soil_temp_1
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • soil_heat_flow_4
  • soil_heat_capacity_3
  • energy_storage_change_4
  • Barometric Pressure(atmos_pressure)
  • soil_temp_2
  • sensible_heat_flux
  • soil_heat_capacity_1
  • soil_heat_capacity_2
  • soil_temp_4
  • surface_soil_heat_flux_5
  • soil_temp_5
  • wdir_vec_std
  • soil_moisture_1
  • temp_trh_top
  • corr_soil_heat_flow_1
  • temp_reference
  • temp_air_bottom
  • energy_storage_change_5
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D030711.13
Start DateStart TimeEnd DateEnd Time
01/09/2003173001/09/20031730
Subject:
SGP/EBBR/E19 - Several Soil Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
During replacement of a J-panel, several soil measurements were incorrect.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 3(rr_ts3)

sgp15ebbrE19.b1:
  • soil_temp_3_rr

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 5(g5)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • energy_storage_change_2
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • surface_soil_heat_flux_5
  • energy_storage_change_1
  • energy_storage_change_3
  • soil_temp_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • latent_heat_flux
  • energy_storage_change_5
  • energy_storage_change_4
  • surface_soil_heat_flux_4


Back To Table of Contents

DQRID : D030711.14
Start DateStart TimeEnd DateEnd Time
01/26/2003170001/26/20031700
01/26/2003180001/26/20031800
Subject:
SGP/EBBR/E19 -Several Soil Measurements Were Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Several soil measurements were incorrect during replacement of soil temperature probe #2.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • soil_temp_2_rr

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 5(g5)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • energy_storage_change_2
  • surface_soil_heat_flux_3
  • surface_soil_heat_flux_2
  • surface_soil_heat_flux_5
  • energy_storage_change_1
  • energy_storage_change_3
  • surface_soil_heat_flux_avg
  • surface_soil_heat_flux_1
  • latent_heat_flux
  • energy_storage_change_5
  • energy_storage_change_4
  • surface_soil_heat_flux_4
  • soil_temp_2


Back To Table of Contents

DQRID : D030711.17
Start DateStart TimeEnd DateEnd Time
01/09/2003173004/03/20031900
Subject:
SGP/EBBR/E19 - Soil Heat Flow #3 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #3 was a flat -24 W/m2 for most of this period.
Measurements:sgp15ebbrE19.a1:
  • Soil heat flow 3(mv_hft3)

sgp15ebbrE19.b1:
  • soil_heat_flow_3_mv

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow, site 3(shf3)
  • h(h)

sgp30ebbrE19.b1:
  • sensible_heat_flux
  • surface_soil_heat_flux_3
  • soil_heat_flow_3
  • surface_soil_heat_flux_avg
  • latent_heat_flux
  • corr_soil_heat_flow_3


Back To Table of Contents

DQRID : D040525.4
Start DateStart TimeEnd DateEnd Time
04/01/2001000005/06/20032359
Subject:
SGP/EBBR/E19 - Metadata errors
DataStreams:sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The latitude, longitude and altitude of the El Reno (E19) site were
incorrectly entered into the ARM database.  The correct location
of the EBBR.E19 instrument is:
    Lat: 35.557N
    Lon: 98.017W
    Alt:    421m
Measurements:sgp15ebbrE19.b1:
  • lat(lat)
  • Dummy altitude for Zeb(alt)
  • lon(lon)

sgp5ebbrE19.a1:
  • Dummy altitude for Zeb(alt)
  • lat(lat)
  • lon(lon)

sgp30ebbrE19.b1:
  • Dummy altitude for Zeb(alt)
  • lon(lon)
  • lat(lat)


Back To Table of Contents

DQRID : D050719.12
Start DateStart TimeEnd DateEnd Time
04/01/2001000003/31/20051600
Subject:
SGP/EBBR/E19 - Improved EBBR CR10 Program
DataStreams:sgp30ebbrE19.b1
Description:
Effective 20050331.1600, the EBBR.E19 CR10 program was revised to improve 
the quality of the primary measurements as follows:
   
1) An RMS procedure is used to determine max and min limits of acceptable
soil heat flow.  This is applied to the individual soil sets.  Measurements
outside the limits are rejected and the measurements within the limits are
used to calculate the average soil heat flow.
   
2) Max and min limits are used to determine incorrect values of net
radiation, sensible heat flux (h), latent heat flux (e), and automatic
exchange mechanism (AEM) signal.  If AEM signal is outside the limits, h 
and e are set to 999s.  If net radiation is outside the limits, h and e are 
set to 999s. If h or e are outside the limits, h and e are set to 999s.
   
Virtually no incorrect soil measurement will affect the primary 
measurements of h and e.
   
By setting h and e to 999s, it can be easily seen that the primary 
variables are incorrect; no other interpretation is possible.
   
Prior to 20050331, the improved CR10 program was NOT in effect.  DQRs have
been submitted for known instances of incorrect soil measurements which
affected the quality of the primary measurements of h and e.
   
Note: the DQR begin date is the begin date of the sgp30ebbrE19.b1 data
stream.  The earlier version of the CR10 program was also used on previous
EBBR datastream names (e.g. sgp30ebbrE19.a1).
Measurements:sgp30ebbrE19.b1:
  • sensible_heat_flux
  • latent_heat_flux
  • surface_soil_heat_flux_avg


Back To Table of Contents



END OF DATA