DQR ID | Subject | Data Streams Affected |
---|---|---|
D051214.2 | SGP/EBBR/E19 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D051214.3 | SGP/EBBR/E19 - Incorrect Sensible and Latent Heat Fluxes | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
D080527.2 | SGP/EBBR/E19 - Sensible and latent heat flux and net radiation incorrect | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2005 | 0600 | 10/25/2005 | 0100 |
10/31/2005 | 1030 | 12/20/2005 | 1940 |
Subject: | SGP/EBBR/E19 - Net Radiation, Sensible and Latent Heat Fluxes Incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The net radiometer cable was chewed by an animal. Net Radiation, Sensible Heat Flux, and Latent Heat Flux were incorrect during the stated periods. |
Measurements: | sgp5ebbrE19.b1:
sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2005 | 0000 | 01/19/2006 | 1630 |
Subject: | SGP/EBBR/E19 - Incorrect Sensible and Latent Heat Fluxes |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | The gradient measurements and latent and sensible heat fluxes were incorrect because the AEM was not exchanging. The problem was fixed by replacing the AEM. |
Measurements: | sgp5ebbrE19.b1:
sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/07/2008 | 0600 | 05/07/2008 | 0900 |
05/07/2008 | 1530 | 05/07/2008 | 2130 |
05/09/2008 | 1330 | 05/09/2008 | 1330 |
Subject: | SGP/EBBR/E19 - Sensible and latent heat flux and net radiation incorrect |
DataStreams: | sgp15ebbrE19.b1, sgp30ebbrE19.b1, sgp5ebbrE19.b1 |
Description: | Net radiation was too large, resulting in sensible and latent heat fluxes being too large. The cause of the problem is unknown. |
Measurements: | sgp5ebbrE19.b1:
sgp15ebbrE19.b1:
sgp30ebbrE19.b1:
|