DQR ID | Subject | Data Streams Affected |
---|---|---|
D070412.3 | TWP/MWR/C1/C2 - Sun in the field of view | twpmwrlosC1.b1, twpmwrlosC2.b1, twpmwrtipC1.a1, twpmwrtipC2.a1 |
D080707.2 | TWP/SMET/C1 - ORG Precip Data Incorrect | twpsmet60sC1.b1 |
D081013.2 | TWP/MWR/C1 - Short periods of wrong ambient temperature | twpmwrlosC1.b1, twpmwrtipC1.a1 |
D081022.3 | TWP/SMET/C1 - Wind Speed data contain offset due to induced voltage | twpsmet60sC1.b1 |
D090114.4 | TWP/MWR/C1 - Ambient temperature data missing | twpmwrlosC1.b1, twpmwrtipC1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/07/2007 | 0000 | 04/02/2007 | 2359 |
03/01/2008 | 0000 | 04/02/2008 | 2359 |
09/10/2008 | 0000 | 10/12/2008 | 0000 |
03/01/2009 | 0000 | 04/02/2009 | 2359 |
Subject: | TWP/MWR/C1/C2 - Sun in the field of view |
DataStreams: | twpmwrlosC1.b1, twpmwrlosC2.b1, twpmwrtipC1.a1, twpmwrtipC2.a1 |
Description: | Every day between around noon local time (usually between 1 and 2 AM UTC time) there is an increase in the brightness temperatures due to the sun in the field of vie of the radiometer. |
Measurements: | twpmwrlosC2.b1:
twpmwrlosC1.b1:
twpmwrtipC1.a1:
twpmwrtipC2.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2008 | 1300 | 02/09/2008 | 0600 |
Subject: | TWP/SMET/C1 - ORG Precip Data Incorrect |
DataStreams: | twpsmet60sC1.b1 |
Description: | Moisture intrusion into the sensor body allowed condensation on the lens of the ORG. Continual and long-term precipitation was reported during this time. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/2008 | 0000 | 09/30/2008 | 2359 |
Subject: | TWP/MWR/C1 - Short periods of wrong ambient temperature |
DataStreams: | twpmwrlosC1.b1, twpmwrtipC1.a1 |
Description: | During this month there are a few intermittent instances of bad ambient temperature readings. I will summarize here these short time ranges: 20080909 between 02:00 and 03:00 UTC 20080909 between 09:00 and 14:00 UTC Additionally the ambient temperature has short but frequent data gaps. The data gaps are very short (less than one hour). Users could interpolate missing data or use alternative sensors such as the surface meteorological data available for the site. |
Measurements: | twpmwrlosC1.b1:
twpmwrtipC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 11/03/2008 | 0730 |
Subject: | TWP/SMET/C1 - Wind Speed data contain offset due to induced voltage |
DataStreams: | twpsmet60sC1.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/2008 | 1900 | 02/06/2009 | 0530 |
Subject: | TWP/MWR/C1 - Ambient temperature data missing |
DataStreams: | twpmwrlosC1.b1, twpmwrtipC1.a1 |
Description: | The ambient temperature readings (tkair) suddenly failed. The sensor was replaced on Feb 6. |
Measurements: | twpmwrlosC1.b1:
twpmwrtipC1.a1:
|