DQR ID | Subject | Data Streams Affected |
---|---|---|
D061114.4 | TWP/MWR/C3 - Sun in field of view of radiometer | twpmwrlosC3.b1, twpmwrtipC3.a1 |
D070306.2 | TWP/SMET/C3 - Barometer Data Incorrect | twpsmet60sC3.b1 |
D070306.3 | TWP/SMET/C3 - Water in enclosure | twpsmet60sC3.b1 |
D081022.8 | TWP/SMET/C3 - Wind Speed data contain offset due to induced voltage | twpsmet60sC3.b1 |
D090515.1 | TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degrees | twpsmet60sC3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/20/2006 | 0000 | 11/09/2006 | 0400 |
10/15/2008 | 0000 | 11/08/2008 | 0000 |
Subject: | TWP/MWR/C3 - Sun in field of view of radiometer |
DataStreams: | twpmwrlosC3.b1, twpmwrtipC3.a1 |
Description: | Each day from approximately 9/15 to 10/9 around local noon, there is an increase in the brightness temperature due to the sun being in the field of view of the radiometer. |
Measurements: | twpmwrlosC3.b1:
twpmwrtipC3.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2007 | 1723 | 03/03/2007 | 1228 |
Subject: | TWP/SMET/C3 - Barometer Data Incorrect |
DataStreams: | twpsmet60sC3.b1 |
Description: | The barometer data was incorrect or missing due to water intrusion. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/03/2007 | 0056 | 03/03/2007 | 0123 |
Subject: | TWP/SMET/C3 - Water in enclosure |
DataStreams: | twpsmet60sC3.b1 |
Description: | Data became erratic and unreasonable after water intrusion into the enclosure. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0000 | 10/17/2008 | 0627 |
Subject: | TWP/SMET/C3 - Wind Speed data contain offset due to induced voltage |
DataStreams: | twpsmet60sC3.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/19/2005 | 0420 | 04/04/2009 | 0115 |
Subject: | TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degrees |
DataStreams: | twpsmet60sC3.b1 |
Description: | The temperature of the SMET system was biased low by 1.06 Deg C due to a non-precision resistor being used int he measurement circuit. The data will be reprocessed. Saturation Vapor pressure and Vapor pressure will have to be reprocessed also. Add 1.06 degrees to the temperature data. Correct Sat Vap Press using the Buck approximation: Es = 6.1121exp((17.502*T)/(240.97 + T) where T is the temperature in C. Correct Vapor pressure using: E = (RH*Es)/100 |
Measurements: | twpsmet60sC3.b1:
|