Description: | Even though the polarization voltage was reported as switching between circular and cross
polarization states, there was no cross polarization taking place because the
polarization box had not been recycled following a system reboot.
You will not be able to create depolarization values. However all other processes, such
as cloud base determination should be ok. |
Measurements: | nsamplpolC1.b1: - Attenuated backscatter(signal_return)
- Polarization control voltage setting(polarization_control_voltage)
nsamplpolavgC1.c1: - Background signal, cross-pol(background_signal_cross_pol)
- *(aqc_signal_return_cross_pol)
- Number of samples used for cross-pol averages(nsamples_cross_pol_average)
- Background signal standard deviation, cross-pol(background_signal_std_cross_pol)
- Preliminary cloud base height above ground level, cross-pol(preliminary_cbh_cross_pol)
- Polarization control voltage, cross-pol(polarization_control_voltage_cross_pol)
- Attenuated backscatter, cross-pol(signal_return_cross_pol)
nsamplpolavgC1.s1: - Preliminary cloud base height above ground level, cross-pol(preliminary_cbh_cross_pol)
- Background signal standard deviation, cross-pol(background_signal_std_cross_pol)
- Background signal, cross-pol(background_signal_cross_pol)
- *(aqc_signal_return_cross_pol)
- Polarization control voltage, cross-pol(polarization_control_voltage_cross_pol)
- Attenuated backscatter, cross-pol(signal_return_cross_pol)
|