DQR ID | Subject | Data Streams Affected |
---|---|---|
D050104.1 | TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect | twpskyrad60sC3.b1, twpskyradC3.00 |
D060120.1 | TWP/SKYRAD/C1 - IRT serial number incorrect | twpskyradC1.00 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/20/2004 | 0000 | 09/02/2004 | 0400 |
Subject: | TWP/SKYRAD/C3 - Reprocess: NIP calibrations incorrect |
DataStreams: | twpskyrad60sC3.b1, twpskyradC3.00 |
Description: | On 05/20 during installtion of the new Campbell datalogger system and associated program the NIP (S/N 33551E6) was assigned an incorrect calibration coefficient. A coefficient of 116.01 W/m2/mV was applied when it should have been 124.378 W/m2/mV. TWP reports SV0405D and SV0409D detail the activities which lead to this DQR. |
Measurements: | twpskyradC3.00:
twpskyrad60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/02/2004 | 2340 | 01/18/2006 | 2300 |
Subject: | TWP/SKYRAD/C1 - IRT serial number incorrect |
DataStreams: | twpskyradC1.00 |
Description: | The serial number was incorrect in the csi program for the skyrad. It was entered as 868 when the correct number was 866 until 10 jan 2006 at 0145. It then should have been 1845. This does not affect data other than the once per day serial number and calibration field. |
Measurements: | twpskyradC1.00:
|