DQR ID | Subject | Data Streams Affected |
---|---|---|
D071010.2 | SGP/THWAPS/B5 - Reprocess: RH values 10% Above Actual | sgpthwapsB5.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/17/2007 | 1920 | 10/11/2007 | 1620 |
Subject: | SGP/THWAPS/B5 - Reprocess: RH values 10% Above Actual |
DataStreams: | sgpthwapsB5.b1 |
Description: | The T/RH probe was replaced with a recently calibrated sensor. The newly installed probe did not have the correct span 0-110% RH for output. This caused the readings to artificially high by 10%. |
Measurements: | sgpthwapsB5.b1:
|