DQR ID | Subject | Data Streams Affected |
---|---|---|
D040617.2 | SGP/ECOR/E1 - Missing data | sgp30ecorE1.b1 |
D041025.2 | SGP/ECOR/E1 - System time drift | sgp30ecorE1.b1, sgpecorE1.00 |
D051101.6 | SGP/ECOR/E1 - ECOR Pressure and Temperature Missing | sgp30ecorE1.b1 |
D051101.7 | SGP/ECOR/E1 - CO2 and H2O Fluxes Incorrect | sgp30ecorE1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/15/2004 | 0900 | 09/10/2004 | 1800 |
10/03/2004 | 1500 | 10/05/2004 | 1830 |
05/24/2005 | 0800 | 06/07/2005 | 1800 |
09/06/2005 | 1900 | 09/20/2005 | 2030 |
05/31/2006 | 2000 | 08/08/2006 | 2230 |
Subject: | SGP/ECOR/E1 - Missing data |
DataStreams: | sgp30ecorE1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgp30ecorE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2004 | 2030 | 10/25/2004 | 1735 |
Subject: | SGP/ECOR/E1 - System time drift |
DataStreams: | sgp30ecorE1.b1, sgpecorE1.00 |
Description: | Network Time Protocol (NTP) client was not working at the ECOR computer resulting in an ECOR system time drift of about 6s per hour. Time adjustment by +15 minutes occurred at 17:35:25, Oct. 25, 2004. |
Measurements: | sgpecorE1.00:
sgp30ecorE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2005 | 1400 | 05/31/2006 | 1930 |
Subject: | SGP/ECOR/E1 - ECOR Pressure and Temperature Missing |
DataStreams: | sgp30ecorE1.b1 |
Description: | The pressure and temperature from the CO2/H2O sensor is missing due to a failure of some kind. |
Measurements: | sgp30ecorE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/18/2005 | 2000 | 10/21/2005 | 1800 |
Subject: | SGP/ECOR/E1 - CO2 and H2O Fluxes Incorrect |
DataStreams: | sgp30ecorE1.b1 |
Description: | The CO2 and H2O fluxes (lv_e) were incorrect because of a LI-7500 failure. |
Measurements: | sgp30ecorE1.b1:
|