Subject: | SGP/SMOS/E1 - RH high when near saturated |
DataStreams: | sgp1440smosE1.b1, sgp1smosE1.b1, sgp30smosE1.b1
|
Description: | Instrument RH values consistently exceeded MAX and did not compare well to surrounding
sites. Problem was apparent only when local RH values were in excess of 90%, usually
between 02-16Z. |
Measurements: | sgp30smosE1.b1: - qc_rh
- qc_vap_pres
- Standard Deviation of Vapor Pressure(sd_vap_pres)
- Vapor Pressure (kiloPascals)(vap_pres)
- Standard Deviation of Relative Humidity(sd_rh)
- qc_sd_vap_pres
- qc_sd_rh
- Relative humidity inside the instrument enclosure(rh)
sgp1440smosE1.b1: - Time of Minimum Relative Humidity(time_min_rh)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Time of Maximum Relative Humidity(time_max_rh)
- Minimum Vapor Pressure(min_vap_pres)
- Minimum Relative Humidity(min_rh)
- Maximum Vapor Pressure(max_vap_pres)
- Maximum Relative Humidity(max_rh)
sgp1smosE1.b1: - Relative humidity inside the instrument enclosure(rh)
- qc_rh
- qc_vap_pres
- Vapor Pressure (kiloPascals)(vap_pres)
|