Description: | Two sensor connections were switched. The East 125-cm sensor was plugged into the East
85-cm sensor's connection, and vice versa. Technicians corrected the problem on 1440 GMT,
08/02/00, and returned the instrument to service 10 minutes later at 1450 GMT. As the
datalogger was confusing the 2 sensors, incorrect calibrations were applied.
The correct calibrations can be applied to the raw data to obtain the correct temperature
rises, soil moisture, etc. However, this is a rather large amount of work to apply toward
just 1 measurement profile. This has been identified as a medium priority reprocessing
task. |
Measurements: | sgpswatsE1.a1: - East profile sensor serial numbers(serial_numbers_E)
- Soil Water Potential, East Profile(soilwatpot_E)
- Volumetric Water Content, East Profile(watcont_E)
- Sensor Temperature Rise, East Profile(trise_E)
- Soil Temperature, East Profile(tsoil_E)
sgpswatsE1.00: - null(Raw data stream - documentation not supported)
sgpswatsE1.b1: - Volumetric Water Content, East Profile(watcont_E)
- Soil Temperature, East Profile(tsoil_E)
- Sensor Temperature Rise, East Profile(trise_E)
- East profile sensor serial numbers(serial_numbers_E)
- Soil Water Potential, East Profile(soilwatpot_E)
|