DQR ID | Subject | Data Streams Affected |
---|---|---|
D020814.9 | NSA/SKYRAD/C1 - Reprocess: PIR identified as unshaded is really shaded | nsaskyrad60sC1.b1 |
D021111.1 | NSA/SKYRAD/C1 - Unshaded PIR data unreasonable | nsaskyrad60sC1.a1 |
D021111.11 | NSA/SKYRAD/C1 - PIR data suspect | nsaskyrad60sC1.a1 |
D021111.14 | NSA/SKYRAD/C1 - Shaded PIR data unreasonable | nsaskyrad60sC1.a1 |
D021111.15 | NSA/SKYRAD/C1 - Corrupted Shaded PIR data | nsaskyrad60sC1.a1 |
D021111.16 | NSA/SKYRAD/C1 - Shaded PIR data unreasonable | nsaskyrad60sC1.a1 |
D021111.17 | NSA/SKYRAD/C1 - Unshaded and Shaded PIR data bad | nsaskyrad60sC1.a1 |
D021111.19 | NSA/SKYRAD/C1 - PIR data suspect | nsaskyrad60sC1.a1 |
D021111.2 | NSA/SKYRAD/C1 - Corrupted Unshaded PIR data | nsaskyrad60sC1.a1 |
D021111.21 | NSA/SKYRAD/C1 - Shaded PIR data suspect | nsaskyrad60sC1.a1 |
D021111.22 | NSA/SKYRAD/C1 - PIR data suspect | nsaskyrad60sC1.a1 |
D021111.23 | NSA/SKYRAD/C1 - Shaded PIR data incorrect | nsaskyrad60sC1.a1 |
D021111.3 | NSA/SKYRAD/C1 - Spikes in Unshaded PIR data | nsaskyrad60sC1.a1 |
D021111.4 | NSA/SKYRAD/C1 - Discontinuity in Unshaded PIR data | nsaskyrad60sC1.a1 |
D021111.5 | NSA/SKYRAD/C1 - Unshaded PIR data incorrect | nsaskyrad60sC1.a1 |
D021211.10 | NSA/GNDRAD/C1 - Unreasonable/missing PIR data due to maintenance | nsagndrad60sC1.a1 |
D021211.11 | NSA/GNDRAD/C1 - Noisy PIR Data | nsagndrad60sC1.a1 |
D021211.13 | NSA/GNDRAD/C1 - PIR1 Calibration problems resulted in small error | nsagndrad60sC1.a1 |
D021211.8 | NSA/GNDRAD/C1 - PIR offline | nsagndrad60sC1.a1 |
D021211.9 | NSA/GNDRAD/C1 - Instrument maintenance | nsagndrad60sC1.a1 |
D021212.5 | NSA/GNDRAD/C1 - Bad PSP calibrations applied | nsagndrad60sC1.a1 |
D021213.1 | NSA/SKYRAD/C1 - NIP shaded by skyrad stand around local midnight | nsaskyrad60sC1.a1 |
D021213.3 | NSA/SKYRAD/C1 - PSP shaded by MFRSR after local midnight | nsaskyrad60sC1.a1 |
D030301.1 | NSA/GNDRAD/C1 - No power to IRT | nsagndrad60sC1.a1, nsagndrad60sC1.b1 |
D030510.2 | NSA/SKYRAD/C1 - IRT measurements during precipitation events | nsaskyrad60sC1.a1 |
D030819.2 | NSA/SKYRAD/C1 - IRT failed | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
D030820.1 | NSA/SKYRAD/C1 - IRT incorrectly configured | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
D030821.2 | NSA/SKYRAD/C1 - IRT incorrect min/max value | nsaskyrad60sC1.b1 |
D031007.2 | NSA/SKYRAD/C1 - IRT measurements during precipitation events | nsaskyrad60sC1.b1 |
D031216.1 | NSA/GNDRAD/C1 - IRT cable failure | nsagndrad60sC1.b1 |
D031216.2 | NSA/GNDRAD/C1 - IRT returned to vendor | nsagndrad60sC1.b1 |
D040213.4 | NSA/SKYRAD/C1 - Downwelling IRT bias | nsaskyrad60sC1.b1 |
D040421.2 | NSA/SKYRAD/C1 - Tracker Inoperative | nsaskyrad60sC1.b1 |
D040421.3 | NSA/SKYRAD/C1 - itermittant IRT bias | nsaskyrad60sC1.b1 |
D040919.1 | NSA/GNDRAD/C1 - IRT power failure | nsagndrad60sC1.b1 |
D051101.2 | NSA/SKYRAD/C1 - IRT failure | nsaskyrad60sC1.b1 |
D060209.4 | NSA/GNDRAD/C1 - power outage affecting some GNDRAD sensors | nsagndrad20sC1.a0, nsagndrad60sC1.b1, nsagndradC1.00 |
D060209.5 | NSA/SKYRAD/C1 - Power outage affecting some SKYRAD instruments | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1, nsaskyradC1.00 |
D060630.39 | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error | nsagndrad60sC1.b1 |
D071018.1 | NSA/SKYRAD/C1 - Missing data | nsaskyrad60sC1.b1 |
D071018.2 | NSA/GNDRAD/C1 - Missing data | nsagndrad60sC1.b1 |
D071025.1 | NSA/SKYRAD/C1 - Reprocessed: Incorrect calibration coefficient | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D071025.3 | NSA/GNDRAD/C1 - Reprocessed: Incorrect calibration coefficient | nsagndrad20sC1.a0, nsagndrad60sC1.b1 |
D080107.2 | NSA/SKYRAD/C1 - Instrument cleaning effects | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080222.3 | NSA/SKYRAD/C1 - IRT mirror heater failure | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080418.1 | NSA/SKYRAD/C1 - Direct and Diffuse incorrect | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080630.1 | NSA/SKYRAD/C1 - Tracker inoperative | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080725.1 | NSA/GNDRAD/C1 - IRT not connected | nsagndrad60sC1.a1 |
D080725.2 | NSA/GNDRAD/C1 - Reprocess: IRT scale factor | nsagndrad60sC1.a1 |
D080726.1 | NSA/GNDRAD/C1 - IRT failed | nsagndrad60sC1.a1 |
D080728.1 | NSA/GNDRAD/C1 - IRT incorrect | nsagndrad60sC1.a1 |
D080729.2 | NSA/GNDRAD/C1 - IRT scale factor | nsagndrad60sC1.a1 |
D080825.1 | NSA/SKYRAD/C1 - IRT signal bias | nsaskyrad60sC1.a1 |
D080904.5 | NSA/GNDRAD/C1 - Upwelling PIR ventilator fan failure | nsagndrad20sC1.a0, nsagndrad60sC1.b1 |
D080919.1 | NSA/GNDRAD/C1 - Net radiometer removed | nsagndrad60sC1.a1, nsagndrad60sC1.b1 |
D090209.2 | NSA/GNDRAD/C1 - Reprocess: IRT insufficiently heated | nsagndrad60sC1.b1 |
D090209.4 | NSA/SKYRAD/C1 - Reprocess: IRT insufficiently heated | nsaskyrad60sC1.b1 |
D090318.1 | NSA/SKYRAD/C1 - IRT covered with snow | nsaskyrad60sC1.b1 |
D981212.1 | NSA/SKYRAD - Barrow SKYRAD PIRD Misconfiguration 12/9 - 12/12/98 | nsaskyrad20sC1.a1, nsaskyrad60sC1.a1 |
D990108.2 | NSA/SKYRAD/C1 - IRT removed for repair | nsaskyrad60sC1.a1, nsaskyradC1.00 |
D990108.5 | NSA/SKYRAD/C1 - IRT removed from service | nsaskyrad60sC1.a1, nsaskyradC1.00 |
D990924.1 | NSA/GNDRAD/C1 - Reprocess: IRT insufficiently heated | nsagndrad60sC1.a1 |
D990924.2 | NSA/SKYRAD/C1 - Reprocess: IRT insufficiently heated | nsaskyrad60sC1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/15/2002 | 1905 | 06/24/2002 | 2359 |
Subject: | NSA/SKYRAD/C1 - Reprocess: PIR identified as unshaded is really shaded |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | On 5/15/02 the Global PIR was moved from the unshaded position on the Skyrad stand to a shaded spot on the tracker. It is still using the same Logger channel as it did as an unshaded PIR even though it is now shaded. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/1999 | 2200 | 01/12/1999 | 2230 |
01/17/1999 | 0500 | 01/17/1999 | 0730 |
01/24/1999 | 1200 | 01/24/1999 | 1400 |
02/13/1999 | 0700 | 02/13/1999 | 2359 |
02/19/1999 | 0545 | 02/20/1999 | 1030 |
02/20/1999 | 2115 | 02/21/1999 | 0200 |
Subject: | NSA/SKYRAD/C1 - Unshaded PIR data unreasonable |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Periods of very large negative values. Negative readings of thermistors caused concurrent calculation in logger of temperature to fail. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/1998 | 0300 | 10/28/1998 | 0500 |
10/29/1998 | 0600 | 10/29/1998 | 0800 |
11/07/1998 | 0100 | 11/07/1998 | 0400 |
11/12/1998 | 0300 | 11/12/1998 | 1200 |
11/23/1998 | 0900 | 11/23/1998 | 2100 |
Subject: | NSA/SKYRAD/C1 - PIR data suspect |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | PIR shaded and PIR unshaded show unusual differences during this time period, frosting or icing is the most probable explanation but we don't know for sure. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/1999 | 2200 | 01/12/1999 | 2230 |
01/17/1999 | 0500 | 01/17/1999 | 0730 |
01/24/1999 | 1200 | 01/24/1999 | 1400 |
02/13/1999 | 0700 | 02/13/1999 | 2359 |
02/19/1999 | 0545 | 02/20/1999 | 1030 |
02/20/1999 | 2115 | 02/21/1999 | 0200 |
Subject: | NSA/SKYRAD/C1 - Shaded PIR data unreasonable |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Periods of very large negative values. Negative readings of thermistors causes concurrent calculation in logger of temperature to fail. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/18/1998 | 2105 | 11/18/1998 | 2110 |
11/20/1998 | 0340 | 11/20/1998 | 0350 |
11/20/1998 | 2010 | 11/20/1998 | 2020 |
11/24/1998 | 0330 | 11/24/1998 | 0340 |
11/25/1998 | 0145 | 11/25/1998 | 0150 |
11/25/1998 | 2030 | 11/25/1998 | 2035 |
11/26/1998 | 0300 | 11/26/1998 | 0310 |
11/26/1998 | 1530 | 11/26/1998 | 1540 |
11/27/1998 | 0110 | 11/27/1998 | 0120 |
11/27/1998 | 0840 | 11/27/1998 | 0850 |
11/27/1998 | 0925 | 11/27/1998 | 0935 |
11/27/1998 | 1940 | 11/27/1998 | 1950 |
11/28/1998 | 0655 | 11/28/1998 | 0705 |
11/29/1998 | 0450 | 11/29/1998 | 0505 |
11/29/1998 | 2335 | 11/29/1998 | 2350 |
Subject: | NSA/SKYRAD/C1 - Corrupted Shaded PIR data |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Erroneous data due to file truncation problem during logger download and corresponding ingest problems when the truncated lines were read. Problem was identified in a PIF submitted by Rune Storvold |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/1998 | 2225 | 12/11/1998 | 2235 |
12/15/1998 | 1800 | 12/15/1998 | 1810 |
12/16/1998 | 1810 | 12/16/1998 | 1815 |
02/16/1999 | 1945 | 02/16/1999 | 2010 |
02/17/1999 | 2145 | 02/17/1999 | 2200 |
03/02/1999 | 1825 | 03/02/1999 | 1835 |
03/09/1999 | 1830 | 03/09/1999 | 1840 |
03/10/1999 | 0340 | 03/10/1999 | 0350 |
Subject: | NSA/SKYRAD/C1 - Shaded PIR data unreasonable |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | There was a 10-25 W/m2 spike in data probably due to operator working around instrument or cleaning it. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/17/1999 | 1700 | 01/17/1999 | 1815 |
Subject: | NSA/SKYRAD/C1 - Unshaded and Shaded PIR data bad |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Large unreasonable spikes and some missing data points. Cause of problem unknown. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/28/1998 | 0300 | 10/28/1998 | 0500 |
10/29/1998 | 0600 | 10/29/1998 | 0800 |
11/07/1998 | 0100 | 11/07/1998 | 0400 |
11/12/1998 | 0300 | 11/12/1998 | 1200 |
11/23/1998 | 0900 | 11/23/1998 | 2100 |
Subject: | NSA/SKYRAD/C1 - PIR data suspect |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | PIR shaded and PIR unshaded show unusual differences during this time period, frosting or icing is one possible explanation but we don't know. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/1998 | 2000 | 11/14/1998 | 2005 |
11/18/1998 | 2105 | 11/18/1998 | 2110 |
11/20/1998 | 0340 | 11/20/1998 | 0350 |
11/20/1998 | 2010 | 11/20/1998 | 2020 |
11/24/1998 | 0330 | 11/24/1998 | 0340 |
11/25/1998 | 0145 | 11/25/1998 | 0150 |
11/25/1998 | 2030 | 11/25/1998 | 2035 |
11/26/1998 | 0300 | 11/26/1998 | 0310 |
11/26/1998 | 1530 | 11/26/1998 | 1540 |
11/27/1998 | 0110 | 11/27/1998 | 0120 |
11/27/1998 | 0840 | 11/27/1998 | 0850 |
11/27/1998 | 0925 | 11/27/1998 | 0935 |
11/27/1998 | 1940 | 11/27/1998 | 1950 |
11/28/1998 | 0655 | 11/28/1998 | 0705 |
11/29/1998 | 0450 | 11/29/1998 | 0505 |
11/29/1998 | 2335 | 11/29/1998 | 2350 |
Subject: | NSA/SKYRAD/C1 - Corrupted Unshaded PIR data |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Erroneous data due to file truncation problem during logger download and corresponding ingest problems when the truncated lines were read. Problem was identified in a PIF submitted by Rune Storvold |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/1998 | 0000 | 12/15/1998 | 1200 |
Subject: | NSA/SKYRAD/C1 - Shaded PIR data suspect |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | The instrument had incorrect calibration settings on two accounts: 1. From 12/09 to 12/12, the calibration was set incorrectly. 2. The dome correction coefficients were not properly set until the early morning of 12/14. The erroneous settings caused an error probably less than 4%. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/14/1998 | 0000 | 02/04/1999 | 2359 |
Subject: | NSA/SKYRAD/C1 - PIR data suspect |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | The PIR unshaded and the PIR shaded have a persistent offset between them, ranging from 0 to 10 W/m2. It is particularly apparent during clear sky. The line of sight of the PIRd is obstructed making it difficult to assess the problem. The instrument was moved to its proper location in early February. The problem seems to have resolved itself with the move. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/11/1999 | 0000 | 02/11/1999 | 2359 |
Subject: | NSA/SKYRAD/C1 - Shaded PIR data incorrect |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | There were two sudden increases (about 50-100 W/m2) at 23:00 and 24:00. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/1998 | 2200 | 12/01/1998 | 2359 |
01/28/1999 | 0145 | 01/28/1999 | 0200 |
03/02/1999 | 1820 | 03/02/1999 | 1835 |
03/10/1999 | 1740 | 03/10/1999 | 1750 |
03/13/1999 | 1745 | 03/13/1999 | 1755 |
Subject: | NSA/SKYRAD/C1 - Spikes in Unshaded PIR data |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | There was a 10-25 W/m2 spike in data probably due to operator working around instrument or cleaning it. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/03/1999 | 2300 | 02/03/1999 | 2310 |
Subject: | NSA/SKYRAD/C1 - Discontinuity in Unshaded PIR data |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | There was a sudden increase in irradiance (from 140 to 190 W/m2) at 02/03/99 23:00. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/1998 | 0000 | 12/10/1998 | 2359 |
Subject: | NSA/SKYRAD/C1 - Unshaded PIR data incorrect |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Data from this time period is incorrect because instrument was swapped out for calibration and correct calibrations were not entered for replacement. Some missing data during this time interval as well |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/19/1999 | 1830 | 01/20/1999 | 2359 |
Subject: | NSA/GNDRAD/C1 - Unreasonable/missing PIR data due to maintenance |
DataStreams: | nsagndrad60sC1.a1 |
Description: | During this period, there are data gaps and the data that were collected are unreasonable. Instrument maintenance was being performed |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/01/1998 | 0000 | 02/28/1999 | 2359 |
Subject: | NSA/GNDRAD/C1 - Noisy PIR Data |
DataStreams: | nsagndrad60sC1.a1 |
Description: | During the period referenced, the ground PIR was noisy in comparison to the sky PIR data. At the end of November, the noise increased to roughly 10 W/m2 from peak to peak. After 12/23/98, the noise decreased in amplitude but still had noisy patterns. On 01/17-18/99, there was an increase again in the noise level (4 GMT 01/17 - 5 GMT 01/18). During this period the gndrad IRT data also exhibited noise patterns (the IRT was not operational). In February, the instrument exhibited large noise variations again. On 02/16, a wiring problem was corrected and the noise dropped back to 1-5 W/m2. However, the noise levels began to increase again. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/1998 | 2051 | 12/15/1998 | 1200 |
Subject: | NSA/GNDRAD/C1 - PIR1 Calibration problems resulted in small error |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The dome correction coefficients were not properly reset in calibration file until the early morning of 12/15. These errors probably account for less than 4% in error. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/16/1999 | 0030 | 02/16/1999 | 2000 |
02/17/1999 | 0000 | 02/17/1999 | 0100 |
Subject: | NSA/GNDRAD/C1 - PIR offline |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The PIR was taken off line during this period; data recorded during this period are garbage. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/08/1998 | 2030 | 12/08/1998 | 2050 |
Subject: | NSA/GNDRAD/C1 - Instrument maintenance |
DataStreams: | nsagndrad60sC1.a1 |
Description: | Instrument replaced. Data affected by the work done during this time. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/21/1999 | 0000 | 02/09/1999 | 2359 |
Subject: | NSA/GNDRAD/C1 - Bad PSP calibrations applied |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The ground radiation data are bad during this period due to erroneous calibration factors in the logger configuration file. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/1998 | 0000 | 08/10/1998 | 0000 |
Subject: | NSA/SKYRAD/C1 - NIP shaded by skyrad stand around local midnight |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Around local midnight 9:30-10:30 GMT during summer clear sky the normal incidence pyranometer is shaded by the instrument stand.This causes the reading to be 0 W/m2. If cosine correction is applied for irradiance calulations this effect is negligible due to the low solar angle. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/1998 | 0000 | 08/10/1998 | 0000 |
Subject: | NSA/SKYRAD/C1 - PSP shaded by MFRSR after local midnight |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Just past local midnight (11:00-11:40 GMT during summer clear sky) the global hemispherical pyranometer is shaded by the MFRSR instrument for about 40 minutes. This causes the reading to be 0 W/m2. This effect would be negligible in a daily average due to the low solar angle. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/24/2002 | 1725 | 11/16/2002 | 0003 |
Subject: | NSA/GNDRAD/C1 - No power to IRT |
DataStreams: | nsagndrad60sC1.a1, nsagndrad60sC1.b1 |
Description: | The data logger failed to supply power to the IRT. |
Measurements: | nsagndrad60sC1.b1:
nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1998 | 2330 | 03/03/2003 | 2359 |
Subject: | NSA/SKYRAD/C1 - IRT measurements during precipitation events |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | The IRT measurement of sky/cloud temperature was in question during precipitation events because rain or snow could accumulate in the upward-pointed lens barrel. This was corrected by reorienting the IRT to measure the sky temperature as reflected by a gold mirror. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2002 | 2330 | 06/27/2002 | 1920 |
Subject: | NSA/SKYRAD/C1 - IRT failed |
DataStreams: | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
Description: | |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2002 | 1920 | 07/10/2002 | 1900 |
Subject: | NSA/SKYRAD/C1 - IRT incorrectly configured |
DataStreams: | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
Description: | The IRT was removed for repair and replaced with a spare that was not configured properly. Data are incorrect. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/03/2003 | 2359 |
Subject: | NSA/SKYRAD/C1 - IRT incorrect min/max value |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The values of valid_min and valid_max for field sky_ir_temp were incorrectly assigned to -49.7 and +50.0 C, respectively. The actual values should be -60 and +40 C. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/1999 | 0000 | 03/28/2003 | 0200 |
Subject: | NSA/SKYRAD/C1 - IRT measurements during precipitation events |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The IRT measurement of sky/cloud temperature was in question during precipitation events because rain or snow could accumulate in the upward-pointed lens barrel. This was corrected by reorienting the IRT to measure the sky temperature as reflected by a gold mirror. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/2003 | 1400 | 09/11/2003 | 1906 |
Subject: | NSA/GNDRAD/C1 - IRT cable failure |
DataStreams: | nsagndrad60sC1.b1 |
Description: | The IRT signal became very noisy and the problem was traced to the cable (after the instrument was returned to the vendor for diagnosis). The problem started occurring when ice surrounding the cable began to melt. The cable was repaired and the instrument reinstalled. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/11/2003 | 1906 | 10/15/2003 | 0015 |
Subject: | NSA/GNDRAD/C1 - IRT returned to vendor |
DataStreams: | nsagndrad60sC1.b1 |
Description: | The IRT was returned to the vendor for diagnosis. It was found to be operating as expected and the signal noise problem was traced to the cable. The cable was repaired and the instrument reinstalled. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1200 | 01/09/2004 | 2120 |
Subject: | NSA/SKYRAD/C1 - Downwelling IRT bias |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The downwelling IRT was measuring higher sky/cloud temperatures than the AERI. The problem was corrected when the mirror was replaced. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/10/2003 | 0000 | 02/16/2004 | 0230 |
Subject: | NSA/SKYRAD/C1 - Tracker Inoperative |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | A blown fuse resulted in the tracker not shading the PIR and B&W and the NIP not following the sun. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2003 | 0200 | 06/21/2006 | 0215 |
Subject: | NSA/SKYRAD/C1 - itermittant IRT bias |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky temperature measurement then a sudden correction, as compared to AERI, appears to have been caused by a combination of a failing IR detector, a poor signal cable, a contaminated (with condensation or frost) mirror or lens, and low ambient temperatures. The positive bias was apparently not consistant in either frequency or duration but on days when the unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. The instrument was repaired and recalibrated and the signal cable replaced. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/15/2004 | 0000 | 09/22/2004 | 0552 |
Subject: | NSA/GNDRAD/C1 - IRT power failure |
DataStreams: | nsagndrad60sC1.b1 |
Description: | After performing corrective maintenance which involved turning off power to the GNDRAD instruments and data logger, power to the IRT was not restored. The problem was corrected by replacing the data logger. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2005 | 0600 | 01/18/2006 | 0032 |
Subject: | NSA/SKYRAD/C1 - IRT failure |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The IRT failed and was replaced with a spare. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2006 | 0910 | 02/07/2006 | 2350 |
Subject: | NSA/GNDRAD/C1 - power outage affecting some GNDRAD sensors |
DataStreams: | nsagndrad20sC1.a0, nsagndrad60sC1.b1, nsagndradC1.00 |
Description: | During the power outage the upwelling global measurement from the PSP instrument shows a slight increase from -5 to +0 W/m2. This is due to removal of thermal offsets induced by the AC heater during the power outage. This heater is installed within the PSP ventilator - it is necessary for the NSA/C1 climate. The upwelling longwave PIR instrument is unaffected due to case and dome temperature corrections. The IRT is AC powered and was not operational during this time period. |
Measurements: | nsagndrad60sC1.b1:
nsagndradC1.00:
nsagndrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2006 | 0910 | 02/07/2006 | 2350 |
Subject: | NSA/SKYRAD/C1 - Power outage affecting some SKYRAD instruments |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1, nsaskyradC1.00 |
Description: | During the power outage the downwelling global measurement from the PSP instrument and downwelling diffuse 8-48 instrument shows a slight increase from -5 to +0 W/m2. This is due to removal of thermal offsets induced by the AC heater during the power outage. This heater is installed within the PSP ventilator - it is necessary for the NSA/C1 climate. The downwelling longwave PIR instrument is unaffected due to case and dome temperature corrections. The IRT is AC powered and was not operational during this time period. |
Measurements: | nsaskyradC1.00:
nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2004 | 1600 | 12/21/2005 | 1855 |
Subject: | NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error |
DataStreams: | nsagndrad60sC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/29/1999 | 2041 | 08/05/1999 | 2359 |
09/23/2000 | 2257 | 09/25/2000 | 0119 |
03/15/2002 | 2036 | 03/18/2002 | 2035 |
09/07/2003 | 1103 | 09/19/2003 | 0201 |
11/05/2003 | 0339 | 11/07/2003 | 2234 |
01/03/2004 | 1153 | 01/06/2004 | 0031 |
Subject: | NSA/SKYRAD/C1 - Missing data |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/27/1999 | 2150 | 08/05/1999 | 2359 |
08/06/1999 | 1208 | 08/07/1999 | 2359 |
10/03/1999 | 0000 | 11/04/1999 | 2359 |
11/23/1999 | 0000 | 11/28/1999 | 2359 |
01/28/2000 | 0000 | 02/03/2000 | 2341 |
09/21/2000 | 1619 | 09/24/2000 | 2258 |
03/15/2002 | 2052 | 03/21/2002 | 1502 |
Subject: | NSA/GNDRAD/C1 - Missing data |
DataStreams: | nsagndrad60sC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2007 | 0359 | 10/25/2007 | 2148 |
Subject: | NSA/SKYRAD/C1 - Reprocessed: Incorrect calibration coefficient |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | Installed Skyrad PSP (downwelling shortwave hemispheric irradiance) had calibration coefficient Rs applied when it should have used Rsc. ARM is using IR corrected Rsc and CFc coefficients for PSPs deployed at all facilities. PSP #32026 should use RSc@45=8.7578 ?V/W/m? and not Rs@45=8.6335. Reprocessed to correct downwelling shortwave hemispheric irradiance (DS) by: DS(new) = (DS(old) / 115.83 ) * 114.18 Reprocessing completed March 2008. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/04/2007 | 0130 | 10/25/2007 | 2149 |
Subject: | NSA/GNDRAD/C1 - Reprocessed: Incorrect calibration coefficient |
DataStreams: | nsagndrad20sC1.a0, nsagndrad60sC1.b1 |
Description: | Installed Gndrad PSP (upwelling shortwave hemispheric irradiance) had calibration coefficient RS applied when it should have used Rsc. ARM is using IR corrected coefficients Rsc@45 and CFc@45 for PSPs deployed at all facilities. PSP #32012F3 should use RSc@45=109.39 and not Rs@45=110.97. Reprocessed to correct upwelling shortwave hemispheric irradiance (US) by: US(new) = (US(old) / 110.97 ) * 109.39 Reprocessing completed March 2008. |
Measurements: | nsagndrad60sC1.b1:
nsagndrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/2007 | 0000 | 12/24/2007 | 2359 |
Subject: | NSA/SKYRAD/C1 - Instrument cleaning effects |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | Starting Dec 16th the Skyrad PIR1 and PIR2 (down_long_hemispheric)started differing by greater than 30W/m2 at times. This is much greater than expected for these instruments (typical is <10-15 W/m2). NSA OMIS reports and observer comments reveal that PM periodically clean the snow from the IRT and ventilator fans on Skyrad and Gndrad. In the case of the Skyrad PIRs, one lies in the wind shadow of the other and builds up more snow which is periodically cleaned out and then builds up again causing the PIR1 PIR2 discrepancies. Disregard down_long_hemispheric during this time period. However, data suggests PIR2 is most likely at fault - potentially use PIR1 data during this time period. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/2007 | 0000 | 02/11/2008 | 1135 |
Subject: | NSA/SKYRAD/C1 - IRT mirror heater failure |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | The heater of the IRT gold mirror failed due to a broken wire (that probably occurred on 12/13 during maintenance of the solar tracker) allowing snow and frost to accumulate on the mirror until it was cleared during daily preventative maintenance. The problem was corrected when the heater wires were replaced. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2008 | 0632 | 04/23/2008 | 1815 |
Subject: | NSA/SKYRAD/C1 - Direct and Diffuse incorrect |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | The short_direct_normal (NIP) instrument and cable were damaged by tracker rewind malfunction. Tracker and NIP replaced on 4/22 2200GMT but tracker not operational until 4/23 1815GMT. Thus, down_short_diffuse_hemisp (8-48) was not shaded 4/22 to 4/23. Do not use 8-48 data from 4/22 to 4/23. NIP data is not recoverable during entire time but can be estimated using component sum for each 1 minute data: Direct = (global - diffuse)/cos(zenith angle) |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2008 | 2000 | 06/30/2008 | 1740 |
Subject: | NSA/SKYRAD/C1 - Tracker inoperative |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | Tracker stopped working at this site during the time range specified. Reason for failure is unknown. Tracker power cycled and restarted by maintenance 06/30 2008 1740 GMT. The short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data are bad during this time. Data are not recoverable. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/1998 | 2200 | 01/20/1999 | 0059 |
Subject: | NSA/GNDRAD/C1 - IRT not connected |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The IRT was unplugged from the Zeno data logger on 12/10/1998 for a interference test and when it was plugged back in on 12/22, the connector at the data logger was damaged. The data logger and connecter were replaced on 1/20/1999. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/20/1999 | 0100 | 03/04/1999 | 1506 |
Subject: | NSA/GNDRAD/C1 - Reprocess: IRT scale factor |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The IRT scale factors in the data logger were incorrectly set to m = 100 degC/V, b = -60 degC. They were corrected to 25 degC/V and -50 degC. Correct with: IRT(C) = (IRT_bad+60C)*(25/100) - 50C |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/1999 | 2138 | 03/03/1999 | 1200 |
Subject: | NSA/GNDRAD/C1 - IRT failed |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The IRT failed and was replaced with one installed inside an insulated enclosure. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/1998 | 0000 | 05/13/1998 | 2152 |
Subject: | NSA/GNDRAD/C1 - IRT incorrect |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The IRT data were very noisy and unrepresentative. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/13/1998 | 2152 | 06/04/1998 | 2208 |
Subject: | NSA/GNDRAD/C1 - IRT scale factor |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The IRT scale factors in the data logger were incorrectly set to m = 100 degC/V, b = -60 degC. They were corrected to 25 degC/V and -50 degC. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1998 | 0330 | 06/17/1999 | 2327 |
Subject: | NSA/SKYRAD/C1 - IRT signal bias |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | Additional heat tape that was installed around the IRT appears to have produced a negative bias of about 10 deg (at a sky temperature of -30C). This was corrected when the heat tape was replaced with thermostatically-controlled 75W heaters (that were still insufficient in maintaining the IRT above freezing - D990924.2). |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2008 | 0330 | 09/15/2008 | 1930 |
Subject: | NSA/GNDRAD/C1 - Upwelling PIR ventilator fan failure |
DataStreams: | nsagndrad20sC1.a0, nsagndrad60sC1.b1 |
Description: | The PIR exhibited a positive bias of approximately 6 W/m2 compared to the upwelling IRT due to ventilator fan failure. Thus, the up_long_hemisp, data stream is high by about 6 W/m2. Bias is not consistent and depends on atmospheric conditions. Either disregard the up_long_hemisp data or use the IRT for an estimate of the up_long_hemisp values during this time period. |
Measurements: | nsagndrad60sC1.b1:
nsagndrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/03/2000 | 0000 | 03/03/2003 | 2359 |
Subject: | NSA/GNDRAD/C1 - Net radiometer removed |
DataStreams: | nsagndrad60sC1.a1, nsagndrad60sC1.b1 |
Description: | The net radiometer was removed from the NSA.C1 GNDRAD instrument suite on Feb 3, 2000. Net rad variables were filled with garbage data until Sept 23, 2000 and with missing values (-9999) until 20041103. The net rad variables were removed from the netcdf files after 20041103. A reprocessing task (RID-526) has been submitted to the queue to address the need to add metadata to the NSA.C1 GNDRAD netcdf files documenting the missing data as well as to move toward a consistent datastream format throughout time. |
Measurements: | nsagndrad60sC1.b1:
nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/1999 | 0000 | 12/22/2000 | 2020 |
Subject: | NSA/GNDRAD/C1 - Reprocess: IRT insufficiently heated |
DataStreams: | nsagndrad60sC1.b1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of surface temperature were over-estimated when instrument was below freezing. Data will be reprocessed, but users can correct data using the following correction factor: If T_reference < -5?C, then IRT_corrected = (IRT_original - 32.993K)/0.87238 where T_reference can be estimated with the ambient temperature (e.g. inst_up_long_hemisp_dome_temp or inst_up_long_dome_temp) |
Measurements: | nsagndrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/1999 | 0000 | 12/21/2000 | 1925 |
Subject: | NSA/SKYRAD/C1 - Reprocess: IRT insufficiently heated |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of sky temperature were over-estimated when instrument was below freezing. Data will be reprocessed, but users can correct data using the following correction factor: If T_reference < -5?C, then IRT_corrected = (IRT_original - 32.993K)/0.87238 where T_reference can be estimated with the ambient temperature (e.g. inst_down_long_hemisp_dome_temp or inst_down_long_dome_temp) |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2009 | 1900 | 02/24/2009 | 0130 |
Subject: | NSA/SKYRAD/C1 - IRT covered with snow |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The IRT mirror was buried with snow by the dew-line snow-blower. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1998 | 0000 | 12/12/1998 | 2359 |
Subject: | NSA/SKYRAD - Barrow SKYRAD PIRD Misconfiguration 12/9 - 12/12/98 |
DataStreams: | nsaskyrad20sC1.a1, nsaskyrad60sC1.a1 |
Description: | The PIRD was incorrectly set from 12/9/98 to 12/12/98 that resulted in an error of approximately 3.7% (was set to 243902 instead of 235294). |
Measurements: | nsaskyrad60sC1.a1:
nsaskyrad20sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/16/1998 | 1900 | 10/28/1998 | 1300 |
Subject: | NSA/SKYRAD/C1 - IRT removed for repair |
DataStreams: | nsaskyrad60sC1.a1, nsaskyradC1.00 |
Description: | The instrument was damaged by internal water condensation caused by the exterior foam insulation used to keep the IRT above freezing. It was returned in July 1998 to the manufacturer for repair. |
Measurements: | nsaskyradC1.00:
nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/21/1998 | 1800 | 12/03/1998 | 0000 |
Subject: | NSA/SKYRAD/C1 - IRT removed from service |
DataStreams: | nsaskyrad60sC1.a1, nsaskyradC1.00 |
Description: | Instrument was removed to check performance. |
Measurements: | nsaskyradC1.00:
nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/1998 | 2156 | 12/22/2000 | 2020 |
Subject: | NSA/GNDRAD/C1 - Reprocess: IRT insufficiently heated |
DataStreams: | nsagndrad60sC1.a1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of surface temperature were over-estimated when instrument was below freezing. Data will be reprocessed, but users can correct data using the following correction factor: If T_reference < -5?C, then IRT_corrected = (IRT_original - 32.993K)/0.87238 where T_reference can be estimated with the ambient temperature (e.g. inst_up_long_hemisp_dome_temp or inst_up_long_dome_temp) |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1998 | 0130 | 12/21/2000 | 1925 |
Subject: | NSA/SKYRAD/C1 - Reprocess: IRT insufficiently heated |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of sky temperature were over-estimated when instrument was below freezing. Data will be reprocessed, but users can correct data using the following correction factor: If T_reference < -5?C, then IRT_corrected = (IRT_original - 32.993K)/0.87238 where T_reference can be estimated with the ambient temperature (e.g. inst_down_long_hemisp_dome_temp or inst_down_long_dome_temp) |