Description: | Even though the polarization voltage was reported as switching between circular and cross
polarization states, there was no cross polarization taking place because the
polarization box had not been recycled following a system reboot.
You will not be able to create depolarization values. However all other processes, such
as cloud base determination should be ok. |
Measurements: | nsamplpolavgC1.s1: - Attenuated backscatter, cross-pol(signal_return_cross_pol)
- Background signal, cross-pol(background_signal_cross_pol)
- *(aqc_signal_return_cross_pol)
- Preliminary cloud base height above ground level, cross-pol(preliminary_cbh_cross_pol)
- Background signal standard deviation, cross-pol(background_signal_std_cross_pol)
- Polarization control voltage, cross-pol(polarization_control_voltage_cross_pol)
nsamplpolC1.b1: - Polarization control voltage setting(polarization_control_voltage)
- Attenuated backscatter(signal_return)
nsamplpolavgC1.c1: - Attenuated backscatter, cross-pol(signal_return_cross_pol)
- Polarization control voltage, cross-pol(polarization_control_voltage_cross_pol)
- *(aqc_signal_return_cross_pol)
- Number of samples used for cross-pol averages(nsamples_cross_pol_average)
- Background signal standard deviation, cross-pol(background_signal_std_cross_pol)
- Background signal, cross-pol(background_signal_cross_pol)
- Preliminary cloud base height above ground level, cross-pol(preliminary_cbh_cross_pol)
|