DQR ID | Subject | Data Streams Affected |
---|---|---|
D070223.1 | SGP/ECOR/E14 - Fluxes and Most Measurements Incorrect | sgp30ecorE14.b1 |
D070227.1 | SGP/ECOR/E5 - Reprocess: Time shifted -7 hours | sgpecorE5.00, sgp30ecorE5.b1 |
D070402.1 | SGP/ECOR/E21 - Intermittent incorrect flux measurements | sgp30ecorE21.b1 |
D070416.4 | SGP/CO2FLX/C1 - downwelling longwave hemispheric irradiance incorrect | sgp30co2flx4mmetC1.b1 |
D070510.1 | SGP/ECOR/E5 - Flux Measurements Incorrect | sgp30ecorE5.b1 |
D070514.2 | SGP/ECOR/E14 - IRGA Temperature Incorrect or Missing | sgp30ecorE14.b1 |
D070619.1 | SGP/ECOR/E5 - Flux Data Incorrect | sgp30ecorE5.b1 |
D070803.2 | SGP/CO2FLX/C1 - Disturbance around site | sgp30co2flx4mC1.b1, sgp30co2flx25mC1.b1, sgp30co2flx60mC1.b1, sgp30co2flx4mmetC1.b1 |
D070918.3 | SGP/CO2FLX/C1 - Farm operation | sgp30co2flx4mmetC1.b1 |
D071011.2 | SGP/CO2FLX/C1 - Reprocess: 4m instruments height | sgp30co2flx4mC1.b1 |
D080428.1 | SGP/ECOR/E24 - Half Hours of Data Missing | sgp30ecorE24.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/2007 | 0500 | 02/02/2007 | 1600 |
Subject: | SGP/ECOR/E14 - Fluxes and Most Measurements Incorrect |
DataStreams: | sgp30ecorE14.b1 |
Description: | The ECOR system was having major problems during this period and most measurements were incorrect most of the time. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2007 | 1900 | 02/23/2007 | 2330 |
Subject: | SGP/ECOR/E5 - Reprocess: Time shifted -7 hours |
DataStreams: | sgpecorE5.00, sgp30ecorE5.b1 |
Description: | An inadvertant power outage caused the ECOR program to shut down abnormally. When power was restored, the timestamps on the ECOR files were 7 hours behind the correct time and the site ID was not accessed from the ECOR configuration file resulting in site ID (05) being replaced with "AN". These conditions caused the ECOR files to be unrecognizable and therefore are presently missing in the Archive. Correction of the timestamp and ID in the files should allow them to be processed and added to the Archive. |
Measurements: | sgpecorE5.00:
sgp30ecorE5.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2007 | 0800 | 08/29/2007 | 1730 |
Subject: | SGP/ECOR/E21 - Intermittent incorrect flux measurements |
DataStreams: | sgp30ecorE21.b1 |
Description: | At various times during the period, the LI-7500 CO2/H2O serial and analog signals from the LI-7500 were missing, but usually not at the same time. When the CO2/H2O serial datastream was missing, the Pressure and Temperature were missing; this did not, in itself, result in incorrect fluxes, as default values for pressure and temperature are automatically used. Fluxes were incorrect when the CO2/H2O analog datastream was missing; these times are identified by the qc checks. |
Measurements: | sgp30ecorE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/28/2004 | 0000 | 04/02/2007 | 1930 |
Subject: | SGP/CO2FLX/C1 - downwelling longwave hemispheric irradiance incorrect |
DataStreams: | sgp30co2flx4mmetC1.b1 |
Description: | Downwelling longwave radiation measurements (variable "r_down_long_hemisp") are bad due to a wiring issue. Users should not use the downwelling longwave hemispheric irradiance during this period. The downwelling shortwave and longwave radiation at CF is also available from the Broadband Radiometer Station (BRS). |
Measurements: | sgp30co2flx4mmetC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/20/2005 | 1930 | 12/14/2005 | 1930 |
04/18/2007 | 1930 | 05/17/2007 | 1815 |
Subject: | SGP/ECOR/E5 - Flux Measurements Incorrect |
DataStreams: | sgp30ecorE5.b1 |
Description: | The ECOR datastream was corrupted; this happens when the ECOR program produces a datastream of the wrong length (i.e., leaves one or more measurements out). Fluxes and some other measurements are incorrect. Note: Mentor reports reprocessing will not correct this problem. |
Measurements: | sgp30ecorE5.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2007 | 1300 | 05/19/2007 | 1700 |
Subject: | SGP/ECOR/E14 - IRGA Temperature Incorrect or Missing |
DataStreams: | sgp30ecorE14.b1 |
Description: | The temp_irga was too large 05/09/07 1300 GMT - 05/10/07 2000 GMT. The temp_irga was flat lined at 49.96 C 05/10/07 2030 GMT - 05/17/07 2000 GMT. The temp_irga was missing 05/17/07 2030 GMT - 05/19/07 1700 GMT. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/16/2007 | 0300 | 12/12/2007 | 2030 |
Subject: | SGP/ECOR/E5 - Flux Data Incorrect |
DataStreams: | sgp30ecorE5.b1 |
Description: | CO2 flux (lv_e) and latent heat flux (fc) were incorrect because the sonic anemometer was not working. |
Measurements: | sgp30ecorE5.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2007 | 1300 | 07/26/2007 | 2359 |
Subject: | SGP/CO2FLX/C1 - Disturbance around site |
DataStreams: | sgp30co2flx4mC1.b1, sgp30co2flx25mC1.b1, sgp30co2flx60mC1.b1, sgp30co2flx4mmetC1.b1 |
Description: | A bulldozer used to excavate the SGP Tower anchors was right around (within 10 feet) our systems. |
Measurements: | sgp30co2flx4mC1.b1:
sgp30co2flx4mmetC1.b1:
sgp30co2flx60mC1.b1:
sgp30co2flx25mC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/22/2007 | 1830 | 09/07/2007 | 1800 |
Subject: | SGP/CO2FLX/C1 - Farm operation |
DataStreams: | sgp30co2flx4mmetC1.b1 |
Description: | Soil temperature, soil moisture, soil heat flux and upwelling surface radiation (long, short, and PAR) measurements were not valid from 20070822 to 20070907 due to farm operations. |
Measurements: | sgp30co2flx4mmetC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2003 | 0000 | 10/11/2007 | 0000 |
Subject: | SGP/CO2FLX/C1 - Reprocess: 4m instruments height |
DataStreams: | sgp30co2flx4mC1.b1 |
Description: | The height of the 4m instruments is 4.8 m and not 4.65 m as shown in b1 datastream |
Measurements: | sgp30co2flx4mC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/27/2006 | 0530 | 11/08/2007 | 1600 |
Subject: | SGP/ECOR/E24 - Half Hours of Data Missing |
DataStreams: | sgp30ecorE24.b1 |
Description: | Individual half hours of data were missing infrequently because of a UPS problem. The extent of missing data is minimal, with only a few to several half hours missing most months. Data quality of the surrounding data times was not affected. |
Measurements: | sgp30ecorE24.b1:
|