DQR ID | Subject | Data Streams Affected |
---|---|---|
D050722.1 | SGP/MWR/C1 - REPROCESS - Revised Retrieval Coefficients | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgp1mwravgC1.c1, sgp5mwravgC1.c1, sgpqmemwrcolC1.c1 |
D050915.1 | SGP/MWR/C1 - Instrument noise problem | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgp1mwravgC1.c1, sgp5mwravgC1.c1 |
D051011.6 | SGP/MWR/C1 - New software version (4.15) installed | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D051031.3 | SGP/ECOR/E14 - CO2/H2O Incorrect | sgp30ecorE14.b1 |
D051216.2 | SGP/ECOR/E14 - Flux Measurements Incorrect | sgp30ecorE14.b1 |
D060717.2 | SGP/MWR/C1 - Spikes in ambient temperature readings | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D070223.1 | SGP/ECOR/E14 - Fluxes and Most Measurements Incorrect | sgp30ecorE14.b1 |
D070314.2 | SGP/MWR/C1/E14 - Freezing rain-Incorrect rain flag | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgpmwrlosE14.b1, sgpmwrtipE14.a1 |
D070514.2 | SGP/ECOR/E14 - IRGA Temperature Incorrect or Missing | sgp30ecorE14.b1 |
D070802.3 | SGP/MWR/C1 - Noisy data | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D071128.2 | SGP/MWR/C1 - Reprocess: Radiometer reinstalled- Please reprocess | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgp5mwravgC1.c1 |
D080103.2 | SGP/MWR/C1 - Missing data | sgpmwrlosC1.b1 |
D080604.2 | SGP/ECOR/E14 - Sensor failure | sgp30ecorE14.b1 |
D081013.5 | SGP/MWR/C1 - Increased noise in 31.4 GHz channel | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
D110204.1 | SGP/SEBS/E14 - Instrument logger problem | sgp30ecorE14.b1 |
D110322.1 | SGP/ECOR/E14 - Instrument logger problem | sgpecorE14.00, sgp30ecorE14.b1 |
D110322.2 | SGP/SEBS/E14 - Soil Moisture Measurements Incorrect | sgp30ecorE14.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2002 | 2000 | 06/28/2005 | 2300 |
Subject: | SGP/MWR/C1 - REPROCESS - Revised Retrieval Coefficients |
DataStreams: | sgpmwrlosC1.a1, sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgp1mwravgC1.c1, sgp5mwravgC1.c1, sgpqmemwrcolC1.c1 |
Description: | IN THE BEGINNING (June 1992), the retrieval coefficients used to derive the precipitable water vapor (PWV) and liquid water path (LWP) from the MWR brightness temperatures were based on the Liebe and Layton (1987) water vapor and oxygen absorption model and the Grant (1957) liquid water absorption model. Following the SHEBA experience, revised retrievals based on the more recent Rosenkranz (1998) water vapor and oxygen absorption models and the Liebe (1991) liquid waer absorption model were developed. The Rosenkranz water vapor absorption model resulted a 2 percent increase in PWV relative to the earlier Liebe and Layton model. The Liebe liquid water absorption model decreased the LWP by 10% relative to the Grant model. However, the increased oxygen absorption caused a 0.02-0.03 mm (20-30 g/m2) reduction in LWP, which was particularly significant for low LWP conditions (i.e. thin clouds encountered at SHEBA). Recently, it has been shown (Liljegren, Boukabara, Cady-Pereira, and Clough, TGARS v. 43, pp 1102-1108, 2005) that the half-width of the 22 GHz water vapor line from the HITRAN compilation, which is 5 percent smaller than the Liebe and Dillon (1969) half-width used in Rosenkranz (1998), provided a better fit to the microwave brightness temperature measurements at 5 frequencies in the range 22-30 GHz, and yielded more accurate retrievals. Accordingly, revised MWR retrieval coefficients have been developed using MONORTM, which utilizes the HITRAN compilation for its spectroscopic parameters. These new retrievals provide 3 percent less PWV and 2.6 percent greater LWP than the previous retrievals based on Rosenkranz (1998). Although the MWR data will be reprocessed to apply the new monortm-based retrievals, for most purposes it will be sufficient to correct the data using the following factors: PWV_MONORTM = 0.9695 * PWV_ROSENKRANZ LWP_MONORTM = 1.026 * LWP_ROSENKRANZ The Rosenkranz-based retrieval coefficients became active as follows (BCR 456): SGP/C1 (Lamont) 4/16/2002, 2000 SGP/B1 (Hillsboro) 4/12/2002, 1600 SGP/B4 (Vici) 4/15/2002, 2300 SGP/B5 (Morris) 4/15/2002, 2300 SGP/B6 (Purcell) 4/16/2002, 2200 SGP/E14(Lamont) 4/16/2002, 0000 NSA/C1 (Barrow) 4/25/2002, 1900 NSA/C2 (Atqasuk) 4/18/2002, 1700 TWP/C1 (Manus) 5/04/2002, 0200 TWP/C2 (Nauru) 4/27/2002, 0600 TWP/C3 (Darwin) inception The MONORTM-based retrieval coefficients became active as follows (BCR 984): SGP/C1 (Lamont) 6/28/2005, 2300 SGP/B1 (Hillsboro) 6/24/2005, 2100 SGP/B4 (Vici) 6/24/2005, 2100 SGP/B5 (Morris) 6/24/2005, 2100 SGP/B6 (Purcell) 6/24/2005, 1942 SGP/E14(Lamont) 6/28/2005, 2300 NSA/C1 (Barrow) 6/29/2005, 0000 NSA/C2 (Atqasuk) 6/29/2005, 0000 TWP/C1 (Manus) 6/30/2005, 2100 TWP/C2 (Nauru) 6/30/2005, 2100 TWP/C3 (Darwin) 6/30/2005, 2100 PYE/M1 (Pt. Reyes) 4/08/2005, 1900** ** At Pt. Reyes, the original retrieval coefficients implemented in March 2005 were based on a version of the Rosenkranz model that had been modified to use the HITRAN half-width at 22 GHz and to be consistent with the water vapor continuum in MONORTM. These retrievals yield nearly identical results to the MONORTM retrievals. Therefore the Pt. Reyes data prior to 4/08/2005 may not require reprocessing. |
Measurements: | sgp5mwravgC1.c1:
sgpmwrtipC1.a1:
sgpqmemwrcolC1.c1:
sgp1mwravgC1.c1:
sgpmwrlosC1.a1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/28/2005 | 1400 | 08/05/2005 | 1700 |
Subject: | SGP/MWR/C1 - Instrument noise problem |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgp1mwravgC1.c1, sgp5mwravgC1.c1 |
Description: | Various variables including the mixer temperatures were very noisy. After several attempts to fix the problem, the instrument was taken off line and returned to the manufacturer for repair. |
Measurements: | sgp5mwravgC1.c1:
sgpmwrtipC1.a1:
sgp1mwravgC1.c1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/31/2002 | 2027 | 08/04/2005 | 1959 |
Subject: | SGP/MWR/C1 - New software version (4.15) installed |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Description: | A problem began with the installation of MWR.EXE version 4.12 in July 2002. The software had been upgraded from a "DOS" to a "Windows"-compiled program to address an earlier problem. The software upgrade corrected the earlier problem but introduced a new one that caused line-of-sight observing cycles to be skipped, a 15% reduction in the number of tip curves, and saturation of CPU usage. Software versions 4.13 and 4.14 also produced these problems. The new MWR software, version 4.15, was installed on 08/04/2005. As a consequence of this upgrade, the tip curve frequency increased. The tip cycle time decreased from ~60s to ~50s. |
Measurements: | sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/25/2005 | 0000 | 08/29/2005 | 0230 |
Subject: | SGP/ECOR/E14 - CO2/H2O Incorrect |
DataStreams: | sgp30ecorE14.b1 |
Description: | The LI-7500 CO2/H2O sensor stopped working after a rainstorm. CO2 and H2O data are incorrect during the stated period. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/04/2005 | 1330 | 11/17/2005 | 1630 |
12/01/2005 | 1930 | 12/13/2005 | 1630 |
01/24/2006 | 1500 | 02/13/2006 | 1930 |
05/05/2006 | 2100 | 06/13/2006 | 1400 |
07/07/2006 | 1930 | 07/14/2006 | 1500 |
Subject: | SGP/ECOR/E14 - Flux Measurements Incorrect |
DataStreams: | sgp30ecorE14.b1 |
Description: | The ECOR datastream was corrupted; this happens when the ECOR program produces a datastream of the wrong length (i.e., leaves one or more measurements out). Fluxes and some other measurements are incorrect. Reprocessing will not correct this problem; the raw data are corrupted. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/13/2006 | 1200 | 07/28/2006 | 1600 |
Subject: | SGP/MWR/C1 - Spikes in ambient temperature readings |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Description: | There are intermittent spikes in the Tkair temperature (T ~ 400 K). On 7/28 the sensor was cleaned and the instrument power cycled following a power outage. Readings came back to normal. |
Measurements: | sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/14/2007 | 0500 | 02/02/2007 | 1600 |
Subject: | SGP/ECOR/E14 - Fluxes and Most Measurements Incorrect |
DataStreams: | sgp30ecorE14.b1 |
Description: | The ECOR system was having major problems during this period and most measurements were incorrect most of the time. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/12/2007 | 0000 | 01/15/2007 | 0000 |
Subject: | SGP/MWR/C1/E14 - Freezing rain-Incorrect rain flag |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgpmwrlosE14.b1, sgpmwrtipE14.a1 |
Description: | Between 01/12 and 01/15 brightness temperatures show high values (Tb > 100 K) indicative of rain, however the wet_window flag is 0 (indicative of no rain). During that time there where freezing conditions with ice pellets that were not detected by the sensor. The high brightness temperatures may be due to melting ice on the window. |
Measurements: | sgpmwrtipE14.a1:
sgpmwrtipC1.a1:
sgpmwrlosE14.b1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2007 | 1300 | 05/19/2007 | 1700 |
Subject: | SGP/ECOR/E14 - IRGA Temperature Incorrect or Missing |
DataStreams: | sgp30ecorE14.b1 |
Description: | The temp_irga was too large 05/09/07 1300 GMT - 05/10/07 2000 GMT. The temp_irga was flat lined at 49.96 C 05/10/07 2030 GMT - 05/17/07 2000 GMT. The temp_irga was missing 05/17/07 2030 GMT - 05/19/07 1700 GMT. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2007 | 0000 | 08/03/2007 | 1508 |
Subject: | SGP/MWR/C1 - Noisy data |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Description: | Since 7/14 the MWR is experiencing intermittent periods of increased noise. The cause for this is under investigation. |
Measurements: | sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/19/2007 | 2136 | 11/22/2007 | 1400 |
Subject: | SGP/MWR/C1 - Reprocess: Radiometer reinstalled- Please reprocess |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1, sgp5mwravgC1.c1 |
Description: | After the radiometer was reinstalled it had to update the calibration coefficients. Data will be reprocessed |
Measurements: | sgp5mwravgC1.c1:
sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/21/1993 | 1809 | 07/23/1993 | 1541 |
08/07/1993 | 2143 | 08/09/1993 | 1308 |
08/14/1993 | 2233 | 08/16/1993 | 1300 |
08/29/1993 | 1207 | 08/31/1993 | 1206 |
09/17/1993 | 2158 | 09/20/1993 | 1311 |
10/02/1993 | 1612 | 10/04/1993 | 1338 |
10/09/1993 | 0741 | 10/11/1993 | 1319 |
11/12/1993 | 1207 | 11/15/1993 | 1413 |
11/19/1993 | 2027 | 11/21/1993 | 1957 |
12/14/1993 | 2300 | 03/21/1994 | 1759 |
03/28/1994 | 1459 | 03/30/1994 | 2225 |
05/16/1994 | 1900 | 05/19/1994 | 2018 |
06/22/1994 | 1600 | 07/11/1994 | 1859 |
07/16/1994 | 0300 | 07/19/1994 | 2006 |
09/03/1994 | 1647 | 09/05/1994 | 1341 |
09/09/1994 | 2325 | 09/11/1994 | 1543 |
12/17/1994 | 0049 | 12/19/1994 | 1441 |
01/05/1995 | 0000 | 01/05/1995 | 2359 |
01/28/1995 | 0054 | 01/30/1995 | 1443 |
04/01/1995 | 0000 | 04/03/1995 | 1610 |
05/09/1995 | 2200 | 05/11/1995 | 2350 |
06/24/1995 | 0903 | 06/26/1995 | 0159 |
07/07/1995 | 1625 | 07/10/1995 | 1409 |
07/19/1995 | 1741 | 07/21/1995 | 1906 |
07/28/1995 | 1900 | 07/31/1995 | 1918 |
08/22/1995 | 1616 | 08/25/1995 | 2359 |
12/01/1995 | 1434 | 01/11/1996 | 2214 |
01/12/1996 | 2000 | 01/23/1996 | 2030 |
01/23/1996 | 2332 | 02/08/1996 | 2359 |
08/06/1996 | 0000 | 08/06/1996 | 2359 |
09/04/1996 | 1910 | 09/08/1996 | 2359 |
09/14/1996 | 0700 | 09/16/1996 | 1759 |
09/30/1996 | 0000 | 09/30/1996 | 2359 |
12/01/1996 | 0000 | 12/01/1996 | 2359 |
07/01/1997 | 0000 | 07/01/1997 | 2359 |
02/20/1998 | 2339 | 02/22/1998 | 0042 |
05/04/1998 | 0000 | 05/04/1998 | 2359 |
07/18/1998 | 0337 | 07/20/1998 | 1355 |
09/28/1998 | 0000 | 09/28/1998 | 2359 |
10/10/1998 | 2359 | 10/12/1998 | 1959 |
10/17/1998 | 1705 | 10/19/1998 | 1535 |
12/25/1998 | 2233 | 12/28/1998 | 2026 |
02/06/1999 | 1235 | 02/08/1999 | 1652 |
03/13/1999 | 2048 | 03/15/1999 | 2040 |
07/23/1999 | 2059 | 07/26/1999 | 1408 |
12/04/1999 | 0800 | 12/06/1999 | 1821 |
01/07/2000 | 1446 | 02/02/2000 | 2224 |
05/24/2000 | 1455 | 05/31/2000 | 1505 |
06/06/2000 | 1620 | 06/14/2000 | 1451 |
09/01/2000 | 2200 | 09/05/2000 | 1526 |
09/15/2000 | 2057 | 09/18/2000 | 1339 |
09/23/2000 | 1236 | 09/25/2000 | 1430 |
01/13/2001 | 2011 | 01/16/2001 | 1647 |
02/24/2001 | 0230 | 02/26/2001 | 1508 |
03/03/2001 | 0701 | 03/05/2001 | 1546 |
03/10/2001 | 0702 | 03/12/2001 | 1535 |
03/17/2001 | 0702 | 03/19/2001 | 1422 |
03/24/2001 | 0703 | 03/26/2001 | 1516 |
03/30/2001 | 2038 | 04/02/2001 | 1423 |
05/05/2001 | 1042 | 05/07/2001 | 1514 |
05/18/2001 | 1534 | 05/21/2001 | 1451 |
06/22/2001 | 2200 | 06/25/2001 | 1438 |
06/30/2001 | 0616 | 07/02/2001 | 1435 |
07/28/2001 | 1944 | 07/30/2001 | 1357 |
08/24/2001 | 1348 | 08/27/2001 | 1751 |
09/07/2001 | 1842 | 09/10/2001 | 1439 |
09/15/2001 | 1533 | 09/17/2001 | 1459 |
10/13/2001 | 0033 | 10/15/2001 | 1526 |
11/03/2001 | 0840 | 11/05/2001 | 1613 |
11/10/2001 | 0558 | 11/13/2001 | 1518 |
11/16/2001 | 1701 | 11/19/2001 | 1706 |
12/21/2001 | 2153 | 12/24/2001 | 1442 |
01/30/2002 | 0845 | 02/11/2002 | 1632 |
03/02/2002 | 0339 | 03/04/2002 | 1533 |
03/09/2002 | 0340 | 03/11/2002 | 1621 |
06/26/2002 | 1639 | 06/28/2002 | 2057 |
09/01/2002 | 1002 | 09/03/2002 | 1332 |
12/21/2002 | 1124 | 12/23/2002 | 1510 |
01/04/2003 | 0703 | 01/06/2003 | 1529 |
08/09/2003 | 0037 | 08/11/2003 | 1456 |
08/23/2003 | 2101 | 08/25/2003 | 1456 |
10/12/2003 | 2025 | 10/14/2003 | 1414 |
11/29/2003 | 1840 | 12/01/2003 | 1502 |
01/24/2004 | 0743 | 01/26/2004 | 1505 |
02/14/2004 | 2316 | 02/17/2004 | 1520 |
09/04/2004 | 0250 | 09/07/2004 | 1419 |
10/09/2004 | 0745 | 10/11/2004 | 1739 |
10/16/2004 | 2250 | 10/18/2004 | 1355 |
11/13/2004 | 1849 | 11/15/2004 | 1521 |
12/03/2004 | 1940 | 12/06/2004 | 1454 |
01/15/2005 | 0652 | 01/18/2005 | 1446 |
01/22/2005 | 1449 | 01/24/2005 | 1444 |
06/10/2005 | 2118 | 06/13/2005 | 1603 |
08/05/2005 | 1701 | 08/30/2005 | 2001 |
10/07/2005 | 1658 | 11/16/2005 | 1945 |
11/18/2005 | 1801 | 11/21/2005 | 1521 |
11/23/2005 | 2201 | 11/25/2005 | 1550 |
11/26/2005 | 0101 | 11/28/2005 | 1543 |
12/03/2005 | 1043 | 12/05/2005 | 1559 |
01/14/2006 | 1822 | 01/17/2006 | 1512 |
01/21/2006 | 0708 | 01/23/2006 | 1515 |
02/03/2006 | 1903 | 02/06/2006 | 1533 |
02/25/2006 | 0730 | 02/27/2006 | 1420 |
07/01/2006 | 2225 | 07/03/2006 | 1413 |
11/28/2006 | 2143 | 12/01/2006 | 1911 |
06/09/2007 | 1150 | 06/11/2007 | 1300 |
07/07/2007 | 1642 | 07/09/2007 | 1304 |
08/03/2007 | 1601 | 11/19/2007 | 2135 |
12/22/2007 | 2112 | 12/24/2007 | 1504 |
05/25/2008 | 1815 | 05/27/2008 | 1335 |
06/06/2008 | 1512 | 06/09/2008 | 0238 |
05/08/2009 | 1535 | 05/11/2009 | 1254 |
10/30/2010 | 0400 | 11/01/2010 | 1200 |
Subject: | SGP/MWR/C1 - Missing data |
DataStreams: | sgpmwrlosC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/02/2008 | 1330 | 06/05/2008 | 1930 |
Subject: | SGP/ECOR/E14 - Sensor failure |
DataStreams: | sgp30ecorE14.b1 |
Description: | The water vapor output of the LI-7500 was too low, indicating that water vapor leaked into the sensor and that the chemicals are spent or there was an electronic failure. Latent heat flux was not normally affected by this problem. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2008 | 0000 | 11/30/2008 | 2300 |
Subject: | SGP/MWR/C1 - Increased noise in 31.4 GHz channel |
DataStreams: | sgpmwrlosC1.b1, sgpmwrtipC1.a1 |
Description: | The 31.4 GHz channel has been showing an increased level of noise which results in increased noise in the LWP retrievals. The instrument is tipping and all other parameters seem to be in the range. The elevate noise started in May, but it seems to have become slightly worse after summer. After November the noise level seems to have gone back to normal. |
Measurements: | sgpmwrtipC1.a1:
sgpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2010 | 2100 | 01/31/2011 | 1900 |
Subject: | SGP/SEBS/E14 - Instrument logger problem |
DataStreams: | sgp30ecorE14.b1 |
Description: | SEBS soil moisture measurements had been recorded as minimum soil moisture since the start of the SEBS installation at E14. Soil heat capacity calculations were also incorrect as a result of the problem. Instrument was mis-identified as ECOR in the DQPR. Since the SEBS datastream was not available to list, I have had to select an ECOR datastream just to be able to submit this DQR. Please add the SEBS to my listing of instrument systems for the purpose of submitting DQRs. |
Measurements: | sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2010 | 2100 | 01/31/2011 | 1900 |
Subject: | SGP/ECOR/E14 - Instrument logger problem |
DataStreams: | sgpecorE14.00, sgp30ecorE14.b1 |
Description: | Soil moisture measurements have been recorded as minimum soil moisture since the start of the SEBS installation at E14. Please check the wiring of the soil moisture sensors to the multiplexer as follows: 1) Check that the yellow/green leads run from MUX 2, Common, Even, H/L to CR1000 4H/Gr. 2) Check that the 50K resistor is connected between CR1000 4H and excitation VX3. Please report the results of these checks here. Thanks. |
Measurements: | sgpecorE14.00:
sgp30ecorE14.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/19/2010 | 2100 | 01/31/2011 | 1900 |
Subject: | SGP/SEBS/E14 - Soil Moisture Measurements Incorrect |
DataStreams: | sgp30ecorE14.b1 |
Description: | Soil Moisture measurements were incorrect because of a loose lead. |
Measurements: | sgp30ecorE14.b1:
|