DQR ID | Subject | Data Streams Affected |
---|---|---|
D110811.6 | PGH/TSI/M1 - Obstruction | pghtsiM1.00, pghtsimovieM1.a1, pghtsicldmaskM1.a1, pghtsiskycoverM1.b1, pghtsiskyimageM1.a1 |
D110815.1 | PGH/MWR/M1 - Incorrect calibration -- Reprocess | pghmwrlosM1.b1, pghmwrtipM1.a1 |
D110817.2 | PGH/MWRP/M1 - Incorrect time-please reprocess | pghmwrpM1.00, pghmwrpM1.b1, pghmwrprawM1.00 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/24/2011 | 0000 | 07/15/2011 | 0359 |
Subject: | PGH/TSI/M1 - Obstruction |
DataStreams: | pghtsiM1.00, pghtsimovieM1.a1, pghtsicldmaskM1.a1, pghtsiskycoverM1.b1, pghtsiskyimageM1.a1 |
Description: | The TSI was located in an obstructed view location for the beginning of the deployment. Trees and the AMF VANs are within the field of view of the instrument and appears to have affected the data processing. On 7/15/2011 the instrument was moved. |
Measurements: | pghtsicldmaskM1.a1:
pghtsiM1.00:
pghtsiskycoverM1.b1:
pghtsiskyimageM1.a1:
pghtsimovieM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/20/2011 | 0532 | 07/15/2011 | 0000 |
Subject: | PGH/MWR/M1 - Incorrect calibration -- Reprocess |
DataStreams: | pghmwrlosM1.b1, pghmwrtipM1.a1 |
Description: | When the radiometer started to collect data it was not at its final location with obstructed field of view. The radiometer was moved to its final location on 7/13/2011. After the radiometer was moved it started to collect good data for calibration. |
Measurements: | pghmwrtipM1.a1:
pghmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/20/2011 | 0000 | 07/15/2011 | 2025 |
Subject: | PGH/MWRP/M1 - Incorrect time-please reprocess |
DataStreams: | pghmwrpM1.00, pghmwrpM1.b1, pghmwrprawM1.00 |
Description: | The computer clock had the daylight saving on so the measurements time is off 1 hr. The setting was changed on 7/15. Data before this time need to be reprocessed. |
Measurements: | pghmwrpM1.b1:
pghmwrprawM1.00:
pghmwrpM1.00:
|